Detecting ventricular tachycardia and fibrillation by complexity measure

被引:247
作者
Zhang, XS [1 ]
Zhu, YS
Thakor, NV
Wang, ZZ
机构
[1] Shanghai Jiao Tong Univ, Coll Life Sci & Biotechnol, Dept Biomed Engn, Shanghai 200030, Peoples R China
[2] Johns Hopkins Sch Med, Dept Biomed Engn, Baltimore, MD 21205 USA
基金
中国国家自然科学基金;
关键词
arrhythmia detection; automatic external defibrillators; complexity measure; ventricular fibrillation;
D O I
10.1109/10.759055
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Sinus rhythm (SR), ventricular tachycardia (VT) and ventricular fibrillation (VF) belong to different nonlinear physiological processes with different complexity, In this study, we present a novel, and computationally fast method to detect VT and VF, which utilizes a complexity measure suggested by Lempel and Ziv [1]. For a specific window length (i.e., the length of data segment to be analyzed), the method first generates a 0-1 string by comparing the raw electrocardiogram (ECG) data to a selected suitable threshold, The complexity measure can be obtained from the 0-1 string only using two simple operations, comparison and accumulation. When the window length is 7 s, the detection accuracy for each of SR, VT, and VF is 100% for a test set of 204 body surface records (34 SR, 85 monomorphic VT, and 85 VF), Compared with other conventional time- and frequency-domain methods, such as rate and irregularity, VF-filter leakage, and sequential hypothesis testing, the new algorithm is simple, computationally efficient, and well suited for real-time implementation in automatic external defibrillators (AED's).
引用
收藏
页码:548 / 555
页数:8
相关论文
共 31 条
[21]   IDENTIFICATION OF VENTRICULAR-TACHYCARDIA USING INTRACAVITARY VENTRICULAR ELECTROGRAMS - ANALYSIS OF TIME AND FREQUENCY-DOMAIN PATTERNS [J].
LIN, DP ;
DICARLO, LA ;
JENKINS, JM .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1988, 11 (11) :1592-1606
[22]   CORRELATION DIMENSION AND SYSTEMATIC GEOMETRIC EFFECTS [J].
NERENBERG, MAH ;
ESSEX, C .
PHYSICAL REVIEW A, 1990, 42 (12) :7065-7074
[23]   COMPLEX DYNAMICS UNDERLYING THE HUMAN ELECTROCARDIOGRAM [J].
RAVELLI, F ;
ANTOLINI, R .
BIOLOGICAL CYBERNETICS, 1992, 67 (01) :57-65
[24]  
REID PR, 1978, PACING CLIN ELECTROP, V1, P127
[25]   EVALUATION OF TECHNIQUES FOR RECOGNITION OF VENTRICULAR ARRHYTHMIAS BY IMPLANTED DEVICES [J].
RIPLEY, KL ;
BUMP, TE ;
ARZBAECHER, RC .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1989, 36 (06) :618-624
[26]  
Stewart J A, 1996, J Cardiovasc Nurs, V10, P37
[27]   MULTIWAY SEQUENTIAL HYPOTHESIS-TESTING FOR TACHYARRHYTHMIA DISCRIMINATION [J].
THAKOR, NV ;
NATARAJAN, A ;
TOMASELLI, GF .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1994, 41 (05) :480-487
[28]   VENTRICULAR-TACHYCARDIA AND FIBRILLATION DETECTION BY A SEQUENTIAL HYPOTHESIS-TESTING ALGORITHM [J].
THAKOR, NV ;
ZHU, YS ;
PAN, KY .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1990, 37 (09) :837-843
[30]   A COMPARISON OF 4 NEW TIME-DOMAIN TECHNIQUES FOR DISCRIMINATING MONOMORPHIC VENTRICULAR-TACHYCARDIA FROM SINUS RHYTHM USING VENTRICULAR WAVE-FORM MORPHOLOGY [J].
THRONE, RD ;
JENKINS, JM ;
DICARLO, LA .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1991, 38 (06) :561-570