Concordance between capnography and arterial blood gas measurements of carbon dioxide in acute asthma

被引:47
作者
Corbo, J [1 ]
Bijur, P [1 ]
Lahn, M [1 ]
Gallagher, EJ [1 ]
机构
[1] Albert Einstein Coll Med, Dept Emergency Med, Bronx, NY 10461 USA
关键词
D O I
10.1016/j.annemergmed.2004.12.005
中图分类号
R4 [临床医学];
学科分类号
1002 [临床医学]; 100602 [中西医结合临床];
摘要
Study objective: We examine the concordance between end-tidal partial pressure Of CO2 (PetCO(2)) measured by capnography and arterial partial pressure of carbon dioxide (PaCO2) obtained by arterial blood gas in acute asthmatic patients presenting to the emergency department. Methods: This was a prospective observational cohort study of acutely ill adult asthmatic patients undergoing an arterial blood gas measurement as part of their evaluation. PetCO(2) was recorded during exhalation into a capnograph while arterial blood was pulsing in the arterial blood gas tubing. Concordance between PetCO(2) and PaCO2 was displayed as a Bland-Altman matrix, using prespecified limits of agreement of +/- 5 mm Hg difference between PetCO(2) and PaCO2 in each patient. Results: The mean difference between the PetCO(2) and PaCO2 levels was 1.0 mm Hg (95% confidence interval -0.1 to 2.0 mm Hg), with a median of 0 mm Hg. Of the 39 patients enrolled, 37 (95%) fell within the a priori limits of agreement. Conclusion: In adult asthmatic patients with acute exacerbations, concordance between PetCO(2) measured by capnography and PaCO2 measured by arterial blood gas was high. These findings must be validated before capnography replacement of arterial blood gas as an accurate means of assessing alveolar ventilation in acute asthma is recommended.
引用
收藏
页码:323 / 327
页数:5
相关论文
共 13 条
[1]
Noninvasive capnometry in a pediatric population with respiratory emergencies [J].
Abramo, TJ ;
Wiebe, RA ;
Scott, SM ;
Primm, PA ;
McIntyre, D ;
Mydler, T .
PEDIATRIC EMERGENCY CARE, 1996, 12 (04) :252-254
[2]
COMPARISON OF PEDIATRIC END-TIDAL CO2 MEASURED WITH NASAL ORAL CANNULA CIRCUIT AND CAPILLARY PCO2 [J].
ABRAMO, TJ ;
COWAN, MR ;
SCOTT, SM ;
PRIMM, PA ;
WIEBE, RA ;
SIGNS, M .
AMERICAN JOURNAL OF EMERGENCY MEDICINE, 1995, 13 (01) :30-33
[3]
Noninvasive capnometry monitoring for respiratory status during pediatric seizures [J].
Abramo, TJ ;
Wiebe, RA ;
Scott, S ;
Goto, CS ;
McIntire, DD .
CRITICAL CARE MEDICINE, 1997, 25 (07) :1242-1246
[4]
BARTON CW, 1995, ANN EMERG MED, V25, P560
[5]
A NOTE ON THE USE OF THE INTRACLASS CORRELATION-COEFFICIENT IN THE EVALUATION OF AGREEMENT BETWEEN 2 METHODS OF MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
COMPUTERS IN BIOLOGY AND MEDICINE, 1990, 20 (05) :337-340
[6]
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[7]
BONGARD F, 1992, WESTERN J MED, V156, P57
[8]
FROM RP, 1988, ANESTH ANALG, V67, P884
[9]
Capnometry for noninvasive continuous monitoring of metabolic status in pediatric diabetic ketoacidosis [J].
Garcia, E ;
Abramo, TJ ;
Okada, P ;
Guzman, DD ;
Reisch, JS ;
Wiebe, RA .
CRITICAL CARE MEDICINE, 2003, 31 (10) :2539-2543
[10]
RELATIONSHIP BETWEEN ARTERIAL CARBON-DIOXIDE AND END-TIDAL CARBON-DIOXIDE WHEN A NASAL SAMPLING PORT IS USED [J].
MCNULTY, SE ;
ROY, J ;
TORJMAN, M ;
SELTZER, JL .
JOURNAL OF CLINICAL MONITORING, 1990, 6 (02) :93-98