摘要
BACKGROUND: High-resolution melting is an emerging technique for detection of nucleic acid sequence variations. Developments in instrumentation and saturating intercalating dyes have made accurate high-resolution melting analysis possible and created opportunities to use this technology in diagnostic settings. We evaluated 2 high-resolution melting instruments for screening BRCA1 and BRCA2 mutations. METHODS: To cover the complete coding region and splice sites, we designed 112 PCR amplicons (136-435 bp), amplifiable with a single PCR program. LCGreen (R) Plus was used as the intercalating dye. High-resolution melting analysis was performed on the 96-well Light-scanner (TM) (Idaho Technology Inc.) and the 96-well LightCycler (R) 480 (Roche) instruments. We evaluated sensitivity by analyzing 212 positive controls scattered over almost all amplicons and specificity by blind screening of 22 patients for BRCA1 and BRCA2. In total, we scanned 3521 fragments. RESULTS: All 212 known heterozygous sequence variants were detected on the Lightscanner by analysis on normal sensitivity setting. On the LightCycler 480, the standard instrument sensitivity setting of 0.3 had to be increased to 0.7 to detect all variants, decreasing the specificity to 95.9% (vs 98.7% for the Lightscanner). CONCLUSIONS: Previously, we screened BRCA1/2by direct sequencing of the large exon 11 and denaturing gel gradient electrophoresis (DGGE) for all other coding exons. Since the introduction of high-resolution melting, our turnaround time has been one third of that with direct sequencing and DGGE, as post-PCR handling is no longer required and the software allows fast analyses. High-resolution melting is a rapid, cost-efficient, sensitive method simple enough to be readily implemented in a diagnostic laboratory. (c) 2008 American Association for Clinical Chemistry.
引用
收藏
页码:982 / 989
页数:8
相关论文
共 19 条
[1]  
Arnold N, 1999, HUM MUTAT, V14, P333, DOI 10.1002/(SICI)1098-1004(199910)14:4<333::AID-HUMU9>3.3.CO
[2]  
2-3
[3]   A comparison of high-resolution melting analysis with denaturing high-performance liquid chromatography for mutation scanning - Cystic fibrosis transmembrane conductance regulator gene as a model [J].
Chou, LS ;
Lyon, E ;
Wittwer, CT .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2005, 124 (03) :330-338
[4]   BRCA1 and BRCA2 germline mutation spectrum and frequencies in Belgian breast/ovarian cancer families [J].
Claes, K ;
Poppe, B ;
Coene, I ;
De Paepe, A ;
Messiaen, L .
BRITISH JOURNAL OF CANCER, 2004, 90 (06) :1244-1251
[5]   Streamlined assessment of gene variants by high resolution melt profiling utilizing the ornithine Transcarbamylase gene as a model system [J].
Dobrowolski, Steven F. ;
Ellingson, Clinton E. ;
Caldovic, Ljubica ;
Tuchman, Mendel .
HUMAN MUTATION, 2007, 28 (11) :1133-1140
[6]   High throughput fluorescence-based conformation-sensitive gel electrophoresis (F-CSGE) identifies six unique BRCA2 mutations and an overall low incidence of BRCA2 mutations in high-risk BRCA1-negative breast cancer families [J].
Ganguly, T ;
Dhulipala, R ;
Godmilow, L ;
Ganguly, A .
HUMAN GENETICS, 1998, 102 (05) :549-556
[7]   Diagnostic accuracy of methods for the detection of BRCA1 and BRCA2 mutations:: a systematic review [J].
Gerhardus, Ansgar ;
Schleberger, Henriette ;
Schlegelberger, Brigitte ;
Gadzicki, Dorothea .
EUROPEAN JOURNAL OF HUMAN GENETICS, 2007, 15 (06) :619-627
[8]   Expanded instrument comparison of amplicon DNA melting analysis for mutation scanning and genotyping [J].
Herrmann, Mark G. ;
Durtschi, Jacob D. ;
Wittwer, Carl T. ;
Voelkerding, Karl V. .
CLINICAL CHEMISTRY, 2007, 53 (08) :1544-1548
[9]   Amplicon DNA melting analysis for mutation scanning and genotyping: Cross-platform comparison of instruments and dyes [J].
Herrmann, MG ;
Durtschi, JD ;
Bromley, LK ;
Wittwer, CT ;
Voelkerding, KV .
CLINICAL CHEMISTRY, 2006, 52 (03) :494-503
[10]   Mutation scanning the GJB1 gene with high-resolution melting analysis:: Implications for mutation scanning of genes for Charcot-Marie-Tooth disease [J].
Kennerson, L. ;
Warburton, Trent ;
Nelis, Eva ;
Brewer, Megan ;
Polly, Patsie ;
De Jonghe, Peter ;
Timmerman, Vincent ;
Nicholson, Garth A. .
CLINICAL CHEMISTRY, 2007, 53 (02) :349-352