Dose distribution in conventional CT and spiral CT and the question of dose reduction in spiral CT

被引:2
作者
Hidajat, N [1 ]
Schroder, RJ [1 ]
Vogl, T [1 ]
Maurer, J [1 ]
Steger, W [1 ]
Felix, R [1 ]
机构
[1] Strahlenklin & Poliklin, D-13353 Berlin, Germany
来源
RADIOLOGE | 1998年 / 38卷 / 05期
关键词
spiral CT; dose distribution; dose reduction;
D O I
10.1007/s001170050378
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective:To compare spiral CT with conventional CT with regard to the dose distribution in the irradiated body and to examine when an increase in the table feed in spiral CT leads to a dose reduction compared with conventional CT. Methods: In order to show the dose distribution,films were exposed in an acrylic phantom. The nominal slice thickness was 5 mm; the slice distance in conventional CT was 5 and 10 mm;the table feed in spiral CT was 5 and 10 mm per rotation (pitch 1 and 2). The integral doses delivered in spiral CT and conventional CT were related to the table feed per rotation or slice distance, respectively, and were compared arithmetically with each other. Results: Spiral CT showed reduced dose fluctuations in axial direction compared with conventional CT. Only in conventional CT with a slice distance of 5 mm did the axial dose distribution show relative dose peaks near the rotation axis. The longer the body region to be examined is, the less table feed is necessary in spiral CT for a dose reduction. From a length of 15 times the slice thickness, a pitch of >1.1 leads to a reduction of the integral dose compared with conventional CT. Conclusions: The continuous ta bie feed in spiral CT results in evenly distributed axial dose distributions compared with the discontinuous scanning in spiral technique can give an important contribution to dose reduction in CT.
引用
收藏
页码:438 / 443
页数:6
相关论文
共 14 条
[1]   PULMONARY NODULE - EVALUATION WITH SPIRAL VOLUMETRIC CT [J].
COSTELLO, P ;
ANDERSON, W ;
BLUME, D .
RADIOLOGY, 1991, 179 (03) :875-876
[2]   Low-dose computed tomography of the thorax. Experimental and clinical investigations [J].
Diederich, S ;
Lenzen, H ;
Puskas, Z ;
Koch, AT ;
Yelbuz, TM ;
Eameri, M ;
Roos, N ;
Peters, PE .
RADIOLOGE, 1996, 36 (06) :475-482
[3]   SPIRAL CT OF THE PANCREAS WITH MULTIPLANAR DISPLAY [J].
FISHMAN, EK ;
WYATT, SH ;
NEY, DR ;
KUHLMAN, JE ;
SIEGELMAN, SS .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1992, 159 (06) :1209-1215
[4]   THE ACCURACY OF CT ANGIOGRAPHY FOR THE DEMONSTRATION OF RENAL-ARTERY STENOSIS [J].
GALANSKI, M ;
PROKOP, M ;
CHAVAN, A ;
SCHAEFER, C ;
JANDELEIT, K ;
OLBRICHT, C .
FORTSCHRITTE AUF DEM GEBIETE DER RONTGENSTRAHLEN UND DER NEUEN BILDGEBENDEN VERFAHREN, 1994, 161 (06) :519-525
[5]  
HIDAJAT N, 1997, BERICHT OTTO HUG STR, V15, P170
[6]  
HIDAJAT N, 1996, ROFO FORTSCHR RONTG, V165, P11
[7]   The quality of 3D reconstructions from 1.0 and 1.5 pitch helical and conventional CT [J].
Hopper, KD ;
Pierantozzi, D ;
Potok, PS ;
Kasales, CJ ;
TenHave, TR ;
Meilstrup, JW ;
VanSlyke, MA ;
Mahraj, R ;
Westacott, S ;
Hartzel, JS .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1996, 20 (05) :841-847
[8]   SPIRAL VOLUMETRIC CT WITH SINGLE-BREATH-HOLD TECHNIQUE, CONTINUOUS TRANSPORT, AND CONTINUOUS SCANNER ROTATION [J].
KALENDER, WA ;
SEISSLER, W ;
KLOTZ, E ;
VOCK, P .
RADIOLOGY, 1990, 176 (01) :181-183
[9]   A COMPARISON OF CONVENTIONAL AND SPIRAL CT - AN EXPERIMENTAL-STUDY ON THE DETECTION OF SPHERICAL LESIONS [J].
KALENDER, WA ;
POLACIN, A ;
SUSS, C .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1994, 18 (02) :167-176
[10]   RADIATION EXPOSURE DURING CT USING STANDARD AND SPIRAL TECHNIQUES [J].
LANGKOWSKI, JH ;
POGODA, P ;
HESS, A .
FORTSCHRITTE AUF DEM GEBIETE DER RONTGENSTRAHLEN UND DER NEUEN BILDGEBENDEN VERFAHREN, 1994, 161 (01) :3-11