Aspects of MR image distortions in radiotherapy treatment planning

被引:118
作者
Fransson, A
Andreo, P
Pötter, R
机构
[1] Univ Hosp Vienna, AKH 3B, Dept Radiotherapy & Radiat Biol, A-1090 Vienna, Austria
[2] IAEA, Div Human Hlth, Dosimetry & Med Radiat Phys Sect, Vienna, Austria
关键词
magnetic resonance imaging; geometric distortion; radiotherapy treatment planning; image registration;
D O I
10.1007/PL00002385
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background: Registration of computed tomography (CT) and magnetic resonance (MR) images are commonly performed to define the different target regions used in radiotherapy treatment planning (RTTP). The accuracy of target definition will then depend on the spatial accuracy of the CT and MR data, and on the technique used to register the images. CT images are usually regarded as geometrically correct, while MR images are known to suffer from geometric distortion. The aim of this paper is to discuss the possible impact of MR image distortions in the radiotherapy treatment planning process. Methods: The origin, magnitude, and relative impact of the different sources of geometric distortions that affect the MR image data at different magnetic fields and for different acquisition settings are described. Techniques for distortion correction are reviewed, and their Limitations are outlined. The sensitivity of image registration techniques to the presence of geometric distortions in the MR data is discussed. Finally, art overview of image registration techniques used and results obtained in clinical radiotherapy treatment planning applications is given. Results: Spatial distortions in MR images vary with field strength and with the image acquisition protocol. The spatial accuracy generally decreases with distance from the magnet isocenter. Distortion correction techniques based on phantom evaluations cannot adequately model patient-induced distortions. Conclusion: Image protocols with high gradient bandwidths should be used to reduce the spatial distortions in MR images. Correction techniques based only on phantom measurements could be sufficient at Low magnetic fields, while at higher fields additional corrections of patient-related distortions might be needed. Registration techniques based on matching of Landmark points Located far from the magnet isocenter are especially prone to MR distortions.
引用
收藏
页码:59 / 73
页数:17
相关论文
共 58 条
[1]  
[Anonymous], 1993, 50 ICRU
[2]  
[Anonymous], REV MAGN RESON MED
[3]   Improvements in absorbed dose measurements for external radiation therapy using ferrous dosimeter gel and MR imaging (FeMRI) [J].
Back, SAJ ;
Magnusson, P ;
Fransson, A ;
Olsson, LE ;
Montelius, A ;
Holmberg, O ;
Andreo, P ;
Mattsson, S .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (02) :261-276
[4]   ANALYSIS OF MACHINE-DEPENDENT AND OBJECT-INDUCED GEOMETRIC DISTORTION IN 2DFT MR IMAGING [J].
BAKKER, CJG ;
MOERLAND, MA ;
BHAGWANDIEN, R ;
BEERSMA, R .
MAGNETIC RESONANCE IMAGING, 1992, 10 (04) :597-608
[5]   Radiotherapy treatment planning of brain tumours using MRI alone [J].
Beavis, AW ;
Gibbs, P ;
Dealey, RA ;
Whitton, VJ .
BRITISH JOURNAL OF RADIOLOGY, 1998, 71 (845) :544-548
[6]   A TECHNIQUE FOR ACCURATE MAGNETIC-RESONANCE-IMAGING IN THE PRESENCE OF FIELD INHOMOGENEITIES [J].
CHANG, H ;
FITZPATRICK, JM .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1992, 11 (03) :319-329
[7]   IMAGE CORRELATION TECHNIQUES IN RADIATION-THERAPY TREATMENT PLANNING [J].
CHEN, GTY ;
PELIZZARI, CA .
COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 1989, 13 (03) :235-240
[8]   TOTAL INHOMOGENEITY CORRECTION INCLUDING CHEMICAL-SHIFTS AND SUSCEPTIBILITY BY VIEW ANGLE TILTING [J].
CHO, ZH ;
KIM, DJ ;
KIM, YK .
MEDICAL PHYSICS, 1988, 15 (01) :7-11
[9]   CORRECTION OF INTENSITY VARIATIONS IN MR-IMAGES FOR COMPUTER-AIDED TISSUE CLASSIFICATION [J].
DAWANT, BM ;
ZIJDENBOS, AP ;
MARGOLIN, RA .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1993, 12 (04) :770-781
[10]  
DEBATIN JF, 1998, INTERVENTIONAL MAGNE