Comparison of the mechanical thrombectomy efficacy of the Amplatz thrombectomy device and the Cragg thrombolytic brush in vitro

被引:6
作者
Grimm, J [1 ]
Müller-Hülsbeck, S [1 ]
Heller, M [1 ]
机构
[1] Univ Kiel, Radiol Diagnost Klin, D-24105 Kiel, Germany
关键词
mechanical thrombectomy; dialysis fistulas; interventional radiology;
D O I
10.1097/00004424-200104000-00002
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
RATIONALE AND OBJECTIVES. TO determine the efficacy of thrombectomy (without thrombolytic agents) for the Amplatz thrombectomy device (ATD) and the Cragg thrombolytic brush catheter (CBC) in vitro, METHODS. Thrombectomy was performed with the ATD or CBC (6F) in a Row model. Embolus sizes, weight, remaining thrombus, and activation time were evaluated, RESULTS. NO significant difference in the activation time was found. The CBC produced significantly less embolism (3.3% vs, 0.03% in the 5-mm and 89% vs. 0.5% in the 7-mm model), but also much more thrombus remained in the system than with the ATD (1% vs. 41% in the 5-mm and 0.1% vs, 62% in the 7-mm model). CONCLUSIONS. The ATD can remove almost all thrombus (99%), whereas the CBC removes only up to 60%, producing fewer emboli than the ATD, This might be due to the lower rotational speed of the CBC compared with the ATD, which is 20 Limes greater. The soft nylon brush offers less resistance and shear force toward the thrombus than the stainless-steel impeller of the ATD, Because of the large amount of remaining thrombus, the CBC should not be used without lytic agents.
引用
收藏
页码:204 / 209
页数:6
相关论文
共 23 条
[1]   MECHANICAL VERSUS PHARMACOMECHANICAL THROMBOLYSIS FOR THE TREATMENT OF THROMBOSED DIALYSIS ACCESS GRAFTS [J].
BEATHARD, GA .
KIDNEY INTERNATIONAL, 1994, 45 (05) :1401-1406
[2]   New thrombolytic brush catheter in thrombosed polytetrafluoroethylene dialysis grafts: Preclinical animal study [J].
Castaneda, F ;
Wyffels, PL ;
Pater, JC ;
Swischuk, JL ;
Li, RZ ;
Cole, BA ;
Cragg, AH .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 1998, 9 (05) :793-798
[3]   Synthetic dialysis shunts: Thrombolysis with the Cragg thrombolytic brush catheter [J].
Dolmatch, BL ;
Casteneda, F ;
McNamara, TO ;
Zemel, G ;
Lieber, M ;
Cragg, AH .
RADIOLOGY, 1999, 213 (01) :180-184
[4]  
EGGERS PW, 1992, INTERNAL RECORD HLTH
[5]   Percutaneous intervention for permanent hemodialysis access: A review [J].
Gray, RJ .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 1997, 8 (03) :313-327
[6]  
LIEPSCH DW, 1986, BIORHEOLOGY, V23, P395
[7]   Enhancement of in vitro effectiveness for hydrodynamic thrombectomy devices -: Simultaneous high-pressure rt-PA application [J].
Müller-Hülsbeck, S ;
Bathe, M ;
Grimm, J ;
Heller, M .
INVESTIGATIVE RADIOLOGY, 1999, 34 (08) :536-542
[8]   Mechanical assistance for enhancement of in vitro effectiveness for hydrodynamic thrombectomy [J].
Müller-Hülsbeck, S ;
Grimm, J ;
Suwelack, D ;
Heller, M .
INVESTIGATIVE RADIOLOGY, 1999, 34 (11) :669-677
[9]   Mechanical thrombectomy of acute thrombosis in transjugular intrahepatic portosystemic shunts [J].
Müller-Hülsbeck, S ;
Höpfner, M ;
Hilbert, C ;
Krämer-Hansen, H ;
Heller, M .
INVESTIGATIVE RADIOLOGY, 2000, 35 (06) :385-391
[10]   Rheolytic hydrodynamic thrombectomy for percutaneous treatment of acutely occluded infra-aortic native arteries and bypass grafts -: Midterm follow-up results [J].
Müller-Hülsbeck, S ;
Kalinowski, M ;
Heller, M ;
Wagner, HJ .
INVESTIGATIVE RADIOLOGY, 2000, 35 (02) :131-140