Novel system for real-time ex vivo lactate monitoring in human whole blood

被引:13
作者
Gfrerer, RJ
Brunner, GA
Trajanoski, Z
Schaupp, L
Sendlhofer, G
Skrabal, F
Jobst, G
Moser, I
Urban, G
Pieber, TR
Wach, P
机构
[1] Graz Univ Technol, Inst Biomed Engn, Dept Biophys, A-8010 Graz, Austria
[2] Karl Franzens Univ Graz, Dept Internal Med Diabet & Metab, A-8036 Graz, Austria
[3] Krankenhaus Barmherzigen Bruderr, Dept Internal Med, A-8010 Graz, Austria
[4] Univ Freiburg, Inst Microsyst Techn, Freiburg, Germany
[5] Ludwig Boltzmann Inst Tech Lebenhilfen, A-8010 Graz, Austria
关键词
blood lactate monitoring; heparin coating; biosensor-array;
D O I
10.1016/S0956-5663(98)00090-6
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
The objective of the study was to evaluate the performance of an amperometric enzyme based lactate sensor and to investigate the possibility of replacing a double lumen catheter based blood withdrawal system with a heparin coated single lumen system. The inner lumen of a double lumen catheter which was placed in a peripheral vein was perfused with heparin solution. The outer lumen was used to collect heparinized blood samples at a defined flow rate. The single lumen system was attached to a heparinized catheter which was also placed in a peripheral vein. The undiluted blood samples were collected at a specified flow rate, A sensor flow chamber incorporating an amperometric thin-film lactate microbiosensor was placed in the sampling line for real-time lactate monitoring. Plasma lactate concentrations were measured during frequently performed hyperlactatemia bicycle ergometer experiments in six healthy volunteers (age 25.8 +/- 2.8 years, BMI 22.7 +/- 1 kg/m(2)). Additionally, plasma lactate was measured in realtime using the lactate sensors. The first three experiments were performed with a double lumen based catheter system whereas the following three experiments were performed with a heparin coated catheter system. The correlation coefficients of sensor readings and laboratory analyzer results in all six experiments were between 0.93 and 0.99, respectively (P < 0.001). The miniaturized lactate sensors showed a linear range up to 25 mmol/l lactate concentration and 95% response times < 30 s in undiluted serum. During the experiments maximum lactate concentrations of 14 mmol/l were achieved. Improvements of system performance using heparin coated catheter systems could be shown. The overall SD of the sensor readings compared to laboratory results using three double lumen catheter based systems was 0.91 mmol/l whereas the SD using three heparin coated systems was 0.65 mmol/l, In summary, real-time monitoring of lactate in human whole blood is feasible with such a device and can be improved by using heparin coated catheter systems. (C) 1998 Elsevier Science S.A. All rights reserved.
引用
收藏
页码:1271 / 1278
页数:8
相关论文
共 17 条
[1]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[2]   PROGRESS IN PRACTICAL ENDOCRINOLOGY - THE GLUCOSENSOR UNITEC ULM - A PORTABLE MONITOR FOR CONTINUOUS BLOOD-GLUCOSE MEASUREMENT [J].
BRUCKEL, J ;
ZIER, H ;
KERNER, W ;
PFEIFFER, EF .
HORMONE AND METABOLIC RESEARCH, 1990, 22 (07) :382-384
[3]   PORTABLE SYSTEM FOR CONTINUOUS EX-VIVO MEASUREMENTS OF LACTATE [J].
HAKANSON, H ;
KYROLAINEN, M ;
MATTIASSON, B .
BIOSENSORS & BIOELECTRONICS, 1993, 8 (3-4) :213-217
[4]   A NEW METHOD FOR COVALENT COUPLING OF HEPARIN AND OTHER GLYCOSAMINOGLYCANS TO SUBSTANCES CONTAINING PRIMARY AMINO-GROUPS [J].
HOFFMAN, J ;
LARM, O ;
SCHOLANDER, E .
CARBOHYDRATE RESEARCH, 1983, 117 (JUN) :328-331
[5]   Mass producible miniaturized flow through a device with a biosensor array [J].
Jobst, G ;
Moser, I ;
Svasek, P ;
Varahram, M ;
Trajanoski, Z ;
Wach, P ;
Kotanko, P ;
Skrabal, F ;
Urban, G .
SENSORS AND ACTUATORS B-CHEMICAL, 1997, 43 (1-3) :121-125
[6]   Thin-film microbiosensors for glucose-lactate monitoring [J].
Jobst, G ;
Moser, I ;
Varahram, M ;
Svasek, P ;
Aschauer, E ;
Trajanoski, Z ;
Wach, P ;
Kotanko, P ;
Skrabal, F ;
Urban, G .
ANALYTICAL CHEMISTRY, 1996, 68 (18) :3173-3179
[7]   A NEW NON-THROMBOGENIC SURFACE PREPARED BY SELECTIVE COVALENT BINDING OF HEPARIN VIA A MODIFIED REDUCING TERMINAL RESIDUE [J].
LARM, O ;
LARSSON, R ;
OLSSON, P .
BIOMATERIALS MEDICAL DEVICES AND ARTIFICIAL ORGANS, 1983, 11 (2-3) :161-173
[8]  
Larm O., 1989, CLIN APPL, P597
[9]  
LICHIEN H, 1996, PERFUSION, V11, P256
[10]   ANTITHROMBIN ACTIVITY OF SURFACE-BOUND HEPARIN STUDIED UNDER FLOW CONDITIONS [J].
LINDHOUT, T ;
BLEZER, R ;
SCHOEN, P ;
WILLEMS, GM ;
FOUACHE, B ;
VERHOEVEN, M ;
HENDRIKS, M ;
CAHALAN, L ;
CAHALAN, PT .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1995, 29 (10) :1255-1266