Development of a contactless conductivity detector cell for 1.6 mm OD (1/16th inch) HPLC tubing and micro-bore columns with on-column detection

被引:19
作者
Gillespie, Eoin [2 ]
Connolly, Damian [2 ]
Macka, Mirek [1 ]
Hauser, Peter [3 ]
Paull, Brett [1 ,2 ]
机构
[1] Dublin City Univ, Natl Ctr Sensor Res, Dublin 9, Ireland
[2] Dublin City Univ, Ctr Bioanalyt Sci, Dublin 9, Ireland
[3] Univ Basel, Dept Chem, Basel, Switzerland
关键词
D O I
10.1039/b803038c
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A capacitively coupled contactless conductivity detector cell was designed and constructed suitable for standard HPLC 1.6 mm o.d. (1/16 '') tubing and columns. Bode plots were acquired in order to determine the optimum input frequency for the new detector cell, with three feedback resistors of increasing resistance (1 M Omega, 3.3 M Omega and 4.7 M Omega) alternately fitted to maximize sensitivity. To evaluate the new detector geometry, a 190 mm x 0.381 mm i.d. (1.59 mm o.d.) column packed with Dionex OmniPac 8.5 mu m PAX-100 anion exchange resin was used, with an eluent of 0.5 mM sodium benzoate containing 2% methanol, pumped at 20 mu L min(-1), with detection taking place 'on-column'. Standard analytical performance criteria were used to compare the test detector with a commercially available TraceDec (CD)-D-4 capillary ( 360 mu m o.d.) detector, with the new detector cell matching the performance of the commercial instrument in terms of linearity, sensitivity and reproducibility. In 'on-column' detection mode, the detector cell could be moved during ( or between) chromatographic runs along the length of the packed column bed, to shorten the effective length of the column and thus reduce the retention time of strongly retained sample components, whilst maintaining the resolution of weakly retained components. In addition, the detector could also be applied to the evaluation of the stationary phase packing homogeneity by physically scanning the entire length of the micro-bore column with the moveable detector cell.
引用
收藏
页码:1104 / 1110
页数:7
相关论文
共 30 条
  • [1] SIMULTANEOUS CONDUCTIVITY AND PERMITTIVITY DETECTOR WITH A SINGLE CELL FOR LIQUID-CHROMATOGRAPHY
    ALDER, JF
    FIELDEN, PR
    CLARK, AJ
    [J]. ANALYTICAL CHEMISTRY, 1984, 56 (06) : 985 - 988
  • [2] ANDERSEN, 1989, Patent No. 4888824
  • [3] BURKHALTER TS, 1964, HIGH FREQUENCY CONDU, P215
  • [4] Fabrication of poly(methyl methacrylate) microfluidic chips by redox-initiated polymerization
    Chen, Jiang
    Lin, Yuehe
    Chen, Gang
    [J]. ELECTROPHORESIS, 2007, 28 (16) : 2897 - 2903
  • [5] Analysis of inorganic and small organic ions by CE with amperometric detection
    Chen, Rongsheng
    Cheng, Han
    Wu, Wenzhan
    Ai, Xiaoni
    Huang, Weihua
    Wang, Zongli
    Cheng, Jieke
    [J]. ELECTROPHORESIS, 2007, 28 (19) : 3347 - 3361
  • [6] A thin cover glass chip for contactless conductivity detection in microchip capillary electrophoresis
    Chen, Zuanguang
    Li, Quanwen
    Li, Oulian
    Zhou, Xie
    Lan, You
    Wei, Yuanfang
    Mo, Jinyuan
    [J]. TALANTA, 2007, 71 (05) : 1944 - 1950
  • [7] da Silva JAF, 1998, ANAL CHEM, V70, P4339
  • [8] CE microchips:: An opened gate to food analysis
    Escarpa, Alberto
    Gonzalez, Maria Cristina
    Crevillen, Agustin Gonzalez
    Blasco, Antonio Javier
    [J]. ELECTROPHORESIS, 2007, 28 (06) : 1002 - 1011
  • [9] Use of contactless conductivity detection for non-invasive characterisation of monolithic stationary-phase coatings for application in capillary ion chromatography
    Gillespie, Eoin
    Connolly, Damian
    Macka, Miroslav
    Nesterenko, Pavel N.
    Paull, Brett
    [J]. ANALYST, 2007, 132 (12) : 1238 - 1245
  • [10] Evaluation of capillary ion exchange stationary phase coating distribution and stability using radial capillary column contactless conductivity detection
    Gillespie, Eoin
    Macka, Miroslav
    Connolly, Damien
    Paull, Brett
    [J]. ANALYST, 2006, 131 (08) : 886 - 888