Image quality and reconstruction intervals of dual-source CT coronary angiography - Recommendations for ECG-pulsing windowing

被引:129
作者
Leschka, Sebastian
Scheffel, Hans
Desbiolles, Lotus
Plass, Andre
Gaemperli, Oliver
Valenta, Ines
Husmann, Lars
Flohr, Thomas G.
Genoni, Michele
Marincek, Borut
Kaufmann, Philipp A.
Alkadhi, Hatem
机构
[1] Univ Zurich Hosp, Inst Diagnost Radiol, CH-8091 Zurich, Switzerland
[2] Univ Zurich Hosp, Cardiovasc Surg Clin, CH-8091 Zurich, Switzerland
[3] Univ Zurich Hosp, Ctr Cardiovasc, CH-8091 Zurich, Switzerland
[4] Siemens Med Solut, Forchheim, Germany
[5] Univ Zurich Hosp, Zurich Ctr Integrat Human Physiol, CH-8091 Zurich, Switzerland
关键词
dual-source computed tomography; coronary angiography; reconstruction interval; image quality;
D O I
10.1097/RLI.0b013e31803b93cf
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To evaluate reconstruction intervals and image quality in dual-source computed tomography (DSCT) coronary angiography for optimal placement of the ECG-pulsing window. Materials and Methods: DSCT coronary angiography was performed in 60 patients. Thirteen datasets were reconstructed in 5% increments from 20-80%. Two readers independently assessed image quality of each segment in each percentage-interval, using scores ranging from 1 (no motion artifacts) to 4 (nonevaluable). Results: Mean heart rate (HR) was 69.0 +/- 18.9 beats per minute (bpm) (range, 35-117 bpm). Diagnostic image quality (scores 1-3) was found in 97.8% of all segments (763 of 780). The 70% RR-interval provided best image quality in all patients and all HRs. The narrowest reconstruction window providing diagnostic image quality was 60-70% for HR <60 bpm, 60-80% for 60-70 bpm, 55-80% for 70-80 bpm, and 30-80% for HR >80 bpm. Conclusions: DSCT coronary angiography provides best image quality for various HRs at 70%. The ECG-pulsing window can be adapted according to the HR while maintaining diagnostic image quality.
引用
收藏
页码:543 / 549
页数:7
相关论文
共 35 条
[1]   Contrast-enhanced coronary artery visualization by dual-source computed tomography - Initial experience [J].
Achenbach, S ;
Ropers, D ;
Kuettner, A ;
Flohr, T ;
Ohnesorge, B ;
Bruder, H ;
Theessen, H ;
Karakaya, M ;
Daniel, WG ;
Bautz, W ;
Kalender, WA ;
Anders, K .
EUROPEAN JOURNAL OF RADIOLOGY, 2006, 57 (03) :331-335
[2]   In-plane coronary arterial motion velocity: Measurement with electron-beam CT [J].
Achenbach, S ;
Ropers, D ;
Holle, J ;
Muschiol, G ;
Daniel, WG ;
Moshage, W .
RADIOLOGY, 2000, 216 (02) :457-463
[3]  
Austen W G, 1975, Circulation, V51, P5
[4]  
BELHAYE D, 2007, EUR RADIOL, V17, P591
[5]   Computed tomography coronary angiography with 370-millisecond gantry rotation time - Evaluation of the best image reconstruction interval [J].
Bley, TA ;
Ghanem, NA ;
Foell, D ;
Uhl, M ;
Geibel, A ;
Bode, C ;
Langer, M .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2005, 29 (01) :1-5
[6]   Multisegment and halfscan reconstruction of 16-slice computed tomography for assessment of regional and global left ventricular myocardial function [J].
Dewey, M ;
Müller, M ;
Teige, F ;
Schnapauff, D ;
Schink, T ;
Hamm, B ;
Lembcke, A .
INVESTIGATIVE RADIOLOGY, 2006, 41 (04) :400-409
[7]   Quantitative parameters of image quality in 64-slice computed tomography angiography of the coronary arteries [J].
Ferencik, M ;
Nomura, CH ;
Maurovich-Horvat, P ;
Hoffmann, U ;
Pena, AJ ;
Cury, RC ;
Abbara, S ;
Nieman, K ;
Fatima, U ;
Achenbach, S ;
Brady, TJ .
EUROPEAN JOURNAL OF RADIOLOGY, 2006, 57 (03) :373-379
[8]   Heart rate adaptive optimization of spatial and temporal resolution for electrocardiogram-gated multislice spiral CT of the heart [J].
Flohr, T ;
Ohnesorge, B .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2001, 25 (06) :907-923
[9]   First performance evaluation of a dual-source CT (DSCT) system [J].
Flohr, TG ;
McCollough, CH ;
Bruder, H ;
Petersilka, M ;
Gruber, K ;
Süss, C ;
Grasruck, M ;
Stierstorfer, K ;
Krauss, B ;
Raupach, R ;
Primak, AN ;
Küttner, A ;
Achenbach, S ;
Becker, C ;
Kopp, A ;
Ohnesorge, BM .
EUROPEAN RADIOLOGY, 2006, 16 (02) :256-268
[10]   Guidelines on the management of stable angina pectoris: executive summary [J].
Fox, Kim ;
Angeles Alonso Garcia, Maria ;
Ardissino, Diego ;
Buszman, Pawel ;
Katowice ;
Camici, Paolo G. ;
Crea, Filippo ;
Daly, Caroline ;
De Backer, Guy ;
Ghent ;
Hjemdahl, Paul ;
Lopez-Sendon, Jose ;
Marco, Jean ;
Morais, Joao ;
Leiria ;
Pepper, John ;
Sechtem, Udo ;
Simoons, Maarten ;
Thygesen, Kristian ;
Priori, Silvia G. ;
Blanc, Jean-Jacques ;
Budaj, Andrzej ;
Camm, John ;
Dean, Veronica ;
Deckers, Jaap ;
Dickstei, Kenneth ;
Lekakis, John ;
McGregor, Keith ;
Metra, Marco ;
Morais, Joao ;
Osterspey, Ady ;
Tamargo, Juan ;
Zamorano, Jose L. ;
Andreotti, Felicita ;
Becher, Harald ;
Dietz, Rainer ;
Fraser, Alan ;
Hernandez Antolin, Rosa Ana ;
Huber, Kurt ;
Kremastinos, Dimitris T. ;
Maseri, Attilio ;
Nesser, Hans-Joachim ;
Pasierski, Tomasz ;
Sigwart, Ulrich ;
Tubaro, Marco ;
Weis, Michael .
EUROPEAN HEART JOURNAL, 2006, 27 (11) :1341-1381