Repeatability and reproducibility of retention data and band profiles on reversed-phase liquid chromatography columns V.: Results obtained with Vydac 218TP C18 columns

被引:37
作者
Kele, M
Guiochon, G
机构
[1] Univ Tennessee, Dept Chem, Knoxville, TN 37996 USA
[2] Oak Ridge Natl Lab, Div Chem & Analyt Sci, Oak Ridge, TN 37831 USA
关键词
retention data; band profiles; stationary phases; LC; Vydac 218TP C-18 columns;
D O I
10.1016/S0021-9673(00)01042-6
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The general goal of this work is to investigate the precision of chromatographic data and to determine which properties of chromatographic columns influence this factor. Chromatographic data were acquired under five different sets of experimental conditions for 30 neutral, acidic and basic test compounds on columns packed with Vydac 218TP C-18, a polymeric, wide-pore silica-based stationary phase. Five columns packed with samples from the same batch of this packing material were used to measure the column-to-column reproducibility and six columns packed with material from six different batches to measure the batch-to-batch reproducibility. The parameters studied were the retention time, the retention and separation factors, the hydrophobic and the steric selectivities, the column efficiency, and the tailing factor. (C) 2001 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:89 / 112
页数:24
相关论文
共 31 条
[1]   Characterization of wide-pore reversed-phase columns for biopolymer separations .1. Single-parametric evaluation [J].
Bede, A ;
Rippel, G ;
Szepesy, L ;
Claessens, HA .
JOURNAL OF CHROMATOGRAPHY A, 1996, 728 (1-2) :179-188
[2]   SYNTHESES OF STATIONARY PHASES FOR REVERSED-PHASE LC USING SILANIZATION AND POLYMER COATING [J].
BIENVOGELSANG, U ;
DEEGE, A ;
FIGGE, H ;
KOHLER, J ;
SCHOMBURG, G .
CHROMATOGRAPHIA, 1984, 19 :170-179
[3]  
DOVE PM, 1994, SILICA PHYSICAL BEHA
[4]   Chromatographic determination of metallic impurities in reversed-phase HPLC columns [J].
Engelhardt, H ;
Lobert, T .
ANALYTICAL CHEMISTRY, 1999, 71 (09) :1885-1892
[5]   COMPARISON AND CHARACTERIZATION OF REVERSED PHASES [J].
ENGELHARDT, H ;
JUNGHEIM, M .
CHROMATOGRAPHIA, 1990, 29 (1-2) :59-68
[6]   POLYMER ENCAPSULATED STATIONARY PHASES - ADVANTAGES, PROPERTIES AND SELECTIVITIES [J].
ENGELHARDT, H ;
LOW, H ;
EBERHARDT, W ;
MAUSS, M .
CHROMATOGRAPHIA, 1989, 27 (11-12) :535-543
[7]   USE OF METHYL SPACERS IN A MIXED HORIZONTALLY POLYMERIZED STATIONARY-PHASE [J].
FAIRBANK, RWP ;
XIANG, Y ;
WIRTH, MJ .
ANALYTICAL CHEMISTRY, 1995, 67 (21) :3879-3885
[8]   SI-29 AND C-13 NMR CHARACTERIZATION OF MIXED HORIZONTALLY POLYMERIZED MONOLAYERS ON SILICA-GEL [J].
FATUNMBI, HO ;
BRUCH, MD ;
WIRTH, MJ .
ANALYTICAL CHEMISTRY, 1993, 65 (15) :2048-2054
[9]  
FLEINGER A, 2000, IN PRESS J CHROMAT A
[10]   Repeatability and reproducibility of retention data and band profiles on reversed-phase liquid chromatography columns III.: Results obtained with Kromasil C18 columns [J].
Kele, M ;
Guiochon, G .
JOURNAL OF CHROMATOGRAPHY A, 1999, 855 (02) :423-453