Solid-phase FT-Raman determination of caffeine in energy drinks

被引:57
作者
Armenta, S [1 ]
Garrigues, S [1 ]
de la Guardia, M [1 ]
机构
[1] Univ Valencia, Dept Analyt Chem, E-46100 Burjassot, Valencia, Spain
关键词
caffeine; energy drink; FT-Raman; solid phase;
D O I
10.1016/j.aca.2005.05.032
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A solid-phase vibrational spectrometry-based methodology (solid-phase Fourier transform-Raman spectrometry, SP-FT-Raman) has been developed for caffeine determination in commercial energy drink samples. The Raman spectra of caffeine, fixed on a C18 solid phase packed into a glass tube of 5 mm i.d., was obtained directly between 3500 and 70 cm(-1). In order to quantify caffeine, Raman intensity between 573 and 542 cm(-1) corrected using a baseline defined between 580 and 540 cm(-1) was used. A repeatability of 3%, as relative standard deviation of five analysis of a 200 mg l(-1) concentration, and a limit of detection of 18 mg l(-1) were obtained. The SP-FT-Raman procedure provides a sampling frequency of 13.3 h(-1), higher than that of liquid chromatography (LC), which was 7.0 h(-1). The use of FT-Raman reduces the reagent consumption and waste generation, also minimizing the sample handling. Results obtained by the developed procedure were statistically comparable with those found by a reference LC method. (c) 2005 Elsevier B.V. All rights reserved.
引用
收藏
页码:197 / 203
页数:7
相关论文
共 23 条