Application of Raman spectroscopy for on-line monitoring of low dose blend uniformity

被引:59
作者
Hausman, DS
Cambron, RT
Sakr, A
机构
[1] Procter & Gamble Co, Norwich, NY 13815 USA
[2] Univ Cincinnati, Coll Pharm, Ind Pharm Grad Program, Cincinnati, OH 45267 USA
关键词
Raman; on-line spectroscopy; blend uniformity; low dose; content uniformity;
D O I
10.1016/j.ijpharm.2005.04.011
中图分类号
R9 [药学];
学科分类号
1007 ;
摘要
On-line Raman spectroscopy was used to evaluate the effect of blending time on low dose, 1%, blend uniformity of azimilide dihydrochloride. An 8qt blender was used for the experiments and instrumented with a Raman probe through the I-bar port. The blender was slowed to 6.75 rpm to better illustrate the blending process (normal speed is 25 rpm). Uniformity was reached after 20 min of blending at 6.75 rpm (135 revolutions or 5.4 min at 25 rpm). On-line Raman analysis of blend uniformity provided more benefits than traditional thief sampling and off-line analysis. On-line Raman spectroscopy enabled generating data rich blend profiles, due to the ability to collect a large number of samples during the blending process (sampling every 20 s). In addition, the Raman blend profile was rapidly generated, compared to the lengthy time to complete a blend profile with thief sampling and off-line analysis. The on-line Raman blend uniformity results were also significantly correlated (p-value < 0.05) to the HPLC uniformity results of thief samples. (c) 2005 Published by Elsevier B.V.
引用
收藏
页码:80 / 90
页数:11
相关论文
共 22 条
[1]   CONTROL ANALYSIS OF A PHARMACEUTICAL PREPARATION BY NEAR-INFRARED REFLECTANCE SPECTROSCOPY - A COMPARATIVE-STUDY OF A SPINNING MODULE AND FIBER OPTIC PROBE [J].
BLANCO, M ;
COELLO, J ;
ITURRIAGA, H ;
MASPOCH, S ;
DELAPEZUELA, C ;
RUSSO, E .
ANALYTICA CHIMICA ACTA, 1994, 298 (02) :183-191
[2]   Quantitative characterization of mixing of dry powders in V-blenders [J].
Brone, D ;
Alexander, A ;
Muzzio, FJ .
AICHE JOURNAL, 1998, 44 (02) :271-278
[3]   Enhanced mixing in double-cone blenders [J].
Brone, D ;
Muzzio, FJ .
POWDER TECHNOLOGY, 2000, 110 (03) :179-189
[4]  
CARSTENSEN J, 1993, TECHNOMIC LANCASTER, P15
[5]  
Chang R.-K., 2004, PHARM TECH MAR, V78
[6]   The Mahalanobis distance [J].
De Maesschalck, R ;
Jouan-Rimbaud, D ;
Massart, DL .
CHEMOMETRICS AND INTELLIGENT LABORATORY SYSTEMS, 2000, 50 (01) :1-18
[7]   Near-infrared spectroscopy and imaging for the monitoring of powder blend homogeneity [J].
El-Hagrasy, AS ;
Morris, HR ;
D'Amico, F ;
Lodder, RA ;
Drennen, JK .
JOURNAL OF PHARMACEUTICAL SCIENCES, 2001, 90 (09) :1298-1307
[8]  
*FDA, 2000, GUID IND POWD BLENDS
[9]  
HUSSAIN A, 2002, PAT SUBC ADV COMM PH
[10]  
HWANG R, 2004, AM PHARM REV, V7, P101