Cross-calibration of eight-polar bioelectrical impedance analysis versus dual-energy X-ray absorptiometry for the assessment of total and appendicular body composition in healthy subjects aged 21-82 years

被引:292
作者
Malavolti, M
Mussi, C
Poli, M
Fantuzzi, AL
Salvioli, G
Battistini, N
Bedogni, G
机构
[1] Univ Modena & Reggio Emilia, Fac Med & Surg, Dept Hyg Microbiol & Biostat, Human Nutr Chair, I-41100 Modena, Italy
[2] Univ Modena & Reggio Emilia, Geriatr Chair, Modena, Italy
[3] USL, Food Sci & Nutr Unit, Modena, Italy
关键词
D O I
10.1080/0301446031000095211
中图分类号
Q98 [人类学];
学科分类号
030303 ;
摘要
Aim: To calibrate eight-polar bioelectrical impedance analysis (BIA) against dual-energy X-ray absorptiometry (DXA) for the assessment of total and appendicular body composition in healthy adults. Research design: A cross-sectional study was carried out. Subjects: Sixty-eight females and 42 males aged 21-82 years participated in the study. Methods: Whole-body fat-free mass (FFM) and appendicular lean tissue mass (LTM) were measured by DXA; resistance (R) of arms, trunk and legs was measured by eight-polar BIA at frequencies of 5, 50, 250 and 500 kHz; whole-body resistance was calculated as the sum R of arms, trunk and legs. Results: The resistance index (RI), i.e. the height(2)/resistance ratio, was the best predictor of FFM and appendicular LTM. As compared with weight (Wt), RI at 500 kHz explained 35% more variance of FFM (R-adj(2) =0.92 vs 0.57), 45% more variance of LTMarm (R-adj(2) = 0.93 vs 0.48) and 36% more variance of LTleg (R-adj(2) = 0.86 vs 0.50) (p < 0.001 for all). The contribution of age to the unexplained variance of FFM and appendicular LTM was nil or negligible and the RI x sex interactions were either not significant or not important on practical grounds. The percent root mean square error of the estimate was 6% for FFM and 8% for LTMarm and LTMleg. Conclusion: Eight-polar BIA offers accurate estimates of total and appendicular body composition. The attractive hypothesis that eight-polar BIA is influenced minimally by age and sex should be tested on larger samples including younger individuals.
引用
收藏
页码:380 / 391
页数:12
相关论文
共 40 条
[1]  
[Anonymous], AM J PHYSIOL
[2]   The prediction of total body water and extracellular water from bioelectric impedance in obese children [J].
Bedogni, G ;
Bollea, MR ;
Severi, S ;
Trunfio, O ;
Manzieri, AM ;
Battistini, N .
EUROPEAN JOURNAL OF CLINICAL NUTRITION, 1997, 51 (03) :129-133
[3]   Accuracy of an eight-point tactile-electrode impedance method in the assessment of total body water [J].
Bedogni, G ;
Malavolti, M ;
Severi, S ;
Poli, M ;
Mussi, C ;
Fantuzzi, AL ;
Battistini, N .
EUROPEAN JOURNAL OF CLINICAL NUTRITION, 2002, 56 (11) :1143-1148
[4]  
BEDOGNI G, IN PRESS EUROPEAN J
[5]   MEASUREMENT OF BODY-COMPOSITION IN THE ELDERLY - DUAL-ENERGY X-RAY ABSORPTIOMETRY, UNDERWATER WEIGHING, BIOELECTRICAL-IMPEDANCE ANALYSIS, AND ANTHROPOMETRY [J].
BRODOWICZ, GR ;
MANSFIELD, RA ;
MCCLUNG, MR ;
ALTHOFF, SA .
GERONTOLOGY, 1994, 40 (06) :332-339
[6]   Assessment of body composition in elderly:: Accuracy of bioelectrical impedance analysis [J].
Bussolotto, M ;
Ceccon, A ;
Sergi, G ;
Giantin, V ;
Benincà, P ;
Enzi, G .
GERONTOLOGY, 1999, 45 (01) :39-43
[7]  
CHUMLEA WC, 1995, J GERONTOL A-BIOL, V50, P45
[8]   Validity of total and segmental impedance measurements for prediction of body composition across ethnic population groups [J].
Deurenberg, P ;
Deurenberg-Yap, M ;
Schouten, FJM .
EUROPEAN JOURNAL OF CLINICAL NUTRITION, 2002, 56 (03) :214-220
[9]   ASSESSMENT OF BODY-COMPOSITION BY BIOELECTRICAL IMPEDANCE IN A POPULATION AGED GREATER-THAN-60-Y [J].
DEURENBERG, P ;
VANDERKOOIJ, K ;
EVERS, P ;
HULSHOF, T .
AMERICAN JOURNAL OF CLINICAL NUTRITION, 1990, 51 (01) :3-6
[10]   New equations for estimating body cell mass from bioimpedance parallel models in healthy older Germans [J].
Dittmar, M ;
Reber, H .
AMERICAN JOURNAL OF PHYSIOLOGY-ENDOCRINOLOGY AND METABOLISM, 2001, 281 (05) :E1005-E1014