IABP-related vascular complications: Who is responsible - the patient, the surgeon or the sheath? Part I: Sheath-related complications

被引:18
作者
Kocogullari, Cevdet Ugur [1 ]
Emmiler, Mustafa [1 ]
Ayva, Ercument [1 ]
Cekirdekct, Ahmet [1 ]
机构
[1] Afyon Kocatepe Univ, Fac Med, Dept Cardiovasc Surg, Afyon, Turkey
关键词
catheter sheaths; intra-aortic balloon pump; percutaneous Seldinger; sheathless; vascular complications;
D O I
10.1007/s12325-008-0027-9
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 [基础医学];
摘要
Introduction: An intra-aortic balloon pump (IABP) is frequently used as a mechanical support during the treatment of low cardiac output syndrome. The use of a sheath during IABP treatment can be associated with ischaemic complications in the extremities. The aim of our study was to determine the isolated role of sheaths in the development of vascular complications following IABP catheterisation. Methods: We evaluated a total of 81 patients who had received an IABP for haemodynamic stability, between January 2003 and October 2007, following cardiovascular surgery. To eliminate surgical-related variables, only patients with IABPs and size 8.5 or 9 French catheters (inserted by experienced cardiovascular surgeons using the percutaneous Seldinger technique) were included in the study. To eliminate patient-related variables, patients with a known risk factor for either thrombosis, embolus or peripheral arterial disease were excluded from the study. Results: The patients were divided into two groups: Group 1 (sheathed catheter group; n=42), Group 2 (sheathless catheter group; n=39). Ischaemic complications were observed in a total of five patients. Four of these patients were from Group 1 and one was from Group 2. Conclusion: Sheathed catheterisation was found to increase the risk of vascular complications in patients who underwent IABP application. Therefore sheathless catheterisation is recommended for IABP application.
引用
收藏
页码:225 / 230
页数:6
相关论文
共 10 条
[1]
Vascular complications of the intraaortic balloon pump in patients undergoing open heart operations: 15-year experience [J].
Arafa, OE ;
Pedersen, TH ;
Svennevig, JL ;
Fosse, E ;
Geiran, OR .
ANNALS OF THORACIC SURGERY, 1999, 67 (03) :645-651
[2]
Complications related to intra-aortic balloon pump in cardiac surgery: A decade later [J].
Elahi, MM ;
Chetty, GK ;
Kirke, R ;
Azeem, T ;
Hartshorne, R ;
Spyt, TJ .
EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 2005, 29 (06) :591-594
[3]
In which patients should sheathless IABP be used?<ΤΒ> An analysis of vascular complications in 1211 cases [J].
Erdogan, Hasan Basri ;
Goksedef, Deniz ;
Erentug, Vedat ;
Polat, Adil ;
Bozbuga, Nilguen ;
Mansuroglu, Denyan ;
Guler, Mustafa ;
Akinci, Esat ;
Yakut, Cevat .
JOURNAL OF CARDIAC SURGERY, 2006, 21 (04) :342-346
[4]
Erentug Vedat, 2004, Ulus Travma Acil Cerrahi Derg, V10, P28
[5]
INITIAL CLINICAL EXPERIENCE WITH INTRAAORTIC BALLOON PUMPING IN CARDIOGENIC SHOCK [J].
KANTROWITZ, A ;
TJONNELAND, S ;
FREED, PS ;
PHILLIPS, SJ ;
BUTNER, AN ;
SHERMAN, JL .
JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1968, 203 (02) :113-+
[6]
Kovacs Endre, 2006, Magy Seb, V59, P105
[7]
Vascular complications of intra-aortic balloon insertion in patients undergoing coronary reavscularization: analysis of 911 cases [J].
Meharwal, ZS ;
Trehan, N .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2002, 21 (04) :741-747
[8]
MOULOPOULOS SD, 1962, T AM SOC ART INT ORG, V8, P85
[9]
Timing of intra-aortic balloon pump support and 1-year survival [J].
Ramnarine, IR ;
Grayson, AD ;
Dihmis, WC ;
Mediratta, NK ;
Fabri, BM ;
Chalmers, JAC .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2005, 27 (05) :887-892
[10]
Effects of intra-aortic balloon counterpulsation on coronary pressure in patients with stencotic coronary arteries [J].
Yoshitani, Hidetoshi ;
Akasaka, Takashi ;
Kaji, Shuichiro ;
Kawamoto, Takahiro ;
Kume, Teruyoshi ;
Neishi, Yoji ;
Koyanta, Yuji ;
Yoshida, Kiyoshi .