Pressure-driven reverse-phase liquid chromatography separations in ordered nonporous pillar array columns

被引:152
作者
De Malsche, Wim
Eghbali, Harned
Clicq, David
Vangelooven, Joris
Gardeniers, Han
Desmet, Gert [1 ]
机构
[1] Vrije Univ Brussel, Dept Chem Engn, Brussels, Belgium
[2] Res Inst, MESA, Res Program Mesofluid, Enschede, Netherlands
关键词
D O I
10.1021/ac070352p
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Building upon the micromachined column idea proposed by the group of Regnier in 1998, we report on the first high-resolution reversed-phase separations in micromachined pillar array columns under pressure-driven LC conditions. A three component mixture could be separated in 3 s using arrays of nonporous silicon pillars with a diameter of approximately 4.3 mu m and an external porosity of 55%. Under slightly retained component conditions (retention factor k' = 0.65- 1.2), plate heights of about H = 4 pm were obtained at a mobile phase velocity around u = 0.5 mm/s. In reduced terms, such plate heights are as low as h(min) = 1. Also, since the flow resistance of the column is much smaller than in a packed column (mainly because of the higher external porosity of the pillar array), the separation impedance of the array was as small as E = 150, i.e., of the same order as the best currently existing monolithic columns. At pH = 3, yielding very low retention factors (k' = 0. 13 and 0.23), plate heights as low as H = 2 mu m were realized, yielding a separation of the three component mixture with an efficiency of N = 4000-5000 plates over a column length of I cm. At higher retention factors, significantly larger plate heights were obtained. More experimental work is needed to investigate this more in depth. The study is completed with a discussion of the performance limits of the pillar array column concept in the frame of the current state-of-the-art in microfabrication precision.
引用
收藏
页码:5915 / 5926
页数:12
相关论文
共 67 条
[11]   Experimental study on band dispersion in channels structured with micropillars [J].
De Pra, M. ;
Kok, W. Th. ;
Gardeniers, J. G. E. ;
Desmet, G. ;
Eeltink, S. ;
van Nieuwkasteele, J. W. ;
Schoenmakers, P. J. .
ANALYTICAL CHEMISTRY, 2006, 78 (18) :6519-6525
[12]   Influence of the pillar shape on the band broadening and the separation impedance of perfectly ordered 2-D porous chromatographic media [J].
De Smet, J ;
Gzil, P ;
Vervoort, N ;
Verelst, H ;
Baron, GV ;
Desmet, G .
ANALYTICAL CHEMISTRY, 2004, 76 (13) :3716-3726
[13]   Geometry-independent plate height representation methods for the direct comparison of the kinetic performance of LC supports with a different size or morphology [J].
Desmet, G ;
Clicq, D ;
Gzil, P .
ANALYTICAL CHEMISTRY, 2005, 77 (13) :4058-4070
[14]   Simultaneous optimization of the analysis time and the concentration detectability in open-tubular liquid chromatography [J].
Desmet, G ;
Baron, GV .
JOURNAL OF CHROMATOGRAPHY A, 2000, 867 (1-2) :23-43
[15]   Future of high pressure liquid chromatography: Do we need porosity or do we need pressure? [J].
Desmet, Gert ;
Cabooter, Deirdre ;
Gzil, Piotr ;
Verelst, Harry ;
Mangelings, Debby ;
Vander Heyden, Yvan ;
Clicq, David .
JOURNAL OF CHROMATOGRAPHY A, 2006, 1130 (01) :158-166
[16]   Dispersion reduction in pressure driven flow through microetched channels [J].
Dutta, D ;
Leighton, DT .
ANALYTICAL CHEMISTRY, 2001, 73 (03) :504-513
[17]   Computer simulations of electrokinetic injection techniques in microfluidic devices [J].
Ermakov, SV ;
Jacobson, SC ;
Ramsey, JM .
ANALYTICAL CHEMISTRY, 2000, 72 (15) :3512-3517
[18]   Electrokinetic injection techniques in microfluidic chips [J].
Fu, LM ;
Yang, RJ ;
Lee, GB ;
Liu, HH .
ANALYTICAL CHEMISTRY, 2002, 74 (19) :5084-5091
[19]   COMPARISON OF THEORETICAL LIMIT OF SEPARATING SPEED IN GAS AND LIQUID CHROMATOGRAPHY [J].
GIDDINGS, JC .
ANALYTICAL CHEMISTRY, 1965, 37 (01) :60-&