Comparison of hologic QDR-1000/W and 4500W DXA scanners in 13-to 18-year olds

被引:38
作者
Litaker, MS [1 ]
Barbeau, P
Humphries, MC
Gutin, B
机构
[1] Med Coll Georgia, Off Biostat & Bioinformat, Augusta, GA 30912 USA
[2] Georgia Prevent Inst, Off Biostat & Bioinformat, Augusta, GA 30912 USA
[3] Med Coll Georgia, Dept Pediat, Augusta, GA 30912 USA
[4] Med Coll Georgia, Dept Physiol, Augusta, GA 30912 USA
来源
OBESITY RESEARCH | 2003年 / 11卷 / 12期
关键词
reliability; children; fat mass; bone mineral content;
D O I
10.1038/oby.2003.206
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: Body composition measurements made using Hologic QDR-1000/W pencil-beam and QDR-4500W fan-beam scanners (Bedford, MA) were compared in a sample of 13- to 18-year-old white and black youth (n = 219). Research Methods and Procedures: Total fat (FAT), fat-free soft tissue (FFST), bone mineral content (BMC), bone mineral density (BMD), and percent body fat (%BF) were compared between repeated measurements using the QDR-4500 and between the two scanners using mixed model ANOVA. Intraclass correlation coefficients and Bland-Altman limits of agreement were used to evaluate inter- and intrascanner reliability. Results: Intraclass correlation coefficients for repeated measurements using the QDR-4500 ranged from 0.997 to 0.999 for FAT, %BF, FFST, and BMC and 0.987 for BMD. Mean measurements made using the two scanners differed significantly for FAT, %BF, BMC, and BMD (p < 0.0001), and scan by sex interactions were significant (all p < 0.0005). There were no significant differences in mean measurements between repeat scans using the QDR-4500 (all p > 0.19). Limits of agreement for measurements of FAT, FFST, and %BF made using the two scanners were approximately three times as wide as those for two measurements using the QDR-4500. For lower values of FAT and %BF, the QDR-4500 gave higher measurements than the QDR-1000, whereas at higher values, this relationship was reversed. The QDR-1000 tended to give higher BMC measurements, with larger differences for higher values. Discussion: Using different models of DXA scanners within a study may reduce precision of body composition measurement. This issue needs to be considered in the design of longitudinal studies.
引用
收藏
页码:1545 / 1552
页数:8
相关论文
共 19 条
[1]   Fat mass measured by DXA varies with scan velocity [J].
Black, E ;
Petersen, L ;
Kreutzer, M ;
Toubro, S ;
Sorensen, TIA ;
Pedersen, O ;
Astrup, A .
OBESITY RESEARCH, 2002, 10 (02) :69-77
[2]   Technical principles of dual energy x-ray absorptiometry [J].
Blake, GM ;
Fogelman, I .
SEMINARS IN NUCLEAR MEDICINE, 1997, 27 (03) :210-228
[3]   DUAL X-RAY ABSORPTIOMETRY - A COMPARISON BETWEEN FAN-BEAM AND PENCIL BEAM SCANS [J].
BLAKE, GM ;
PARKER, JC ;
BUXTON, FMA ;
FOGELMAN, I .
BRITISH JOURNAL OF RADIOLOGY, 1993, 66 (790) :902-906
[4]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[5]   Cross-calibration of a fan-beam X-ray densitometer with a pencil-beam system [J].
Bouyoucef, SE ;
Cullum, ID ;
Jell, P .
BRITISH JOURNAL OF RADIOLOGY, 1996, 69 (822) :522-531
[6]   SWITCHING FROM DXA PENCIL-BEAM TO FAN-BEAM .2. STUDIES IN-VIVO [J].
EIKEN, P ;
KOLTHOFF, N ;
BARENHOLDT, O ;
HERMANSEN, F ;
NIELSEN, SP .
BONE, 1994, 15 (06) :671-676
[7]   Bone mineral and body composition measurements: Cross-calibration of pencil-beam and fan-beam dual-energy X-ray absorptiometers [J].
Ellis, KJ ;
Shypailo, RJ .
JOURNAL OF BONE AND MINERAL RESEARCH, 1998, 13 (10) :1613-1618
[8]   CROSS-CALIBRATION OF DXA EQUIPMENT - UPGRADING FROM A HOLOGIC QDR 1000/W TO A QDR-2000 [J].
FAULKNER, KG ;
GLUER, CC ;
ESTILO, M ;
GENANT, HK .
CALCIFIED TISSUE INTERNATIONAL, 1993, 52 (02) :79-84
[9]   Correcting the magnification error of fan beam densitometers [J].
Griffiths, MR ;
Noakes, KA ;
Pocock, NA .
JOURNAL OF BONE AND MINERAL RESEARCH, 1997, 12 (01) :119-123
[10]   Body-composition measurement in 9-11-y-old children by dual-energy X-ray absorptiometry, skinfold-thickness measurements, and bioimpedance analysis [J].
Gutin, B ;
Litaker, M ;
Islam, S ;
Manos, T ;
Smith, C ;
Treiber, F .
AMERICAN JOURNAL OF CLINICAL NUTRITION, 1996, 63 (03) :287-292