Noninvasive measurement of heartbeat, respiration, snoring and body movements of a subject in bed via a pneumatic method

被引:169
作者
Watanabe, K [1 ]
Watanabe, T [1 ]
Watanabe, H [1 ]
Ando, H [1 ]
Ishikawa, T [1 ]
Kobayashi, K [1 ]
机构
[1] Hosei Univ, Fac Engn, Syst Control Engn Dept, Koganei, Tokyo 1848584, Japan
关键词
body movements; health monitoring; heartbeat; noninvasive measurement; pressure sensor; respiration; snoring; vital sign;
D O I
10.1109/TBME.2005.857637
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We have developed a noninvasive pneumatics-based system by which to measure heartbeat, respiration, snoring, and body movements of a subject in bed. A thin, air-sealed cushion is placed under the bed mattress of the subject and the small movements attributable to human automatic vital functions are measured as changes in pressure using a pressure sensor having an almost flat frequency response from 0.1 to 5 kHz and a sensitivity of 56 mV/Pa. Using the newly developed system, heartbeat, respiration, apnea, snoring and body movements are clearly measured. In addition, the optimal signal-to-noise (S/N) ratio by which to evaluate the reliability of the heart rate measurement is presented. Heart rates were measured for four different body postures, 13 different subjects, four different bed mattresses, and three different sensor positions. For these measurements, the S/N ratios ranged from 15.9 to 23.5 dB, and so were determined to be reliable.
引用
收藏
页码:2100 / 2107
页数:8
相关论文
共 10 条
[1]   STATIC CHARGE SENSITIVE BED - NEW METHOD FOR RECORDING BODY MOVEMENTS DURING SLEEP [J].
ALIHANKA, J ;
VAAHTORANTA, K .
ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1979, 46 (06) :731-734
[2]   AUTOMATIC SLEEP WAKE IDENTIFICATION FROM WRIST ACTIVITY [J].
COLE, RJ ;
KRIPKE, DF ;
GRUEN, W ;
MULLANEY, DJ ;
GILLIN, JC .
SLEEP, 1992, 15 (05) :461-469
[3]   MONITORING OF ELECTROCARDIOGRAMS IN BED WITHOUT UTILIZING BODY-SURFACE ELECTRODES [J].
ISHIJIMA, M .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1993, 40 (06) :593-594
[4]  
ISHIJIMA M, 1997, MED BIOMED ENG COMPU, V34
[5]  
Rechtschaffen A., 1968, BRAIN INFORM SERVICE
[6]   Artifact-resistant power-efficient design of finger-ring plethysmographic sensors [J].
Rhee, S ;
Yang, BH ;
Asada, HH .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2001, 48 (07) :795-805
[7]   Unconstrained and non-invasive measurement of heart-beat and respiration periods using a phonocardiographic sensor [J].
Tanaka, S ;
Matsumoto, Y ;
Wakimoto, K .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2002, 40 (02) :246-252
[8]  
Watanabe H., 1999, Transactions of the Society of Instrument and Control Engineers, V35, P1012
[9]  
WATANABE H, 2000, T SICE, V36, P894
[10]  
WATANABE K, 2000, IMEKO WORLD C, V17