Reliability and exactness of MRI-based volumetry: A phantom study

被引:60
作者
Luft, AR [1 ]
Skalej, M [1 ]
Welte, D [1 ]
Kolb, R [1 ]
Klose, U [1 ]
机构
[1] UNIV TUBINGEN, DEPT NEURORADIOL, D-72076 TUBINGEN, GERMANY
关键词
image processing; MRI; volumetry;
D O I
10.1002/jmri.1880060421
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This study investigated the influence slice thickness, section orientation, contrast, shape, and sequence type on the exactness of MRI-based volumetry, Ni-doped agarose gel phantoms (4 to 46 ml) were scanned with a T1-weighted three-dimensional Fourier transform (FT) fast low-angle shot (FLASH) and a multiecho two-dimensional FT-Turbo spin-echo (SE) sequence, After segmentation with a three-dimensional legion-growing algorithm, the geometric volume was measured considering the partial volume effect, The variability coefficient (Pearson) was .7%, The volumetric error increased with slice thickness, depending on the size and form of the object, Cross sections resulted in smaller error than longitudinal sections (finger-shaped phantoms, nonisotropic image data), Three-dimensional FT was superior to two-dimensional FT imaging. Results of slice thickness and section orientation experiments can be explained by the partial volume effect Higher errors in two-dimensional FT imaging were caused by object movements between two interleaved acquisitions, The study shows a considerable influence of the imaging parameters on the exactness, which depends on size and form of the structure of interest.
引用
收藏
页码:700 / 704
页数:5
相关论文
共 18 条
[1]  
ASHTARI M, 1991, AM J NEURORADIOL, V12, P941
[2]   RELIABILITY OF IN-VIVO VOLUME MEASURES OF HIPPOCAMPUS AND OTHER BRAIN STRUCTURES USING MRI [J].
BARTZOKIS, G ;
MINTZ, J ;
MARX, P ;
OSBORN, D ;
GUTKIND, D ;
CHIANG, F ;
PHELAN, CK ;
MARDER, SR .
MAGNETIC RESONANCE IMAGING, 1993, 11 (07) :993-1006
[3]   MEASUREMENT OF WHOLE TEMPORAL-LOBE AND HIPPOCAMPUS FOR MR VOLUMETRY - NORMATIVE DATA [J].
BHATIA, S ;
BOOKHEIMER, SY ;
GAILLARD, WD ;
THEODORE, WH .
NEUROLOGY, 1993, 43 (10) :2006-2010
[4]  
BLATTER DD, 1995, AM J NEURORADIOL, V16, P241
[5]   GRAPHICAL ANALYSIS OF MR FEATURE SPACE FOR MEASUREMENT OF CSF, GRAY-MATTER, AND WHITE-MATTER VOLUMES [J].
BONAR, DC ;
SCHAPER, KA ;
ANDERSON, JR ;
ROTTENBERG, DA ;
STROTHER, SC .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1993, 17 (03) :461-470
[6]  
BUCCIOLINI M, 1989, ACTA RADIOL, V30, P667
[7]   NICKEL-DOPED AGAROSE-GEL PHANTOMS IN MR IMAGING [J].
CHRISTOFFERSSON, JO ;
OLSSON, LE ;
SJOBERG, S .
ACTA RADIOLOGICA, 1991, 32 (05) :426-431
[8]  
EHRICKE HH, 1995, C P COMP MED IM GRAP
[9]   MR-BASED HIPPOCAMPAL VOLUMETRY IN THE DIAGNOSIS OF ALZHEIMERS-DISEASE [J].
JACK, CR ;
PETERSEN, RC ;
OBRIEN, PC ;
TANGALOS, EG .
NEUROLOGY, 1992, 42 (01) :183-188
[10]   ANALYSIS OF BRAIN AND CEREBROSPINAL-FLUID VOLUMES WITH MR IMAGING .1. METHODS, RELIABILITY, AND VALIDATION [J].
KOHN, MI ;
TANNA, NK ;
HERMAN, GT ;
RESNICK, SM ;
MOZLEY, PD ;
GUR, RE ;
ALAVI, A ;
ZIMMERMAN, RA ;
GUR, RC .
RADIOLOGY, 1991, 178 (01) :115-122