Artefacts in spiral-CT images and their relation to pitch and subject morphology

被引:32
作者
Wilting, JE [1 ]
Timmer, J [1 ]
机构
[1] Philips Med Syst Nederland BV, Dept Comp Tomog, NL-5680 DA Best, Netherlands
关键词
spiral CT; pitch; artefact; phantoms; cadaver;
D O I
10.1007/s003300050673
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This qualitative study is intended to create awareness of artefacts that are associated with spiral-CT imaging. A simple description of spiral-CT reconstruction is used to er;plain how these artefacts depend on the pitch and subject morphology, and shows when these artefacts are likely to impair the diagnostic value of the acquired images. We scanned a cone and rod phantom with pitch 2, and used the acquired images to demonstrate how spiral data acquisition and interpolation leads to artefacts in the reconstructed images. We then demonstrated the effects of various pitches in scans of a human cadaver. whereas the slice thickness was kept constant. Some patient studies are presented in order to show the possible clinical consequences. Spiral acquisition may cause geometric distortions and apparent inhomogeneity of homogeneous structures. We were able to link these artefacts to the way in the acquisitions were done, and the reconstructions were performed. We have shown how these artefacts can be anticipated in clinical studies. When areas of low contrast, surrounded by hypo- or hyperdense structures, are scanned with a large pitch and viewed with a narrow window, spiral artefacts may influence the diagnostic quality of the images. These effects should he considered when choosing the pitch.
引用
收藏
页码:316 / 322
页数:7
相关论文
共 13 条
[1]  
*1 GERM CONS C SPI, 1996, EUR RADIOL, P2
[2]   HELICAL CT - PRINCIPLES AND TECHNICAL CONSIDERATIONS [J].
BRINK, JA ;
HEIKEN, JP ;
WANG, G ;
MCENERY, KW ;
SCHLUETER, FJ ;
VANNIER, MW .
RADIOGRAPHICS, 1994, 14 (04) :887-893
[3]   COMPUTED-TOMOGRAPHY SCANNING WITH SIMULTANEOUS PATIENT TRANSLATION [J].
CRAWFORD, CR ;
KING, KF .
MEDICAL PHYSICS, 1990, 17 (06) :967-982
[4]  
FISHMAN EK, 1995, SPIRAL CT PRINCIPLES
[5]   TECHNICAL FOUNDATIONS OF SPIRAL CT [J].
KALENDER, WA .
SEMINARS IN ULTRASOUND CT AND MRI, 1994, 15 (02) :81-89
[6]   A COMPARISON OF CONVENTIONAL AND SPIRAL CT - AN EXPERIMENTAL-STUDY ON THE DETECTION OF SPHERICAL LESIONS [J].
KALENDER, WA ;
POLACIN, A ;
SUSS, C .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1994, 18 (02) :167-176
[7]  
MACOVSKI A, 1981, RADIOLOGY SKULL BRAI, P3877
[8]   MEASUREMENT OF SLICE SENSITIVITY PROFILES IN SPIRAL CT [J].
POLACIN, A ;
KALENDER, WA ;
BRINK, J ;
VANNIER, MA .
MEDICAL PHYSICS, 1994, 21 (01) :133-140
[9]   INITIAL EXPERIENCE WITH VOLUME CT SCANNING [J].
RIGAUTS, H ;
MARCHAL, G ;
BAERT, AL ;
HUPKE, R .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1990, 14 (04) :675-682
[10]  
VIECO P, 1995, HELICAL SPIRAL CT PR, P27