Elution-extrusion countercurrent chromatography. Use of the liquid nature of the stationary phase to extend the hydrophobicity window

被引:159
作者
Berthod, A [1 ]
Ruiz-Angel, MJ
Carda-Broch, S
机构
[1] Univ Lyon 1, Sci Analyt Lab, F-69622 Villeurbanne, France
[2] Univ Valencia, Dept Quim Analit, Burjassot 46100, Spain
[3] Univ Jaime 1, Area Quim Analit, Castellon de La Plana 12006, Spain
关键词
D O I
10.1021/ac030208d
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Countercurrent chromatography (CCC) is a liquid chromatography technique with a liquid stationary phase. Taking advantage of the liquid nature of the stationary phase, it is possible to perform unique operations not possible in classical liquid chromatography with a solid stationary phase. It is easy to avoid any solute-irreversible absorption in the CCC column. If the retention volumes of solutes become too high, the dual mode will be used. The roles of the phases are reversed. The stationary phase becomes the mobile phase, and the CCC column is started again. The solutes elute rapidly in what was previously the stationary phase. The theoretical basis of the dualmode method is recalled. The dual-mode method is a discontinuous method. The separation should be stopped when the phase switch is performed. The elution-extrusion procedure is another way to avoid any irreversible adsorption of solutes in the column. The method uses the fact that the liquid volumes occupied by the solutes highly retained inside the column can be orders of magnitude lower than the mobile-phase volume that would be needed to elute them. The elution-extrusion method also has two steps: the first step is a regular CCC chromatogram. Next, the stationary phase containing the partially separated hydrophobic solutes is extruded out of the column in a continuous way using the liquid stationary phase. The theory of the process is developed and compared to the dual-mode theory. Alkylbenzene homologues are experimentally used as model compounds with the heptane/methanol/water biphasic liquid system to establish the theoretical treatment and compare the performance of two types, hydrodynamic and hydrostatic, of CCC columns. It is shown that the method can dramatically boost the separation power of the CCC technique. An apparent efficiency higher than 20 000 plates was obtained for extruded octylbenzene and a 160-mL hydrodynamic CCC column with less than 500 plates when conventionally used.
引用
收藏
页码:5886 / 5894
页数:9
相关论文
共 28 条
[11]  
DEFOLTER J, CHROMATOGRAPHIC EXPT
[12]   COUNTERCURRENT CHROMATOGRAPHY [J].
FOUCAULT, AP .
ANALYTICAL CHEMISTRY, 1991, 63 (10) :A569-&
[13]   Counter-current chromatography: instrumentation, solvent selection and some recent applications to natural product purification [J].
Foucault, AP ;
Chevolot, L .
JOURNAL OF CHROMATOGRAPHY A, 1998, 808 (1-2) :3-22
[14]   EXTENDED OCTANOL-WATER PARTITION-COEFFICIENT DETERMINATION BY DUAL-MODE CENTRIFUGAL PARTITION CHROMATOGRAPHY [J].
GLUCK, SJ ;
MARTIN, EJ .
JOURNAL OF LIQUID CHROMATOGRAPHY, 1990, 13 (18) :3559-3570
[15]  
Gunawardana G, 1999, CHROMATOGR SCI SERIE, V82, P249
[16]   Isolation of a bronchodilator flavonoid from the Thai medicinal plant Clerodendrum petasites [J].
Hazekamp, A ;
Verpoorte, R ;
Panthong, A .
JOURNAL OF ETHNOPHARMACOLOGY, 2001, 78 (01) :45-49
[17]   Application of centrifugal partition chromatography in a general separation and dereplication procedure for plant extracts [J].
Ingkaninan, K ;
Hazekamp, A ;
Hoek, AC ;
Balconi, S ;
Verpoorte, R .
JOURNAL OF LIQUID CHROMATOGRAPHY & RELATED TECHNOLOGIES, 2000, 23 (14) :2195-2208
[18]   HIGH-SPEED PREPARATIVE COUNTERCURRENT CHROMATOGRAPHY WITH A COIL PLANET CENTRIFUGE [J].
ITO, Y ;
SANDLIN, J ;
BOWERS, WG .
JOURNAL OF CHROMATOGRAPHY, 1982, 244 (02) :247-258
[19]  
Ito Y., 1996, CHEM ANAL, V132
[20]  
MAILLARD M, 1996, CHEM ANAL SERIES MON, V132, P179