Fluidics cube for biosensor miniaturization

被引:15
作者
Dodson, JM [1 ]
Feldstein, MJ [1 ]
Leatzow, DM [1 ]
Flack, LK [1 ]
Golden, JP [1 ]
Ligler, FS [1 ]
机构
[1] Naval Res Lab, Ctr Biomol Sci & Engn, Washington, DC 20375 USA
关键词
D O I
10.1021/ac010168i
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
To create a small, portable, fully automated biosensor, a compact means of fluid handling is required. We designed, manufactured, and tested a "fluidics cube" for such a purpose. This cube, made of thermoplastic, contains reservoirs and channels for liquid samples and reagents and operates without the use of any internal valves or meters; it is a passive fluid circuit that relies on pressure relief vents to control fluid movement. We demonstrate the ability of pressure relief vents to control fluid movement and show how to simply manufacture or modify the cube. Combined with the planar array biosensor developed at the Naval Research Laboratory, it brings us one step closer to realizing our goal of a handheld biosensor capable of analyzing multiple samples for multiple analytes.
引用
收藏
页码:3776 / 3780
页数:5
相关论文
共 10 条
  • [1] Biotechnology at low Reynolds numbers
    Brody, JP
    Yager, P
    Goldstein, RE
    Austin, RH
    [J]. BIOPHYSICAL JOURNAL, 1996, 71 (06) : 3430 - 3441
  • [2] BRODY JP, 1996, SOL STAT SENS ACT WO, P105
  • [3] BRODY JP, 1998, Patent No. 5726404
  • [4] Array Biosensor: Optical and Fluidics Systems
    Feldstein, Mark J.
    Golden, Joel P.
    Rowe, Chris A.
    MacCraith, Brian D.
    Ligler, Frances S.
    [J]. BIOMEDICAL MICRODEVICES, 1999, 1 (02) : 139 - 153
  • [5] LEATZOW DM, IN PRESS BIOSENS BIO
  • [6] MANZ A, 1995, MESA MG, P5
  • [7] MANZ A, 1993, ADV CHROMATOGR, V33, P1
  • [8] An array immunosensor for simultaneous detection of clinical analytes
    Rowe, CA
    Scruggs, SB
    Feldstein, MJ
    Golden, JP
    Ligler, FS
    [J]. ANALYTICAL CHEMISTRY, 1999, 71 (02) : 433 - 439
  • [9] Verlee D, 1996, SOL STAT SENS ACT WO, p9~14
  • [10] Detection of multiple toxic agents using a planar array immunosensor
    Wadkins, RM
    Golden, JP
    Pritsiolas, LM
    Ligler, FS
    [J]. BIOSENSORS & BIOELECTRONICS, 1998, 13 (3-4) : 407 - 415