Dose reduction in spiral CT:: Detection of pulmonary nodules with and without anatomic adaptation of tube current.

被引:14
作者
Greess, H
Baum, U
Wolf, H
Lell, M
Nömayr, A
Schmidt, B
Kalender, WA
Bautz, W
机构
[1] Univ Erlangen Nurnberg, Inst Diagnost Radiol, D-91054 Erlangen, Germany
[2] Univ Erlangen Nurnberg, Inst Med Phys, Erlangen, Germany
来源
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN | 2001年 / 173卷 / 05期
关键词
dose reduction; spiral CT; intrapulmonary nodule;
D O I
10.1055/s-2001-13343
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To investigate the potential of projection angle-dependent anatomical adaptation of tube current to reduce dose in spiral CT examinations of the thorax without loss in detectability of pulmonary nodules and image quality. Patients and Methods: Dose can be reduced for non-circular patient cross-sections without an increase in noise if the tube current is reduced at those angular positions where the patient diameter and, consequently, attenuation are small. The examinations were dose with a SOMATOM Plus 4 (Siemens AG). CT projection data were analysed to determine the optimum tube current for each projection angle in real time. We compared image quality, done. and detection rates of pulmonary nodules for thoracic spiral CT examinations with and without online anatomically adapted tube current control in a group of 38 patients. Three radiologists counted all intrapulmonary nodules in consensus separated in three different groups (< 5mm, 5-10 mm, > 10 mm). Image quality was evaluated in a scale from 1-3 (1=very good, 2=good, 3=worse). Results: On average, the dose was reduced by 21% (15-34%). With a constant tube current, 704 pulmonary nodules were detected, with tube current modulation 707 pulmonary modules. For three patients we saw more pulmonary nodules (five) with tube current modulation, for two patients we saw less pulmonary nodules (two). In a direct comparison, the missed intrapulmonary nodules were also detected in the respective method. Thus, the detection rate of intrapulmonary nodules was uninfluenced. In general, no deterioration of image quality was observed. Conclusion: On average, 21% dose reduction was achieved by an anatomically adapted tube current modulation in spiral CT examinations of the thorax without a loss in detectability of pulmonary nodules and image quality. This method is an important contribution to dose reduction in spiral CT.
引用
收藏
页码:466 / 470
页数:5
相关论文
共 17 条
  • [1] Low-dose computed tomography of the thorax. Experimental and clinical investigations
    Diederich, S
    Lenzen, H
    Puskas, Z
    Koch, AT
    Yelbuz, TM
    Eameri, M
    Roos, N
    Peters, PE
    [J]. RADIOLOGE, 1996, 36 (06): : 475 - 482
  • [2] Pulmonary nodules: detection with low-dose vs conventional-dose spiral CT
    Gartenschlager, M
    Schweden, F
    Gast, K
    Westermeier, T
    Kauczor, HU
    von Zitzewitz, H
    Thelen, M
    [J]. EUROPEAN RADIOLOGY, 1998, 8 (04) : 609 - 614
  • [3] Giacomuzzi S M, 1996, Aktuelle Radiol, V6, P110
  • [4] Dose reduction in CT by anatomically adapted tube current modulation. 1. Simulation studies
    Gies, M
    Kalender, WA
    Wolf, H
    Suess, C
    Madsen, MT
    [J]. MEDICAL PHYSICS, 1999, 26 (11) : 2235 - 2247
  • [5] Dose reduction in computed tomography by attenuation-based on-line modulation of tube current: evaluation of six anatomical regions
    Greess, H
    Wolf, H
    Baum, U
    Lell, M
    Pirkl, M
    Kalender, W
    Bautz, WA
    [J]. EUROPEAN RADIOLOGY, 2000, 10 (02) : 391 - 394
  • [6] Greess H, 1999, ROFO-FORTSCHR RONTG, V170, P246
  • [7] KALENDER W, 2000, COMPUTERTOMOGRAPHIE, P119
  • [8] Dose reduction in CT by anatomically adapted tube current modulation. II. Phantom measurements
    Kalender, WA
    Wolf, H
    Suess, C
    [J]. MEDICAL PHYSICS, 1999, 26 (11) : 2248 - 2253
  • [9] A PC program for estimating organ dose and effective dose values in computed tomography
    Kalender, WA
    Schmidt, B
    Zankl, M
    Schmidt, M
    [J]. EUROPEAN RADIOLOGY, 1999, 9 (03) : 555 - 562
  • [10] Dose reduction in CT by on-line tube current control: principles and validation on phantoms and cadavers
    Kalender, WA
    Wolf, H
    Suess, C
    Gies, M
    Greess, H
    Bautz, WA
    [J]. EUROPEAN RADIOLOGY, 1999, 9 (02) : 323 - 328