Small-volume radial flow cell for all-solid-state ion-selective electrodes

被引:30
作者
Vázquez, M [1 ]
Bobacka, J [1 ]
Ivaska, A [1 ]
Lewenstam, A [1 ]
机构
[1] Abo Akad Univ, Proc Chem Grp, Ctr Proc Analyt Chem & Sensor Technol ProSens, FIN-20500 Turku, Finland
关键词
flow cell; ion-selective electrode; conducting polymer; poly(3,4-ethylenedioxythiophene); Baytron P; clinical analysis;
D O I
10.1016/S0039-9140(03)00408-9
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A flow cell with a radial distribution of four all-solid-state ion selective electrodes (ISEs), or alternatively three ISEs and one reference electrode, was designed and optimized for mass production. The radial distribution of the electrodes reduces the cell volume and is expected to minimize cross-contamination between different electrodes. Two different cell prototypes were developed and tested for all-solid-state K+-ISEs based on a solvent polymeric ion-selective membrane (ISM) and a conducting polymer, poly(3,4-ethylenedioxythiophene), as solid internal contact. In the first prototype, PEDOT was electropolymerized from an aqueous solution of the monomer and the doping ion salt, sodium polystyrenesulfonate (NaPSS). The second prototype employed an aqueous dispersion of PEDOT(PSS) that is commercially available (Baytron P. Bayer AG). Compared to electrochemical synthesis, solution casting of the polymer dispersion was found to be a more advantageous method to deposit the conducting polymer layer aiming at mass production. The resulting prototypes of the flow cell had a small volume (ca. 17-37 mul), which makes them suitable for application in clinical analysis. (C) 2003 Elsevier B.V. All rights reserved.
引用
收藏
页码:57 / 63
页数:7
相关论文
共 24 条
  • [1] Alonso L, 1999, ELECTROANAL, V11, P161
  • [2] Ion sensors: current limits and new trends
    Bakker, E
    Diamond, D
    Lewenstam, A
    Pretsch, E
    [J]. ANALYTICA CHIMICA ACTA, 1999, 393 (1-3) : 11 - 18
  • [3] Carrier-based ion-selective electrodes and bulk optodes. 1. General characteristics
    Bakker, E
    Buhlmann, P
    Pretsch, E
    [J]. CHEMICAL REVIEWS, 1997, 97 (08) : 3083 - 3132
  • [4] BERGQUIST C, 2000, Patent No. 6123820
  • [5] Potential stability of all-solid-state ion-selective electrodes using conducting polymers as ion-to-electron transducers
    Bobacka, J
    [J]. ANALYTICAL CHEMISTRY, 1999, 71 (21) : 4932 - 4937
  • [6] Potentiometric ion sensors based on conducting polymers
    Bobacka, J
    Ivaska, A
    Lewenstam, A
    [J]. ELECTROANALYSIS, 2003, 15 (5-6) : 366 - 374
  • [7] MICROFABRICATED SENSOR ARRAYS SENSITIVE TO PH AND K+ FOR IONIC DISTRIBUTION MEASUREMENTS IN THE BEATING HEART
    COSOFRET, VV
    ERDOSY, M
    JOHNSON, TA
    BUCK, RP
    ASH, RB
    NEUMAN, MR
    [J]. ANALYTICAL CHEMISTRY, 1995, 67 (10) : 1647 - 1653
  • [8] CHEMICAL SENSORS, BIOSENSORS AND THICK-FILM TECHNOLOGY
    GALANVIDAL, CA
    MUNOZ, J
    DOMINGUEZ, C
    ALEGRET, S
    [J]. TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 1995, 14 (05) : 225 - 231
  • [9] Self-assembly of a conducting polymer nanostructure by physical crosslinking:: applications to conducting blends and modified electrodes
    Ghosh, S
    Inganäs, O
    [J]. SYNTHETIC METALS, 1999, 101 (1-3) : 413 - 416
  • [10] Ghosh S, 1999, ADV MATER, V11, P1214, DOI 10.1002/(SICI)1521-4095(199910)11:14<1214::AID-ADMA1214>3.0.CO