Organic acids as the modifier to determine Zn, Cd, Tl and Pb in soil by slurry sampling electrothermal vaporization inductively-coupled plasma mass spectrometry

被引:45
作者
Lu, HH [1 ]
Jiang, SJ [1 ]
机构
[1] Natl Sun Yat Sen Univ, Dept Chem, Kaohsiung 80424, Taiwan
关键词
ICP-MS; ultrasonic slurry sampling; electrothermal vaporization; soil; Zn; Cd; Tl; Pb;
D O I
10.1016/S0003-2670(00)01288-5
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Ultrasonic slurry sampling electrothermal vaporization inductively-coupled plasma mass spectrometry (USS-ETV-ICP-MS) has been applied to the determination of Zn, Cd, T1 and Pb in several soil samples. A mixture of 1% m/v EDTA and 1.5% m/v of ascorbic acid was used as the modifier to enhance the ion signals. The influences of instrument operating conditions and slurry preparation on the ion signals were reported. In this study, a relatively low vaporization temperature was used which separated the analyte from the major matrix components and improved the ion signals significantly. With ETV sample introduction device, spectroscopic and nonspectroscopic interferences were alleviated significantly. Since the sensitivities of Zn, Cd, T1 and Pb in various soil slurries and aqueous solution were different, standard addition method and isotope dilution method were used for the determination of Zn, Cd, T1 and Pb in these soil samples. This method has been applied to the determination of Zn, Cd, T1 and Pb in NIST SRM 2711 Montana soil reference material and several soil samples collected from Kaohsiung area. The analysis results were agreed with the certified values. The precision between sample replicates was better than 9% with USS-ETV-ICP-MS method. Detection limits estimated from standard addition curves were about 260-620, 3-5, 0.3-1 and 9-20 ng g(-1) for Zn, Cd, T1 and Pb, respectively, in different samples. (C) 2001 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:247 / 255
页数:9
相关论文
共 34 条
[1]   DETERMINATION OF TRACE-METALS IN REFERENCE WATER STANDARDS BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY WITH ONLINE PRECONCENTRATION [J].
BEAUCHEMIN, D ;
BERMAN, SS .
ANALYTICAL CHEMISTRY, 1989, 61 (17) :1857-1862
[2]   DETERMINATION OF TRACE-METALS IN A RIVER WATER REFERENCE MATERIAL BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
BEAUCHEMIN, D ;
MCLAREN, JW ;
MYKYTIUK, AP ;
BERMAN, SS .
ANALYTICAL CHEMISTRY, 1987, 59 (05) :778-783
[3]   SOLID SAMPLING IN ELECTROTHERMAL ATOMIC-ABSORPTION SPECTROMETRY USING COMMERCIAL ATOMIZERS - A REVIEW [J].
BENDICHO, C ;
DELOOSVOLLEBREGT, MTC .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1991, 6 (05) :353-374
[4]   DETERMINATION OF URANIUM IN ENVIRONMENTAL-SAMPLES USING INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
BOOMER, DW ;
POWELL, MJ .
ANALYTICAL CHEMISTRY, 1987, 59 (23) :2810-2813
[5]   ELECTROTHERMAL ISOTOPE-DILUTION INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY METHOD FOR THE DETERMINATION OF SUB-NG ML(-1) LEVELS OF LEAD IN HUMAN PLASMA [J].
BOWINS, RJ ;
MCNUTT, RH .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1994, 9 (11) :1233-1236
[6]   Determination of copper, cadmium and lead in biological samples by electrothermal vaporization isotope dilution inductively coupled plasma mass spectrometry [J].
Chang, CC ;
Jiang, SJ .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1997, 12 (01) :75-80
[7]   APPLICATION OF CARBON CUP ATOMIZATION TECHNIQUE IN WATER ANALYSIS BY ATOMIC-ABSORPTION SPECTROSCOPY [J].
DOLINSEK, F ;
STUPAR, J .
ANALYST, 1973, 98 (1173) :841-+
[8]   THE ROLE OF CHEMICAL MODIFIERS IN ANALYTE TRANSPORT LOSS INTERFERENCES WITH ELECTROTHERMAL VAPORIZATION ICP-MASS SPECTROMETRY [J].
EDIGER, RD ;
BERES, SA .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1992, 47 (07) :907-922
[9]   Effect of oxygen ashing on analyte transport efficiency using ETV-ICP-MS [J].
Fonseca, RW ;
Miller-Ihli, NJ ;
Sparks, C ;
Holcombe, JA ;
Shaver, B .
APPLIED SPECTROSCOPY, 1997, 51 (12) :1800-1806
[10]   DIRECT ANALYSIS OF SOLIDS BY ULTRASONIC SLURRY ELECTROTHERMAL VAPORIZATION INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY [J].
GREGOIRE, DC ;
MILLERIHLI, NJ ;
STURGEON, RE .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1994, 9 (05) :605-610