Prospects of atomic resolution imaging with an aberration-corrected STEM

被引:47
作者
Ishizuka, K [1 ]
机构
[1] HREM Res, Higashimatsuyama, Saitama 3550055, Japan
来源
JOURNAL OF ELECTRON MICROSCOPY | 2001年 / 50卷 / 04期
关键词
STEM; image simulation; high-angle annular dark-field imaging; Cs-correction; dynamical scattering;
D O I
10.1093/jmicro/50.4.291
中图分类号
TH742 [显微镜];
学科分类号
摘要
We investigated high-resolution scanning transmission electron microscope (STEM) images obtained from a microscope equipped with a spherical aberration corrector. The probe size (full-width at half-maximum) is reduced to 0.76 Angstrom at 200 kV by assuming the fifth-order spherical aberration coefficient C-5 = 100 mm. For the simulation we have used the recently developed scheme for a STEM image simulation based on the Fast Fourier Transform (FFT) multislice algorithm. The peak-to-background (P/B) ratio of the high-angle annular dark-field (HAADF) image is significantly improved at a thin specimen region. Although the P/B ratio becomes worse at a thicker region, the resolution is kept high even at such a region. An almost true HAADF signal will be obtained even from a weak-scattering phosphorous column in InP [001] when the background is subtracted. in the bright-field image the coherent character of elastic scattering is suppressed by averaging over a large convergence angle, making the specimen effectively self-luminous. The claim that HAADF imaging is relatively insensitive to a defocus as well as a specimen thickness is valid only qualitatively, and a detailed image simulation will be required for a quantitative analysis as in the case of the conventional transmission electron microscope. It was noted that the delta function approximation for the object function may not be applicable for a very fine probe, and that the achievable resolution of the HAADF imaging will be limited by the widths of the high-angle thermal diffuse scattering potential.
引用
收藏
页码:291 / 305
页数:15
相关论文
共 35 条
[21]   Incoherent imaging using dynamically scattered coherent electrons [J].
Nellist, PD ;
Pennycook, SJ .
ULTRAMICROSCOPY, 1999, 78 (1-4) :111-124
[22]   Subangstrom resolution by underfocused incoherent transmission electron microscopy [J].
Nellist, PD ;
Pennycook, SJ .
PHYSICAL REVIEW LETTERS, 1998, 81 (19) :4156-4159
[23]   CHEMICALLY SENSITIVE STRUCTURE-IMAGING WITH A SCANNING-TRANSMISSION ELECTRON-MICROSCOPE [J].
PENNYCOOK, SJ ;
BOATNER, LA .
NATURE, 1988, 336 (6199) :565-567
[24]   HIGH-RESOLUTION INCOHERENT IMAGING OF CRYSTALS [J].
PENNYCOOK, SJ ;
JESSON, DE .
PHYSICAL REVIEW LETTERS, 1990, 64 (08) :938-941
[25]   HIGH-RESOLUTION Z-CONTRAST IMAGING OF CRYSTALS [J].
PENNYCOOK, SJ ;
JESSON, DE .
ULTRAMICROSCOPY, 1991, 37 (1-4) :14-38
[26]  
Pennycook SJ, 2000, MICROSC MICROANAL, V6, P343
[27]   DEBYE-WALLER FACTORS OF ZINC-BLEND-STRUCTURE MATERIALS - A LATTICE DYNAMICAL COMPARISON [J].
REID, JS .
ACTA CRYSTALLOGRAPHICA SECTION A, 1983, 39 (JAN) :1-13
[28]   THE THEORETICAL RESOLUTION LIMIT OF THE ELECTRON MICROSCOPE [J].
SCHERZER, O .
JOURNAL OF APPLIED PHYSICS, 1949, 20 (01) :20-29
[29]  
SCHERZER O, 1970, BERICH BUNSEN GESELL, V74, P1154
[30]  
SPENCE JCH, 1978, OPTIK, V50, P129