Energy modulated electron therapy using a few leaf electron collimator in combination with IMRT and 3D-CRT: Monte Carlo-based planning and dosimetric evaluation

被引:37
作者
Al-Yahya, K
Schwartz, M
Shenouda, G
Verhaegen, F
Freeman, C
Seuntjens, J
机构
[1] McGill Univ, Montreal Gen Hosp, Med Phys Unit, Montreal, PQ H3G 1A4, Canada
[2] McGill Univ, Montreal Gen Hosp, Ctr Hlth, Dept Radiat Oncol, Montreal, PQ H3G 1A4, Canada
基金
加拿大健康研究院;
关键词
Monte Carlo dosimetry; energy modulated electron therapy; intensity modulated radiation therapy; whole body dose;
D O I
10.1118/1.2011089
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Energy modulated electron therapy (EMET) based on Monte Carlo dose calculation is a promising technique that enhances the treatment planning and delivery of superficially located tumors. This study investigated the application of EMET using a novel few-leaf electron collimator (FLEC) in head and neck and breast sites in comparison with three-dimensional conventional radiation therapy (3D-CRT) and intensity modulated radiation therapy (IMRT) techniques. Treatment planning was performed for two parotid cases and one breast case. Four plans were compared for each case: 3D-CRT, IMRT, 3D-CRT in conjunction with EMET (EMET-CRT), and IMRT in conjunction with EMET (EMET-IMRT), all of which were performed and calculated with Monte Carlo techniques. For all patients, dose volume histograms (DVHs) were obtained for all organs of interest and the DVHs were used as a means of comparing the plans. Homogeneity and conformity of dose distributions were calculated, as well as a sparing index that compares the effect of the low isodose lines. In addition, the whole-body dose equivalent (WBDE) was estimated for each plan. Adding EMET delivered with the FLEC to 3D-CRT improves sparing of normal tissues. For the two head and neck cases, the mean dose to the contralateral parotid and brain stem was reduced relative to IMRT by 43% and 84%, and by 57% and 71%, respectively. Improved normal tissue sparing was quantified as an increase in sparing index of 47% and 30% for the head and neck and the breast cases, respectively. Adding EMET to either 3D-CRT or IMRT results in preservation of target conformity and dose homogeneity. When adding EMET to the treatment plan, the WBDE was reduced by between 6% and 19% for 3D-CRT and by between 21% and 33% for IMRT, while WBDE for EMET-CRT was reduced by up to 72% when compared with IMRT. FLEC offers a practical means of delivering modulated electron therapy. Although adding EMET delivered using the FLEC results in perturbation of target conformity when compared to IMRT, it significantly improves normal tissue sparing while offering enhanced target conformity to the 3D-CRT planning. The addition of EMET systematically leads to a reduction in WBDE especially when compared with IMRT. (C) 2005 American Association of Physicists in Medicine.
引用
收藏
页码:2976 / 2986
页数:11
相关论文
共 29 条
[1]   Monte Carlo based modulated electron beam treatment planning using a few-leaf electron collimator - feasibility study [J].
Al-Yahya, K ;
Hristov, D ;
Verhaegen, F ;
Seuntjens, J .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (05) :847-857
[2]   A conformal index (COIN) to evaluate implant quality and dose specification in brachytherapy [J].
Baltas, D ;
Kolotas, C ;
Geramani, K ;
Mould, RF ;
Ioannidis, G ;
Kekchidi, M ;
Zamboglou, N .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1998, 40 (02) :515-524
[3]   IMRT head and neck treatment planning with a commercially available Monte Carlo based planning system [J].
Boudreau, C ;
Heath, E ;
Seuntjens, J ;
Ballivy, O ;
Parker, W .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (05) :879-890
[4]   Intensity-modulated radiotherapy: Current status and issues of interest [J].
Boyer, AL ;
Butler, EB ;
DiPetrillo, TA ;
Engler, MJ ;
Fraass, B ;
Grant, W ;
Ling, CC ;
Low, DA ;
Mackie, TR ;
Mohan, R ;
Purdy, JA ;
Roach, M ;
Rosenman, JG ;
Verhey, LJ ;
Wong, JW ;
Cumberlin, RL ;
Stone, H ;
Palta, JR .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2001, 51 (04) :880-914
[5]   The role of intensity-modulated radiotherapy in the treatment of parotid tumors [J].
Bragg, CM ;
Conway, J ;
Robinson, MH .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2002, 52 (03) :729-738
[6]   Comparison of measured and Monte Carlo calculated dose distributions in inhomogeneous phantoms in clinical electron beams [J].
Doucet, R ;
Olivares, M ;
DeBlois, F ;
Podgorsak, EB ;
Kawrakow, I ;
Seuntjens, J .
PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (15) :2339-2354
[7]   Estimates of whole-body dose equivalent produced by beam intensity modulated conformal therapy [J].
Followill, D ;
Geis, P ;
Boyer, A .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 38 (03) :667-672
[8]   Intensity-modulated radiation therapy: The inverse, the converse, and the perverse [J].
Glatstein, E .
SEMINARS IN RADIATION ONCOLOGY, 2002, 12 (03) :272-281
[9]   Radiation-induced second cancers: The impact of 3D-CRT and IMRT [J].
Hall, EJ ;
Wuu, CS .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2003, 56 (01) :83-88
[10]   Development and validation of a BEAMnrc component module for accurate Monte Carlo modelling of the Varian dynamic Millennium multileaf collimator [J].
Heath, E ;
Seuntjens, J .
PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (24) :4045-4063