Precision of Lunar Achilles plus bone quality measurements: time dependency and multiple machine use in field studies

被引:24
作者
Economos, C. D.
Sacheck, J. M.
Wacker, W.
Shea, K.
Naumova, E. N.
机构
[1] Tufts Univ, Gerald J & Dorothy R Friedman Sch Nutr Sci & Poli, Boston, MA 02111 USA
[2] GE Healthcare, Madison, WI 53717 USA
[3] Tufts Univ, Sch Med, Div Publ Hlth & Family Med, Boston, MA 02111 USA
关键词
D O I
10.1259/bjr/33589854
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 [临床医学]; 100207 [影像医学与核医学]; 1009 [特种医学];
摘要
Qualitative ultrasound (QUS) is a portable, safe and relatively inexpensive technique used to obtain information on bone mineral quality in adults and children. QUS measures bone stiffness index (SI) through the incorporation of speed of sound (SOS) and broadband ultrasound attenuation (BUA). QUS technology may prove to be extremely useful in field research where more than one machine is used over different periods of time. 13 adults (27.6 +/- 4.6 years old) were recruited to determine the internal stability of two Lunar Achilles+ QUS machines (Lunar1, Lunar2), as well as the repeatability in bone stiffness measures between the two machines over time. Triplicate measurements of the calcaneus were taken within the same day (n=258) and at 1 week (n=120), 6 months (n=54) and 1 year (n=18) apart to determine the time-dependent repeatability. Using paired t-tests and separate mixed effects models, there were no differences reported in SI, SOS or BUA values within one machine, or between two machines over these short- and long-term time-frames. These results indicate that QUS machines are internally consistent and different machines may be used over time to provide reliable measurements of changes in bone quality.
引用
收藏
页码:919 / 925
页数:7
相关论文
共 29 条
[1]
Effect of age, weight and lifestyle factors on calcaneal quantitative ultrasound in premenopausal women: The ESOPO study [J].
Adami, S ;
Giannini, S ;
Giorgino, R ;
Isaia, GC ;
Maggi, S ;
Sinigaglia, L ;
Filipponi, P ;
Crepaldi, G .
CALCIFIED TISSUE INTERNATIONAL, 2004, 74 (04) :317-321
[2]
ULTRASOUND ATTENUATION OF THE CALCANEUS - A SENSITIVE AND SPECIFIC DISCRIMINATOR OF OSTEOPENIA IN POSTMENOPAUSAL WOMEN [J].
AGREN, M ;
KARELLAS, A ;
LEAHEY, D ;
MARKS, S ;
BARAN, D .
CALCIFIED TISSUE INTERNATIONAL, 1991, 48 (04) :240-244
[3]
Diagnostic agreement of two calcaneal ultrasound devices: the Sahara bone sonometer and the Achilles [J].
Alenfeld, FE ;
Engelke, K ;
Schmidt, D ;
Brezger, M ;
Diessel, E ;
Felsenberg, D .
BRITISH JOURNAL OF RADIOLOGY, 2002, 75 (899) :895-902
[4]
COMPARING METHODS OF MEASUREMENT - WHY PLOTTING DIFFERENCE AGAINST STANDARD METHOD IS MISLEADING [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1995, 346 (8982) :1085-1087
[5]
Comparison of calcaneus ultrasound and dual X-ray absorptiometry in children at risk of osteopenia [J].
Fielding, KT ;
Nix, DA ;
Bachrach, LK .
JOURNAL OF CLINICAL DENSITOMETRY, 2003, 6 (01) :7-15
[6]
BROAD-BAND ULTRASOUND ATTENUATION IN THE DIAGNOSIS OF OSTEOPOROSIS - CORRELATION WITH OSTEODENSITOMETRY AND FRACTURE [J].
FUNKE, M ;
KOPKA, L ;
VOSSHENRICH, R ;
FISCHER, U ;
UEBERSCHAER, A ;
OESTMANN, JW ;
GRABBE, E .
RADIOLOGY, 1995, 194 (01) :77-81
[7]
Precision and discriminatory ability of calcaneal bone assessment technologies [J].
Greenspan, SL ;
Bouxsein, ML ;
Melton, ME ;
Kolodny, AH ;
Clair, JH ;
Delucca, PT ;
Stek, M ;
Faulkner, KG ;
Orwoll, ES .
JOURNAL OF BONE AND MINERAL RESEARCH, 1997, 12 (08) :1303-1313
[8]
HANS D, 1995, J BONE MINER RES, V10, pS169
[9]
Longitudinal quality control methodology for the quantitative ultrasound Achilles plus in clinical trial settings [J].
Hans, D ;
Wacker, W ;
Genton, L ;
Paris, E ;
Le Floch, C ;
Slosman, D .
OSTEOPOROSIS INTERNATIONAL, 2002, 13 (10) :788-795
[10]
ULTRASOUND BONE-MEASUREMENT IN PEDIATRIC SUBJECTS [J].
JAWORSKI, M ;
LEBIEDOWSKI, M ;
LORENC, RS ;
TREMPE, J .