Practical performance evaluation of a 10k x 10k CCD for electron cryo-microscopy

被引:11
作者
Bammes, Benjamin E. [1 ,2 ]
Rochat, Ryan H. [1 ,2 ]
Jakana, Joanita [1 ]
Chiu, Wah [1 ,2 ]
机构
[1] Baylor Coll Med, Verna & Marrs McLean Dept Biochem & Mol Biol, Natl Ctr Macromol Imaging, Houston, TX 77030 USA
[2] Baylor Coll Med, Grad Program Struct & Computat Biol & Mol Biophys, Houston, TX 77030 USA
基金
美国国家卫生研究院;
关键词
Cryo-EM; Electron cryo-microscopy; 10k x 10k CCD; Charge-coupled devices; Electron detection; SINGLE PARTICLES; RESOLUTION; MICROSCOPY; IMAGES; CAMERA; ACQUISITION; DETECTORS; CRYSTALLOGRAPHY; RECONSTRUCTION; OPTIMIZATION;
D O I
10.1016/j.jsb.2011.05.012
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
Electron cryo-microscopy (cryo-EM) images are commonly collected using either charge-coupled devices (CCD) or photographic film. Both film and the current generation of 16 megapixel (4k x 4k) CCD cameras have yielded high-resolution structures. Yet, despite the many advantages of CCD cameras, more than two times as many structures of biological macromolecules have been published in recent years using photographic film. The continued preference to film, especially for subnanometer-resolution structures, may be partially influenced by the finer sampling and larger effective specimen imaging area offered by film. Large format digital cameras may finally allow them to overtake film as the preferred detector for cryo-EM. We have evaluated a 111-megapixel (10k x 10k) CCD camera with a 9 gm pixel size. The spectral signal-to-noise ratios of low dose images of carbon film indicate that this detector is capable of providing signal up to at least 2/5 Nyquist frequency potentially retrievable for 3D reconstructions of biological specimens, resulting in more than double the effective specimen imaging area of existing 4k x 4k CCD cameras. We verified our estimates using frozen-hydrated epsilon 15 bacteriophage as a biological test specimen with previously determined structure, yielding a similar to 7 angstrom resolution single particle reconstruction from only 80 CCD frames. Finally, we explored the limits of current CCD technology by comparing the performance of this detector to various CCD cameras used for recording data yielding subnanometer resolution cryo-EM structures submitted to the electron microscopy data bank (http://www.emdatabank.org/). (C) 2011 Elsevier Inc. All rights reserved.
引用
收藏
页码:384 / 393
页数:10
相关论文
共 34 条
[1]   Radiation damage effects at four specimen temperatures from 4 to 100 K [J].
Bammes, Benjamin E. ;
Jakana, Joanita ;
Schmid, Michael F. ;
Chiu, Wah .
JOURNAL OF STRUCTURAL BIOLOGY, 2010, 169 (03) :331-341
[2]   Assessing the capabilities of a 4kx4k CCD camera for electron cryo-microscopy at 300kV [J].
Booth, Christopher R. ;
Jakana, Joanita ;
Chiu, Wah .
JOURNAL OF STRUCTURAL BIOLOGY, 2006, 156 (03) :556-563
[3]   A 9 Å single particle reconstruction from CCD captured images on a 200 kV electron cryomicroscope [J].
Booth, CR ;
Jiang, W ;
Baker, ML ;
Zhou, ZH ;
Ludtke, SJ ;
Chiu, W .
JOURNAL OF STRUCTURAL BIOLOGY, 2004, 147 (02) :116-127
[4]   APPLICATIONS OF A SLOW-SCAN CCD CAMERA IN PROTEIN ELECTRON CRYSTALLOGRAPHY [J].
BRINK, J ;
CHIU, W .
JOURNAL OF STRUCTURAL BIOLOGY, 1994, 113 (01) :23-34
[5]   Leginon: An automated system for acquisition of images from vitreous ice specimens [J].
Carragher, B ;
Kisseberth, N ;
Kriegman, D ;
Milligan, RA ;
Potter, CS ;
Pulokas, J ;
Reilein, A .
JOURNAL OF STRUCTURAL BIOLOGY, 2000, 132 (01) :33-45
[6]   Achievable resolution from images of biological specimens acquired from a 4k x 4k CCD camera in a 300-kV electron cryomicroscope [J].
Chen, Dong-Hua ;
Jakana, Joanita ;
Liu, Xiangan ;
Schmid, Michael F. ;
Chiu, Wah .
JOURNAL OF STRUCTURAL BIOLOGY, 2008, 163 (01) :45-52
[7]   Performance of a 2k CCD camera designed for electron crystallography at 400 kV [J].
Downing, KH ;
Hendrickson, FM .
ULTRAMICROSCOPY, 1999, 75 (04) :215-233
[8]   CCD detectors in high-resolution biological electron microscopy [J].
Faruqi, AR ;
Subramaniam, S .
QUARTERLY REVIEWS OF BIOPHYSICS, 2000, 33 (01) :1-27
[9]   2dx -: User-friendly image processing for 2D crystals [J].
Gipson, Bryant ;
Zeng, Xiangyan ;
Zhang, Zi Yan ;
Stahlberg, Henning .
JOURNAL OF STRUCTURAL BIOLOGY, 2007, 157 (01) :64-72
[10]  
Glaeser R M., 2007, ELECT CRYSTALLOGRAPH, DOI DOI 10.1038/nsb891