Improving the detection of simulated masses in mammograms through two different image-processing techniques

被引:15
作者
Hemminger, BM
Zong, SQ
Muller, KE
Coffey, CS
DeLuca, MC
Johnston, RE
Pisano, ED
机构
[1] Univ N Carolina, Dept Radiol, Chapel Hill, NC 27599 USA
[2] Univ N Carolina, Dept Biomed Engn, Chapel Hill, NC 27599 USA
[3] Univ N Carolina, Dept Biostat, Chapel Hill, NC 27599 USA
[4] Univ N Carolina, Lineberger Comprehens Canc Ctr, Chapel Hill, NC 27599 USA
[5] Vanderbilt Univ, Div Biostat, Dept Prevent Med, Nashville, TN USA
关键词
breast neoplasms; diagnosis; breast radiography; diagnostic radiology; observer performance; images; processing;
D O I
10.1016/S1076-6332(03)80762-6
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Rationale and Objectives. The purpose of this study was to determine whether contrast-limited adaptive histogram equalization (CLAHE) or histogram-based intensity windowing (HIW) improves the detection of simulated masses in dense mammograms. Materials and Methods. Simulated masses were embedded in portions of mammograms of patients with dense breasts; the mammograms were digitized at 50 mum per pixel, 12 bits deep. In two different experiments, images were printed both with no processing applied and with related parameter settings of two image-processing methods. A simulated mass was embedded in a realistic background of dense breast tissue, with its position varied. The key variables in each trial included the position of the mass, the contrast levels of the mass relative to the background, and the selected parameter settings for the image-processing method. Results. The success in detecting simulated masses on mammograms with dense backgrounds depended on the parameter settings of the algorithms used. The best HIW setting performed better than the best fixed-intensity window setting and better than no processing. Performance with the best CLAHE settings was no different from that with no processing. In the HIW experiment, there were no significant differences in observer performance between processing conditions for radiologists and nonradiologists. Conclusion. HIW should be tested in clinical images to determine whether the detection of masses by radiologists can be improved. CLAHE processing will probably not improve the detection of masses on clinical mammograms.
引用
收藏
页码:845 / 855
页数:11
相关论文
共 26 条
[1]   IMAGE FEATURE ANALYSIS AND COMPUTER-AIDED DIAGNOSIS IN DIGITAL RADIOGRAPHY .1. AUTOMATED DETECTION OF MICROCALCIFICATIONS IN MAMMOGRAPHY [J].
CHAN, HP ;
DOI, K ;
GALHOTRA, S ;
VYBORNY, CJ ;
MACMAHON, H ;
JOKICH, PM .
MEDICAL PHYSICS, 1987, 14 (04) :538-548
[2]   DIGITAL MAMMOGRAPHY - ROC STUDIES OF THE EFFECTS OF PIXEL SIZE AND UNSHARP-MASK FILTERING ON THE DETECTION OF SUBTLE MICROCALCIFICATIONS [J].
CHAN, HP ;
VYBORNY, CJ ;
MACMAHON, H ;
METZ, CE ;
DOI, K ;
SICKLES, EA .
INVESTIGATIVE RADIOLOGY, 1987, 22 (07) :581-589
[3]  
HALE DA, 1994, J SURG ONCOL, V55, P42
[4]   INTRODUCTION TO PERCEPTUAL LINEARIZATION OF VIDEO DISPLAY SYSTEMS FOR MEDICAL IMAGE PRESENTATION [J].
HEMMINGER, BM ;
JOHNSTON, RE ;
ROLLAND, JP ;
MULLER, KE .
JOURNAL OF DIGITAL IMAGING, 1995, 8 (01) :21-34
[5]   Effect of display luminance on the feature detection rates of masses in mammograms [J].
Hemminger, BM ;
Dillon, AW ;
Johnston, RE ;
Muller, KE ;
Deluca, MC ;
Coffey, CS ;
Pisano, ED .
MEDICAL PHYSICS, 1999, 26 (11) :2266-2272
[6]  
HEMMINGER BM, 1992, SPIE MED IMAGING, V6, P164
[7]  
HOMER MJ, 1991, MAMMOGRAPHIC INTERPR, P4
[8]  
Jain AK., 1989, Fundamentals of Digital Image Processing
[9]   Digital mammography using storage phosphor plate technique - Optimizing image processing parameters for the visibility of lesions and anatomy [J].
Kheddache, S ;
Kvist, H .
EUROPEAN JOURNAL OF RADIOLOGY, 1997, 24 (03) :237-244
[10]   ENHANCED IMAGE MAMMOGRAPHY [J].
MCSWEENEY, MB ;
SPRAWLS, P ;
EGAN, RL .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1983, 140 (01) :9-14