A comparison of the intubating and standard laryngeal mask airways for airway management by inexperienced personnel

被引:25
作者
Choyce, A [1 ]
Avidan, MS [1 ]
Shariff, A [1 ]
Del Aguila, M [1 ]
Radcliffe, JJ [1 ]
Chan, T [1 ]
机构
[1] Kings Coll London Hosp, Dept Anaesthesia, London SE5 9RS, England
关键词
equipment; laryngeal mask airway; intubating laryngeal mask airway; resuscitation; training;
D O I
10.1046/j.1365-2044.2001.01708-3.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Twenty-four inexperienced participants were timed inserting the intubating laryngeal mask airway and the laryngeal mask airway in 75 anaesthetised subjects. Adequacy of ventilation was assessed on a three-point scale. The pressure at which a leak first developed around the device's cuff was also measured. There was no significant difference in insertion time or the likelihood of achieving adequate ventilation between devices. However, the intubating laryngeal mask airway was better at providing adequate ventilation without audible leak (58/75 (77%) vs. 42/75 (56%); p = 0.009). The median (range [IQR]) pressure at which an audible leak developed was higher for the intubating laryngeal mask airway, 34.5 (14-40 [29-40]) cmH(2)O, than for the laryngeal mask airway, 27.5 (14-40 [22-33]) cmH(2)O (p<0.001). The intubating laryngeal mask airway is worthy of further consideration as a tool for emergency airway management for inexperienced personnel.
引用
收藏
页码:357 / 360
页数:4
相关论文
共 18 条
[1]   A COMPARISON OF THE LARYNGEAL MASK AIRWAY AND GUEDEL AIRWAY, BAG AND FACEMASK FOR MANUAL VENTILATION FOLLOWING FORMAL TRAINING [J].
ALEXANDER, R ;
HODGSON, P ;
LOMAX, D ;
BULLEN, C .
ANAESTHESIA, 1993, 48 (03) :231-234
[2]   Appropriate size and inflation of the laryngeal mask airway [J].
Asai, T ;
Howell, TK ;
Koga, K ;
Morris, S .
BRITISH JOURNAL OF ANAESTHESIA, 1998, 80 (04) :470-474
[3]   The intubating laryngeal mask airway compared with direct laryngoscopy [J].
Avidan, MS ;
Harvey, A ;
Chitkara, N ;
Ponte, J .
BRITISH JOURNAL OF ANAESTHESIA, 1999, 83 (04) :615-617
[4]   The intubating laryngeal mask - Results of a multicentre trial with experience of 500 cases [J].
Baskett, PJF ;
Parr, MJA ;
Nolan, JP .
ANAESTHESIA, 1998, 53 (12) :1174-1179
[5]   The intubating laryngeal mask .1. Development of a new device for intubation of the trachea [J].
Brain, AIJ ;
Verghese, C ;
Addy, EV ;
Kapila, A .
BRITISH JOURNAL OF ANAESTHESIA, 1997, 79 (06) :699-703
[6]  
BRAIN AIJ, 1998, INTUBATING LARYNGEAL
[7]  
BRIMACOME JR, 1996, LARYNGEAL MASK AIRWA
[8]  
Choyce A, 2000, BRIT J ANAESTH, V84, P103
[9]   LARYNGEAL MASK AIRWAY AND TRACHEAL TUBE INSERTION BY UNSKILLED PERSONNEL [J].
DAVIES, PRF ;
TIGHE, SQM ;
GREENSLADE, GL ;
EVANS, GH .
LANCET, 1990, 336 (8721) :977-979
[10]   Influence of cuff volume on oropharyngeal leak pressure and fibreoptic position with the laryngeal mask airway [J].
Keller, C ;
Pühringer, F ;
Brimacombe, JR .
BRITISH JOURNAL OF ANAESTHESIA, 1998, 81 (02) :186-187