CNT/PDMS Composite Flexible Dry Electrodes for Long-Term ECG Monitoring

被引:355
作者
Jung, Ha-Chul [2 ]
Moon, Jin-Hee [2 ]
Baek, Dong-Hyun [2 ,3 ]
Lee, Jae-Hee [1 ]
Choi, Yoon-Young [2 ]
Hong, Joung-Sook [1 ]
Lee, Sang-Hoon [2 ]
机构
[1] Soongsil Univ, Dept Chem Engn, Seoul 156743, South Korea
[2] Korea Univ, Dept Biomed Engn, Coll Hlth Sci, Seoul 136703, South Korea
[3] Korea Univ, Sch Elect Engn, Seoul 136703, South Korea
基金
新加坡国家研究基金会;
关键词
Biocompatible; carbon nanotube (CNT)/polydimethylsiloxane (PDMS) composite electrode; ECG electrode; long-term monitoring; CARBON NANOTUBE; FABRICATION; DISPERSION; IMPEDANCE;
D O I
10.1109/TBME.2012.2190288
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We fabricated a carbon nanotube (CNT)/polydimethylsiloxane (PDMS) composite-based dry ECG electrode that can be readily connected to conventional ECG devices, and showed its long-term wearable monitoring capability and robustness to motion and sweat. While the dispersion of CNTs in PDMS is challenging, we optimized the process to disperse untreated CNTs within PDMS by mechanical force only. The electrical and mechanical characteristics of the CNT/PDMS electrode were tested according to the concentration of CNTs and its thickness. The performances of ECG electrodes were evaluated by using 36 types of electrodes which were fabricated with different concentrations of CNTs, and with a differing diameter and thickness. The ECG signals were obtained by using electrodes of diverse sizes to observe the effects of motion and sweat, and the proposed electrode was shown to be robust to both factors. The CNT concentration and diameter of the electrodes were critical parameters in obtaining high-quality ECG signals. The electrode was shown to be biocompatible from the cytotoxicity test. A seven-day continuous wearability test showed that the quality of the ECG signal did not degrade over time, and skin reactions such as itching or erythema were not observed. This electrode could be used for the long-term measurement of other electrical biosignals for ubiquitous health monitoring including EMG, EEG, and ERG.
引用
收藏
页码:1472 / 1479
页数:8
相关论文
共 34 条
[1]   A dry release of polyimide electrodes using Kapton film and application to EEG signal measurements [J].
Baek, Dong-Hyun ;
Moon, Jin-hee ;
Choi, Yoon Young ;
Lee, Mina ;
Choi, Jee Hyun ;
Pak, James Jungho ;
Lee, Sang-Hoon .
MICROSYSTEM TECHNOLOGIES-MICRO-AND NANOSYSTEMS-INFORMATION STORAGE AND PROCESSING SYSTEMS, 2011, 17 (01) :7-14
[2]   Flexible polymeric dry electrodes for the long-term monitoring of ECG [J].
Baek, Ju-Yeoul ;
An, Jin-Hee ;
Choi, Jong-Min ;
Park, Kwang-Suk ;
Lee, Sang-Hoon .
SENSORS AND ACTUATORS A-PHYSICAL, 2008, 143 (02) :423-429
[3]  
Braecklein M., 2004, P 2 OPENECG WORKSH, V51, P51
[4]   Diffusion of gases in silicone polymers: Molecular dynamics simulations [J].
Charati, SG ;
Stern, SA .
MACROMOLECULES, 1998, 31 (16) :5529-5535
[5]   Continuous carbon nanotube reinforced composites [J].
Ci, L. ;
Suhr, J. ;
Pushparaj, V. ;
Zhang, X. ;
Ajayan, P. M. .
NANO LETTERS, 2008, 8 (09) :2762-2766
[6]   Photopatternable conductive PDMS materials for microfabrication [J].
Cong, Hailin ;
Pan, Tingrui .
ADVANCED FUNCTIONAL MATERIALS, 2008, 18 (13) :1912-1921
[7]   Ambulatory cardiopulmonary data capture [J].
Coyle, M .
2ND ANNUAL INTERNATIONAL IEEE-EMBS SPECIAL TOPIC CONFERENCE ON MICROTECHNOLOGIES IN MEDICINE & BIOLOGY, PROCEEDINGS, 2002, :297-300
[8]   Flexible PDMS-based dry electrodes for electro-optic acquisition of ECG signals in wearable devices [J].
Fernandes, M. S. ;
Lee, K. S. ;
Ram, R. J. ;
Correia, J. H. ;
Mendes, P. M. .
2010 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2010, :3503-3506
[9]   Cellular micropatterns on biocompatible materials [J].
Folch, A ;
Toner, M .
BIOTECHNOLOGY PROGRESS, 1998, 14 (03) :388-392
[10]   TEMPORAL CHANGES IN ELECTRODE IMPEDANCE WHILE RECORDING ELECTROCARDIOGRAM WITH DRY ELECTRODES [J].
GEDDES, LA ;
VALENTINUZZI, ME .
ANNALS OF BIOMEDICAL ENGINEERING, 1973, 1 (03) :356-367