Tandem calibration methodology: dual nebulizer sample introduction for inductively coupled plasma atomic emission spectrometry

被引:21
作者
Hamier, J [1 ]
Salin, ED [1 ]
机构
[1] McGill Univ, Dept Chem, Montreal, PQ H3A 2K6, Canada
关键词
tandem calibration method; dual nebulizer sample introduction; inductively coupled plasma atomic emission spectrometry; high salt matrix;
D O I
10.1039/a801689e
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A calibration method based on the simultaneous introduction of sample and standard into the plasma by two independent devices is described. In this specific implementation, two nebulizers are linked to the ICP by a Y-shaped tube. Figures of merit were determined by comparing the results obtained with those produced using the method of external calibration, 'Difficult samples' containing matrices of C, Na, Al or It in concentrations ranging from 1000 to 10000 ppm mere used to provide matrices that produced significant errors when used with external calibration. The test solutions used for this study were mixtures of some of the following elements: Al, Cu, Rig, Fe, Ni, Pb and Zn in one of the matrices described previously. The comparison between the proposed tandem calibration method (TCM) and the method of external calibration clearly indicates that the TCM appears to be a valid alternative to traditional calibration methods, such as standard additions, matrix-matching and internal standards, that are commonly used when external calibration fails to give valid results for 'difficult samples' introduced into the ICP by means of a nebulizer, Improvement factors in accuracy based on the ratio of the relative error were, on average, of the order of 4 and, in some cases, well in excess of a full order of magnitude, For liquid samples, the precision of the results obtained by the proposed method was limited mainly by the noise of the sample introduction device (about 5% RSD on 'difficult samples' rer sus roughly 1% on clean standards) while the accuracy was limited by both the short-and long-term stability of the arrangement, typically around 7% relative error. In addition, the method was found to be easy to implement on existing equipment, inexpensive and potentially well suited for automation.
引用
收藏
页码:497 / 505
页数:9
相关论文
共 16 条
[1]  
ALLEN LA, 1996, 23 FACCS C KANS CIT
[2]   A STUDY OF DIRECT ANALYSIS OF SOLID SAMPLES USING SPARK ABLATION COMBINED WITH EXCITATION IN AN INDUCTIVELY COUPLED PLASMA [J].
AZIZ, A ;
BROEKAERT, JAC ;
LAQUA, K ;
LEIS, F .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1984, 39 (9-11) :1091-1103
[3]   SOLID SAMPLING WITH INDUCTIVELY COUPLED PLASMA-MASS SPECTROMETRY - A SURVEY [J].
BAUMANN, H .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1992, 342 (12) :907-916
[4]   THE PERFORMANCE OF A LOW CONSUMPTION THERMOSPRAY NEBULIZER FOR SPECIFIC USE IN MICRO-HPLC AND GENERAL USE IN FI WITH ICP-AES DETECTION [J].
ELGERSMA, JW ;
BALKE, J ;
MAESSEN, FJMJ .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1991, 46 (6-7) :1073-1088
[5]   CORRECTION OF MINERAL ACID INTERFERENCES IN INDUCTIVELY COUPLED PLASMA OPTICAL-EMISSION SPECTROMETRY ON COPPER AND MANGANESE USING INTERNAL STANDARDIZATION [J].
GARDEN, LM ;
MARSHALL, J ;
LITTLEJOHN, D .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1991, 6 (02) :159-163
[6]   Calibration studies on dried aerosols for laser ablation inductively coupled plasma mass spectrometry [J].
Gunther, D ;
Cousin, H ;
Magyar, B ;
Leopold, I .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1997, 12 (02) :165-170
[7]  
HAMIER J, 1996, 64 C ACFAS MONTR QUE
[8]  
HAMIER J, 1996, 42 ICASS U W ONT LON
[9]  
INGLE JD, 1988, SPECTROCHEMICAL ANAL, P233
[10]   Selection of internal standards for the determination of major and minor elements in silicate rocks and limestones by laser ablation inductively coupled plasma atomic emission spectrometry [J].
Kanicky, V ;
Mermet, JM .
APPLIED SPECTROSCOPY, 1997, 51 (03) :332-336