A sensitive immunosensor using colloidal gold as electrochemical label

被引:38
作者
Chen, Zhao-Peng
Peng, Zhao-Feng
Zhang, Peng
Jin, Xue-Fang
Jiang, Jian-Hui
Zhang, Xiao-Bing
Shen, Guo-Li
Yu, Ru-Qin [1 ]
机构
[1] Hunan Univ, Coll Chem & Chem Engn, State Key Lab Chemo Biosensing & Chemometr, Changsha 410082, Peoples R China
[2] Hunan Normal Univ, Minist Educ, Chem & Chem Engn Coll, Key Lab Chem Biol & Tradit Chinese Med Res, Changsha 410081, Peoples R China
[3] Chinese Acad Sci, Yantai Inst Coastal Zone Res Sustainable Dev, Dept Environm Chem, Yantai 264003, Peoples R China
基金
中国国家自然科学基金;
关键词
immunosensor; colloidal gold; electrochemical label; adsorptive voltammetry;
D O I
10.1016/j.talanta.2007.02.020
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A sensitive immunosensor using colloidal gold as electrochemical label is described. In this method, the capture protein was first immobilized on a carbon paste electrode surface through passive adsorption to bind quantitatively with corresponding antigen and colloidal gold labeled antibody to perform a sandwich assay. To detect the amount of the colloidal gold captured on the electrode surface, the colloid was first oxidized electrochemically to produce AuCl4- ions which were adsorbed strongly on the electrode surface. Adsorptive voltammetry was then employed for the determination of the adsorbed AuCl4- ions. A linear relationship between reduction wave peak cur-rent and the antigen concentration (human IgG) from 10 to 500 ng/ml is obtained with a detection limit of 4.0 ng/ml. (C) 2007 Elsevier B.V. All rights reserved.
引用
收藏
页码:1800 / 1804
页数:5
相关论文
共 16 条
[11]   Colloidal Au-enhanced surface plasmon resonance immunosensing [J].
Lyon, LA ;
Musick, MD ;
Natan, MJ .
ANALYTICAL CHEMISTRY, 1998, 70 (24) :5177-5183
[12]   Detection of subicomolar concentrations of human matrix metalloproteinase-2 by an optical biosensor [J].
Pieper-Fürst, U ;
Kleuser, U ;
Stöcklein, WFM ;
Warsinke, A ;
Scheller, FW .
ANALYTICAL BIOCHEMISTRY, 2004, 332 (01) :160-167
[13]   Au nanoparticle- and silver-enhancement reaction-amplified microgravimetric biosensor [J].
Su, XD ;
Li, SFY ;
O'Shea, SJ .
CHEMICAL COMMUNICATIONS, 2001, (08) :755-756
[14]   Application of the biological conjugate between antibody and colloid Au nanoparticles as analyte to inductively coupled plasma mass spectrometry [J].
Zhang, C ;
Zhang, ZY ;
Yu, BB ;
Shi, JJ ;
Zhang, XR .
ANALYTICAL CHEMISTRY, 2002, 74 (01) :96-99
[15]   Amplified microgravimetric gene sensor using Au nanoparticle modified oligonucleotides [J].
Zhou, XC ;
O'Shea, SJ ;
Li, SFY .
CHEMICAL COMMUNICATIONS, 2000, (11) :953-954
[16]   An amperometric immunosensor based on an electrochemically pretreated carbon-paraffin electrode for complement III (C3) assay [J].
Zhou, YM ;
Hu, SQ ;
Shen, GL ;
Yu, RQ .
BIOSENSORS & BIOELECTRONICS, 2003, 18 (04) :473-481