Validation of the GENEA Accelerometer

被引:463
作者
Esliger, Dale W. [1 ,2 ]
Rowlands, Ann V. [2 ]
Hurst, Tina L. [3 ]
Catt, Michael [3 ,4 ]
Murray, Peter [3 ]
Eston, Roger G. [2 ]
机构
[1] Univ Saskatchewan, Coll Kinesiol, Saskatoon, SK S7N 5B2, Canada
[2] Univ Exeter, Sch Sport & Hlth Sci, Exeter, Devon, England
[3] Unilever Discover, Colworth, W Sussex, England
[4] Newcastle Univ, Inst Ageing & Hlth, Newcastle Upon Tyne NE1 7RU, Tyne & Wear, England
关键词
ACTIVITY MONITOR; MEASUREMENT; ACCELERATION; FREQUENCY; GRAVITY; RELIABILITY; DAILY PHYSICAL-ACTIVITY; ACTIGRAPH ACCELEROMETERS; TRIAXIAL ACCELEROMETER; TECHNICAL RELIABILITY; RT3; ACCELEROMETER; MECHANICAL SETUP; ACTIVITY MONITOR; CALIBRATION; VARIABILITY; CUTPOINTS;
D O I
10.1249/MSS.0b013e31820513be
中图分类号
G8 [体育];
学科分类号
04 ; 0403 ;
摘要
ESLIGER, D. W., A. V. ROWLANDS, T. L. HURST, M. CATT, P. MURRAY, and R. G. ESTON. Validation of the GENEA Accelerometer. Med. Sci. Sports Exerc., Vol. 43, No. 6, pp. 1085-1093, 2011. Purpose: The study aims were: 1) to assess the technical reliability and validity of the GENEA using a mechanical shaker; 2) to perform a GENEA value calibration to develop thresholds for sedentary and light-, moderate-, and vigorous-intensity physical activity; and 3) to compare the intensity classification of the GENEA with two widely used accelerometers. Methods: A total of 47 GENEA accelerometers were attached to a shaker and vertically accelerated, generating 15 conditions of varying acceleration and/or frequency. Reliability was calculated using SD and intrainstrument and inter-instrument coefficients of variation, whereas validity was assessed using Pearson correlation with the shaker acceleration as the criterion. Next, 60 adults wore a GENEA on each wrist and on the waist (alongside an ActiGraph and RT3 accelerometer) while completing 10-12 activity tasks. A portable metabolic gas analyzer provided the criterion measure of physical activity. Analyses involved the use of Pearson correlations to establish criterion and concurrent validity and receiver operating characteristic curves to establish intensity cut points. Results: The GENEA demonstrated excellent technical reliability (CVintra = 1.4%, CVinter = 2.1%) and validity (r = 0.98, P < 0.001) using the mechanical shaker. The GENEA demonstrated excellent criterion validity using V (over bar)O-2 as the criterion (left wrist, r = 0.86; right wrist, r = 0.83; waist, r = 0.87), on par with the waist-worn ActiGraph and RT3. The GENEA demonstrated excellent concurrent validity compared with the ActiGraph (r = 0.92) and the RT3 (r = 0.97). The waist-worn GENEA had the greatest classification accuracy (area under the receiver operating characteristic curve (AUC) = 0.95), followed by the left (AUC = 0.93) and then the right wrist (AUC = 0.90). The accuracy of the waist-worn GENEA was virtually identical with that of the ActiGraph (AUC = 0.94) and RT3 (AUC = 0.95). Conclusion: The GENEA is a reliable and valid measurement tool capable of classifying the intensity of physical activity in adults.
引用
收藏
页码:1085 / 1093
页数:9
相关论文
共 32 条
[1]   A triaxial accelerometer and portable data processing unit for the assessment of daily physical activity [J].
Bouten, CVC ;
Koekkoek, KTM ;
Verduin, M ;
Kodde, R ;
Janssen, JD .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1997, 44 (03) :136-147
[2]  
Brage S., 2003, Measurement in Physical Education and Exercise Science, V7, P101, DOI DOI 10.1207/S15327841MPEE0702_
[3]   Calibration of the RT3 accelerometer for ambulation and nonambulation in children [J].
Chu, Eva Y. W. ;
McManus, Alison M. ;
Yu, Clare C. W. .
MEDICINE AND SCIENCE IN SPORTS AND EXERCISE, 2007, 39 (11) :2085-2091
[4]   Physical activity and inactivity profiling: the next generation [J].
Esliger, Dale W. ;
Tremblay, Mark S. .
APPLIED PHYSIOLOGY NUTRITION AND METABOLISM, 2007, 32 :S195-S207
[5]   Technical reliability assessment of three accelerometer models in a mechanical setup [J].
Esliger, Dale W. ;
Tremblay, Mark S. .
MEDICINE AND SCIENCE IN SPORTS AND EXERCISE, 2006, 38 (12) :2173-2181
[6]   Calibration of two objective measures of physical activity for children [J].
Evenson, Kelly R. ;
Catellier, Diane J. ;
Gill, Karminder ;
Ondrak, Kristin S. ;
McMurray, Robert G. .
JOURNAL OF SPORTS SCIENCES, 2008, 26 (14) :1557-1565
[7]   Validity of a multi-sensor armband in estimating rest and exercise energy expenditure [J].
Fruin, ML ;
Rankin, JW .
MEDICINE AND SCIENCE IN SPORTS AND EXERCISE, 2004, 36 (06) :1063-1069
[8]   Adaptive control for a multi-axis hydraulic test rig [J].
Gizatullin, A. O. ;
Edge, K. A. .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART I-JOURNAL OF SYSTEMS AND CONTROL ENGINEERING, 2007, 221 (I2) :183-198
[9]   The validation of a novel activity monitor in the measurement of posture and motion during everyday activities [J].
Grant, P. M. ;
Ryan, C. G. ;
Tigbe, W. W. ;
Granat, M. H. .
BRITISH JOURNAL OF SPORTS MEDICINE, 2006, 40 (12) :992-997
[10]   Decision boundaries and receiver operating characteristic curves: New methods for determining accelerometer cutpoints [J].
Jago, Russell ;
Zakeri, Issa ;
Baranowski, Tom ;
Watson, Kathy .
JOURNAL OF SPORTS SCIENCES, 2007, 25 (08) :937-944