Countering matrix effects in environmental liquid chromatography-electrospray ionization tandem mass spectrometry water analysis for endocrine disrupting chemicals

被引:174
作者
Benijts, T [1 ]
Dams, R [1 ]
Lambert, W [1 ]
De Leenheer, A [1 ]
机构
[1] Univ Ghent, Toxicol Lab, B-9000 Ghent, Belgium
关键词
water analysis; matrix effects; environmental analysis; endocrine disrupting chemicals;
D O I
10.1016/j.chroma.2003.12.022
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
In recent years, despite the increasing success of liquid chromatography (LC) coupled to tandem mass spectrometry (MS), reports on matrix susceptibility have shown the limitations of the this powerful analytical technique. Matrix effects (MEs) result from co-eluting residual matrix components affecting the ionization efficiency of target analytes and can lead to erroneous results. The present work evaluates the matrix effect of environmental water samples on 35 endocrine disrupting chemicals (EDCs) in negative and positive LC-ESI-MS/MS. It was shown that mobile-phase additives could significantly influence matrix effects. Addition of acids resulted in a severe signal suppression (average ME%: <65%), and 1 mM ammonium formate increased the average ME% to 84%. The importance of an efficient sample clean-up and internal standardization also was demonstrated. Cleaner extracts resulted in reduced matrix effects (average ME%: 89%) and labeled internal standards proved to have a beneficial effect especially on signal reproducibility (average CV% 4.2% versus 2.6%). The results from the present work indicate that evaluation of matrix effects should become an integrated part of quantitative LC-ESI-MS/MS method development and validation. (C) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:153 / 159
页数:7
相关论文
共 49 条
[1]   Quantitative characterization of differential ion suppression on liquid chromatography/atmospheric pressure ionization mass spectrometric bioanalytical methods [J].
Avery, MJ .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2003, 17 (03) :197-201
[2]   Analysis of estrogenic contaminants in river water using liquid chromatography coupled to ion trap based mass spectrometry [J].
Benijts, T ;
Dams, R ;
Günther, W ;
Lambert, W ;
De Leenheer, A .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2002, 16 (14) :1358-1364
[3]  
Bonfiglio R, 1999, RAPID COMMUN MASS SP, V13, P1175, DOI 10.1002/(SICI)1097-0231(19990630)13:12<1175::AID-RCM639>3.3.CO
[4]  
2-S
[5]   Mechanistic aspects of electrospray ionization [J].
Bruins, AP .
JOURNAL OF CHROMATOGRAPHY A, 1998, 794 (1-2) :345-357
[6]   Quantitation of SR 27417 in human plasma using electrospray liquid chromatography tandem mass spectrometry: A study of ion suppression [J].
Buhrman, DL ;
Price, PI ;
Rudewicz, PJ .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 1996, 7 (11) :1099-1105
[7]   LC-MS/MS signal suppression effects in the analysis of pesticides in complex environmental matrices [J].
Choi, BK ;
Hercules, DM ;
Gusev, AI .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 2001, 369 (3-4) :370-377
[8]   Influence of the eluent composition on the ionization efficiency for morphine of pneumatically assisted electrospray, atmospheric-pressure chemical ionization and sonic spray [J].
Dams, R ;
Benijts, T ;
Günther, W ;
Lambert, W ;
De Leenheer, A .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2002, 16 (11) :1072-1077
[9]   High-speed gradient parallel liquid chromatography/tandem mass spectrometry with fully automated sample preparation for bioanalysis: 30 seconds per sample from plasma [J].
Deng, Y ;
Wu, JT ;
Lloyd, TL ;
Chi, CL ;
Olah, TV ;
Unger, SE .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2002, 16 (11) :1116-1123
[10]   Evaluation of a method based on liquid chromatography electrospray mass spectrometry for analyzing carbamate insecticides in fruits and vegetables [J].
DiCorcia, A ;
Crescenzi, C ;
Lagana, A .
JOURNAL OF AGRICULTURAL AND FOOD CHEMISTRY, 1996, 44 (07) :1930-1938