Dose measurements near a non-radioactive gold seed using radiographic film

被引:26
作者
Chow, JCL
Grigorov, GN
机构
[1] Grand River Hosp, Dept Med Phys, Grand River Reg Canc Ctr, Kitchener, ON N2G 1G3, Canada
[2] Univ Waterloo, Dept Phys, Waterloo, ON N2L 3G1, Canada
关键词
D O I
10.1088/0031-9155/50/18/N02
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
The dose distribution near a non-radioactive gold seed under a 6 MV photon beam was measured using radiographic film, water equivalent bolus and solid water slabs. This type of small seed is typically used as a marker in target positional verification using a portal imager for conformal prostate treatment such as intensity modulated radiation therapy. A stack of three films was placed on top of the seed located on a soft bolus. Solid water slabs were then placed on top of the film. The films were exposed using a small I x I cm 2 field. Then, using a similar experimental set-up and exposure, another stack of three films was placed under the seed, which was then covered by the soft bolus and solid water slabs. The cross-plane axial beam profiles at different depths, depending on the thickness of the film package, were measured. From the group of beam profiles above and below the seed, the dose distribution along a selected vertical line within the profiles was easily plotted. Compared to the dose with no seed at the isocentre and 5 cm of solid water, there was about a 21% increase in dose at 0.35 mm above the seed. On the other hand, there was about a 22% decrease in dose at the same distance below the seed. The dosimetry of the calibrated film was verified with a MOSFET detector. The change in dose due to the seed by varying the incident beam angles was also measured for this note.
引用
收藏
页码:N227 / N234
页数:8
相关论文
共 14 条
[1]
Online ultrasound image guidance for radiotherapy of prostate cancer: Impact of image acquisition on prostate displacement [J].
Artignan, X ;
Smitsmans, MHP ;
Lebesque, JV ;
Jaffray, DA ;
van Her, M ;
Bartelink, H .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 59 (02) :595-601
[2]
Intensity-modulated radiation therapy with dynamic multileaf collimators [J].
Boyer, AL ;
Yu, CX .
SEMINARS IN RADIATION ONCOLOGY, 1999, 9 (01) :48-59
[3]
Serum amylase activity disorders in the course of experimental diabetes in rabbits [J].
Burski, K ;
Ueland, T ;
Maciejewski, R .
VETERINARNI MEDICINA, 2004, 49 (06) :197-200
[4]
Extrapolated surface dose measurements with radiochromic film [J].
Butson, MJ ;
Yu, PKN ;
Metcalfe, PE .
MEDICAL PHYSICS, 1999, 26 (03) :485-488
[5]
BUTSON MJ, 1996, MED PHYS, V23, P665
[6]
Intensity modulation delivery techniques: "Step & shoot" MLC auto-sequence versus the use of a modulator [J].
Chang, SX ;
Cullip, TJ ;
Deschesne, KM .
MEDICAL PHYSICS, 2000, 27 (05) :948-959
[7]
Investigation of the use of MOSFET for clinical IMRT dosimetric verification [J].
Chuang, CF ;
Verhey, LJ ;
Xia, P .
MEDICAL PHYSICS, 2002, 29 (06) :1109-1115
[8]
PROSTATE MOTION DURING STANDARD RADIOTHERAPY AS ASSESSED BY FIDUCIAL MARKERS [J].
CROOK, JM ;
RAYMOND, Y ;
SALHANI, D ;
YANG, H ;
ESCHE, B .
RADIOTHERAPY AND ONCOLOGY, 1995, 37 (01) :35-42
[9]
Clinical feasibility study for the use of implanted gold seeds in the prostate as reliable positioning markers during megavoltage irradiation [J].
Dehnad, H ;
Nederveen, AJ ;
van der Heide, UA ;
van Moorselaar, RJA ;
Hofman, P ;
Lagendijk, JJW .
RADIOTHERAPY AND ONCOLOGY, 2003, 67 (03) :295-302
[10]
Prostate motion and isocenter adjustment from ultrasound-based localization during delivery of radiation therapy [J].
Fung, AYC ;
Enke, CA ;
Ayyangar, KM ;
Raman, NV ;
Zhen, WN ;
Thompson, RB ;
Li, SC ;
Nehru, RM ;
Pillai, S .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 61 (04) :984-992