Minimizing UV noise in supercritical fluid chromatography. I. Improving back pressure regulator pressure noise

被引:48
作者
Berger, Terry A. [1 ]
Berger, Blair K. [1 ]
机构
[1] Aurora SFC Syst Inc, Redwood City, CA 94063 USA
关键词
Supercritical fluid chromatography (SFC); Back pressure regulator (BPR); Refractive index (RI); UV detector noise; Pressure noise; Validation; PACKED-COLUMN SFC; DRUG DISCOVERY; AQUEOUS FORMULATION; CHIRAL SEPARATION; DIRECT ASSAY; SUBSTANCE; PURIFICATION; ENANTIOMERS; VALIDATION; IMPURITY;
D O I
10.1016/j.chroma.2011.02.030
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Pressure fluctuations and resulting refractive index changes, induced by the back pressure regulator (BPR) can be a significant source of UV detector noise in supercritical fluid chromatography (SFC). The refractive index (RI) of pure carbon dioxide (CO2) changes approximate to 0.2%/bar at the most commonly used conditions in supercritical fluid chromatography (SFC) (40 degrees C and 100 bar), compared to 0.0045%/bar for water (CO2 IS 44x worse). Changes in RI cause changes in the focal length of the detector cell which results in changes in UV intensity entering the detector. The change in RI (Delta RI/bar) of CO2 decreases 8-fold at 200 bar, compared to 100 bar. A new back pressure regulator (BPR) design representing an order of magnitude improvement in the state of the art is shown to produce peak to peak pressure noise (PNp-p) as low as 0.1 bar, at 200 bar, and 20 Hz, compared to older equipment that attempted to maintain PNp-p < 1 bar, at <5 Hz. With this lower PNp-p, changes in baseline UV offsets could be measured as a function of very small changes in pressure. A pressure change of +/- 1 bar at 100 bar, common with some older BPR's, produced a UV baseline offset >0.5 mAU. A pressure change of +/- 0.5 bar representing the previous state-of-the-art, resulted in a UV offset of 0.3 mAU. Baseline noise <0.05 is required to validate methods for trace analysis. The new BPR, with a PNp-p of 0.1 bar. demonstrated UV peak to peak noise (Np-p) < 0.02 mAU with a >0.03 min (10 Hz) electronic filter under some conditions. This new low noise level makes it possible to validate SFC methods for the first time. (C) 2011 Elsevier B.V. All rights reserved.
引用
收藏
页码:2320 / 2326
页数:7
相关论文
共 28 条
[11]   Superior mirage effect in supercritical CO2:: Experiment and model [J].
Jain, JP ;
Rothe, EW .
JOURNAL OF SUPERCRITICAL FLUIDS, 2005, 35 (03) :260-264
[12]   Effect of system variables involved in packed column supercritical fluid chromatography of stavudine taken as model analyte using response surface methodology along with study of thermodynamic parameters [J].
Kaul, N. ;
Agrawal, H. ;
Paradkar, A. R. ;
Mahadik, K. R. .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2007, 43 (02) :471-480
[13]   Screening approach for chiral separation of pharmaceuticals Part III. Supercritical fluid chromatography for analysis and purification in drug discovery [J].
Maftouh, M ;
Granier-Loyaux, C ;
Chavana, E ;
Marini, J ;
Pradines, A ;
Vander Heyden, Y ;
Picard, C .
JOURNAL OF CHROMATOGRAPHY A, 2005, 1088 (1-2) :67-81
[14]   Validation of direct assay of an aqueous formulation of a drug compound AZY by chiral supercritical fluid chromatography (SFC) [J].
Mukherjee, Partha S. .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2007, 43 (02) :464-470
[15]   A feasibility study on direct assay of an aqueous formulation by chiral supercritical fluid chromatography (SFC) [J].
Mukherjee, Partha S. ;
Cook, Steve E. .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2006, 41 (04) :1287-1292
[16]   Preparative separation and identification of derivatized β-methylphenylalanine enantiomers by chiral SFC, HPLC and NMR for development of new peptide ligand mimetics in drug discovery [J].
Nogle, LM ;
Mann, CW ;
Watts, WL ;
Zhang, YR .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2006, 40 (04) :901-909
[17]   Comparison of drug substance impurity profiles generated with extended length columns during packed-column SFC [J].
Roston, DA ;
Ahmed, S ;
Williams, D ;
Catalano, T .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2001, 26 (03) :339-355
[18]  
SAITO M, Patent No. 4984602
[19]  
SAITO M, 1988, SUPERCRTITICAL FLUID, P211
[20]   A new equation of state for carbon dioxide covering the fluid region from the triple-point temperature to 1100 K at pressures up to 800 MPa [J].
Span, R ;
Wagner, W .