The theoretical benefit of beam fringe compensation and field size reduction for iso-normal tissue complication probability dose escalation in radiotherapy of lung cancer

被引:15
作者
Engelsman, M [1 ]
Remeijer, P [1 ]
van Herk, M [1 ]
Mijnheer, B [1 ]
Damen, E [1 ]
机构
[1] Antoni Van Leeuwenhoek Hosp, Netherlands Canc Inst, Dept Radiotherapy, NL-1066 CX Amsterdam, Netherlands
关键词
dose escalation; IMRT; lung tumor; tumor control probability; conformal radiotherapy;
D O I
10.1118/1.1573208
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
To assess the benefit of beam fringe (50%-90% dose level) sharpening for lung tumors, we performed a numerical simulation in which all geometrical errors (breathing motion, random and systematic errors) are included. A 50 mm diameter lung tumor, located centrally in a lung-equivalent phantom was modeled. Treatment plans were designed with varying number and direction of beams, both with and without the use of intensity modulation to sharpen the beam fringe. Field size and prescribed dose were varied under the constraint of a constant mean lung dose of 20 Gy, which yields a predicted complication probability of about 10%. After numerical simulation of the effect of setup errors and breathing, the resulting dose distribution was evaluated using the minimum dose and the equivalent uniform dose (EUD) in the moving clinical target volume (CTV). When the dose in the CTV was constrained between 95% and 107% of the prescribed dose, the maximum attainable EUD was 71 Gy for a four-field noncoplanar technique with simple conformal beams. When penumbra sharpening was applied using a single beam segment at the edge of the open field, this EUD could be raised to 87 Gy. For a hypothetical infinitely steep penumbra, further escalation to an EUD of 104 Gy was possible. When the dose in the CTV was not constrained, a large escalation of the EUD was possible compared to the constrained case. In this case, the maximum attainable EUD for open fields was 1] 5 Gy, using the four-field noncoplanar technique. The benefit of penumbra sharpening was only modest, with no increase of the EUD for the single-segment technique and a small increase to 125 Gy for the infinitely steep penumbra. From these results we conclude that beam fringe sharpening in combination with field-size reduction leads to a large increase in EUD when a homogeneous target dose is pursued. Further escalation of the EUD is possible when the homogeneity constrained is relaxed, but the relative benefit of beam-fringe sharpening then decreases. 0 2003 American Association of Physicists in Medicine.
引用
收藏
页码:1086 / 1095
页数:10
相关论文
共 34 条
[11]   PITFALLS IN THE USE OF HIGH-ENERGY X-RAYS TO TREAT TUMORS IN THE LUNG [J].
EKSTRAND, KE ;
BARNES, WH .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1990, 18 (01) :249-252
[12]   Field size reduction enables ISO-NTCP escalation of tumor control probability for irradiation of lung tumors [J].
Engelsman, M ;
Remeijer, P ;
van Herk, M ;
Lebesque, JV ;
Mijnheer, BJ ;
Damen, EMF .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2001, 51 (05) :1290-1298
[13]   Impact of simple tissue inhomogeneity correction algorithms on conformal radiotherapy of lung tumours [J].
Engelsman, M ;
Damen, EMF ;
Koken, PW ;
van 't Veld, AA ;
van Ingen, KM ;
Mijnheer, BJ .
RADIOTHERAPY AND ONCOLOGY, 2001, 60 (03) :299-309
[14]   The effect of breathing and set-up errors on the cumulative dose to a lung tumor [J].
Engelsmann, M ;
Damen, EMF ;
De Jaeger, K ;
van Ingen, KM ;
Mijnheer, BJ .
RADIOTHERAPY AND ONCOLOGY, 2001, 60 (01) :95-105
[15]   Clinical dose-volume histogram analysis for pneumonitis after 3D treatment for non-small cell lung cancer (NSCLC) [J].
Graham, MV ;
Purdy, JA ;
Emami, B ;
Harms, W ;
Bosch, W ;
Lockett, MA ;
Perez, CA .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1999, 45 (02) :323-329
[16]  
HOBAN PW, 1992, PHYS ENG SCI MED, V15, P113
[17]   LIMITATIONS OF A PENCIL BEAM APPROACH TO PHOTON DOSE CALCULATIONS IN LUNG-TISSUE [J].
KNOOS, T ;
AHNESJO, A ;
NILSSON, P ;
WEBER, L .
PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (09) :1411-1420
[18]   CHANGES IN THE DOSE-PROFILE OF A 10-MV X-RAY-BEAM WITHIN AND BEYOND LOW-DENSITY MATERIAL [J].
KORNELSEN, RO ;
YOUNG, MEJ .
MEDICAL PHYSICS, 1982, 9 (01) :114-116
[19]   Radiation pneumonitis as a function of mean lung dose: An analysis of pooled data of 540 patients [J].
Kwa, SLS ;
Lebesque, JV ;
Theuws, JCM ;
Marks, LB ;
Munley, MT ;
Bentel, G ;
Oetzel, D ;
Spahn, U ;
Ten Haken, RK ;
Drzymala, RE ;
Purdy, JA ;
Lichter, AS ;
Martel, MK ;
TEN Haken, RK .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1998, 42 (01) :1-9
[20]   OPTIMAL RADIATION BEAM PROFILES CONSIDERING UNCERTAINTIES IN BEAM PATIENT ALIGNMENT [J].
LIND, BK ;
KALLMAN, P ;
SUNDELIN, B ;
BRAHME, A .
ACTA ONCOLOGICA, 1993, 32 (03) :331-342