Signal-to-noise ratio and parallel Imaging performance of a 16-channel receive-only brain coil array at 3.0 Tesla

被引:145
作者
de Zwart, JA
Ledden, PJ
van Gelderen, P
Bodurka, J
Chu, RX
Duyn, JH
机构
[1] NINDS, AMRI Sect, LFMI, NIH, Bethesda, MD 20892 USA
[2] Nova Med Inc, Wakefield, MA USA
[3] NIMH, Funct MRI Facil, NIH, Bethesda, MD 20892 USA
关键词
D O I
10.1002/mrm.10678
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The performance of a 16-channel receive-only RF coil for brain imaging at 3.0 Tesla was investigated using a custom-built 16-channel receiver. Both the image signal-to-noise ratio (SNR) and the noise amplification (g-factor) in sensitivity-encoding (SENSE) parallel imaging applications were quantitatively evaluated. Furthermore, the performance was compared with that of hypothetical coils with one, two, four, and eight elements (n) by combining channels in software during image reconstruction. As expected, both the g-factor and SNR improved substantially with n. Compared to an equivalent (simulated) single-element coil, the 16-channel coil showed a 1.87-fold average increase in brain SNR. This was mainly due to an increase in SNR in the peripheral brain (an up to threefold SNR increase), whereas the SNR increase in the center of the brain was 4%. The incremental SNR gains became relatively small at large n, with a 9% gain observed when n was increased from 8 to 16. Compared to the (larger) product birdcage head coil, SNR increased by close to a factor of 2 in the center, and by up to a factor of 6 in the periphery of the brain. For low SENSE acceleration (rate-2), g-factors leveled off for n > 4, and improved only slightly (1.4% averaged over brain) going from n = 8 to n = 16. Improvements in g for n > 8 were larger for higher acceleration rates, with the improvement for rate-3 averaging 12.0%. Published 2003 Wiley-Liss, Inc.
引用
收藏
页码:22 / 26
页数:5
相关论文
共 14 条
[1]   Scalable multichannel MRI data acquisition system [J].
Bodurka, J ;
Ledden, PJ ;
van Gelderen, P ;
Chu, RX ;
de Zwart, JA ;
Morris, D ;
Duyn, JH .
MAGNETIC RESONANCE IN MEDICINE, 2004, 51 (01) :165-171
[2]  
CHU R, 2003, P 11 ANN M ISMRM TOR, P711
[3]   Application of sensitivity-encoded echo-planar imaging for blood oxygen level-dependent functional brain imaging [J].
de Zwart, JA ;
van Gelderen, P ;
Kellman, P ;
Duyn, JH .
MAGNETIC RESONANCE IN MEDICINE, 2002, 48 (06) :1011-1020
[4]   Design of a SENSE-optimized high-sensitivity MRI receive coil for brain imaging [J].
de Zwart, JA ;
Ledden, PJ ;
Kellman, P ;
van Gelderen, P ;
Duyn, JH .
MAGNETIC RESONANCE IN MEDICINE, 2002, 47 (06) :1218-1227
[5]   AN EFFICIENT, HIGHLY HOMOGENEOUS RADIOFREQUENCY COIL FOR WHOLE-BODY NMR IMAGING AT 1.5-T [J].
HAYES, CE ;
EDELSTEIN, WA ;
SCHENCK, JF ;
MUELLER, OM ;
EASH, M .
JOURNAL OF MAGNETIC RESONANCE, 1985, 63 (03) :622-628
[6]  
KING SB, 2001, P 9 ANN M ISMRM GLAS, P1090
[7]  
Larkman DJ, 2001, J MAGN RESON IMAGING, V13, P313, DOI 10.1002/1522-2586(200102)13:2<313::AID-JMRI1045>3.3.CO
[8]  
2-N
[9]  
LEDDEN P, 2003, P 11 ANN M ISMRM TOR, P466
[10]  
LEDDEN PJ, 2001, P 9 SCI M ISMRM GLAS, P1117