Schemes for the optimization of chest radiography using a computer model of the patient and x-ray imaging system

被引:29
作者
Sandborg, M [1 ]
McVey, G
Dance, DR
Carlsson, GA
机构
[1] Linkoping Univ, Fac Hlth Sci, Dept Radiat Phys, SE-58185 Linkoping, Sweden
[2] Royal Marsden NHS Trust, Joint Dept Phys, London SW3 6JJ, England
[3] Linkoping Univ, Fac Hlth Sci, Dept Radiat Phys, SE-58185 Linkoping, Sweden
关键词
chest radiography; voxel phantom; contrast; SNR; optimization;
D O I
10.1118/1.1405840
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 [临床医学]; 100207 [影像医学与核医学]; 1009 [特种医学];
摘要
A computer program has been developed to model chest radiography. It incorporates a voxel phantom of an adult and includes antiscatter grid, radiographic screen, and film. Image quality is quantified by calculating the contrast (Delta OD) and the ideal observer signal-to-noise ratio (SNRI) for a number of relevant anatomical details at various positions in the anatomy. Detector noise and system unsharpness are modeled and their influence on image quality is considered. A measure of useful dynamic range is computed and defined as the fraction of the image that is reproduced at an optical density such that the film gradient exceeds a preset value. The effective dose is used as a measure of the radiation risk for the patient. A novel approach to patient dose and image quality optimization has been developed and implemented. It is based on a reference system acknowledged to yield acceptable image quality in a clinical trial. Two optimizations schemes have been studied, the first including, the contrast of vessels as measure of image quality and the second scheme using also the signal-to-noise ratio of calcifications. Both schemes make use of our measure of useful dynamic range as a key quantity. A large variety of imaging conditions was simulated by varying the tube voltage, antiscatter device, screen-film system, and maximum optical density in the computed image. It was found that the optical density is crucial in screen-film chest radiography. Significant dose savings (30%-50%) can be accomplished without sacrificing image quality by using low-atomic-number grids with a low grid ratio or an air gap and more sensitive screen-film system. Dose-efficient configurations proposed by the model agree well with the example of good radiographic technique suggested by the European Commission. (C) 2001 American Association of Physicists in Medicine.
引用
收藏
页码:2007 / 2019
页数:13
相关论文
共 32 条
[1]
Optimum tube voltage for chest radiographs obtained by psychophysical analysis [J].
Asai, Y ;
Tanabe, Y ;
Ozaki, Y ;
Kubota, H ;
Matsumoto, M ;
Kanamori, H .
MEDICAL PHYSICS, 1998, 25 (11) :2170-2175
[2]
Berger M. J., 1987, 873597 NBSIR
[3]
Birch R., 1979, SCI REPORT SERIES, V30
[4]
Carmichael J. H. E., 1996, 16260 EUR EN
[5]
CALCULATION OF DOSE AND CONTRAST FOR 2 MAMMOGRAPHIC GRIDS [J].
DANCE, DR ;
PERSLIDEN, J ;
CARLSSON, GA .
PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (01) :235-248
[6]
The use of carbon fibre material in radiographic cassettes: Estimation of the dose and contrast advantages [J].
Dance, DR ;
Lester, SA ;
Carlsson, GA ;
Sandborg, M ;
Persliden, J .
BRITISH JOURNAL OF RADIOLOGY, 1997, 70 (832) :383-390
[7]
Calibration and validation of a voxel phantom for use in the Monte Carlo modeling and optimisation of X-ray imaging systems [J].
Dance, DR ;
McVey, G ;
Sadborg, M ;
Persliden, J ;
Carlsson, GA .
MEDICAL IMAGING 1999: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 1999, 3659 :548-559
[8]
IMAGE-INFORMATION TRANSFER PROPERTIES OF X-RAY INTENSIFYING SCREENS IN THE ENERGY-RANGE FROM 17 TO 320 KEV [J].
GINZBURG, A ;
DICK, CE .
MEDICAL PHYSICS, 1993, 20 (04) :1013-1021
[9]
Hubbell J. H., 1975, Journal of Physical and Chemical Reference Data, V4, P471, DOI 10.1063/1.555523
[10]
RELATIVISTIC ATOMIC FORM-FACTORS AND PHOTON COHERENT SCATTERING CROSS-SECTIONS [J].
HUBBELL, JH ;
OVERBO, I .
JOURNAL OF PHYSICAL AND CHEMICAL REFERENCE DATA, 1979, 8 (01) :69-105