Clinically significant corrosion at the head-neck taper interface in total hip arthroplasty: a systematic review and case series

被引:39
作者
Carli, Alberto [1 ]
Politis, Angelos [2 ]
Zukor, David [2 ]
Huk, Olga [2 ]
Antoniou, John [2 ]
机构
[1] McGill Univ, Montreal, PQ H3T 1E2, Canada
[2] McGill Univ, Jewish Gen Hosp, Lady Davis Inst Med Res, Montreal, PQ H3T 1E2, Canada
关键词
Head-neck corrosion; Taper corrosion; Metallosis; Corrosion and adverse local soft tissue reaction; LARGE-DIAMETER METAL; PSEUDOTUMOR FORMATION; FEMORAL STEMS; FRACTURE; FAILURE; REPLACEMENTS; PROSTHESES; JUNCTION; MODULARITY; RETRIEVAL;
D O I
10.5301/hipint.5000180
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
100224 [整形外科学];
摘要
Corrosion of the head-neck junction of the femoral component in total hip arthroplasty has been associated with symptomatic adverse local tissue reactions, trunion fracture and elevated serum metal ions. An analysis of risk factors and treatment strategies for corrosion at this interface is lacking in the literature. We therefore performed a systematic review of AAOS proceedings, MEDLINE and EMBASE databases, and included our own case series. A total of 24 articles representing 776 cases of head-neck corrosion met inclusion criteria. The combination of large femoral head sizes and small taper dimensions comprised the majority of published corrosion cases. Revision to ceramic head and ceramic/polyethylene liner was the most commonly utilised treatment. Coating precipitation, mixed alloy coupling and head-neck modulus mismatch collectively appear to contribute to the corrosive process.
引用
收藏
页码:7 / 14
页数:8
相关论文
共 36 条
[1]
Metal ion levels in the blood of patients after hip resurfacing: A comparison between twenty-eight and thirty-six-millimeter-head metal-on-metal prostheses [J].
Antoniou, John ;
Zukor, David J. ;
Mwale, Fackson ;
Minarik, William ;
Petit, Alain ;
Huk, Olga L. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2008, 90A :142-148
[2]
Corrosion-Induced Fracture of a Double-Modular Hip Prosthesis A Case Report [J].
Atwood, Sara A. ;
Patten, Eli W. ;
Bozic, Kevin J. ;
Pruitt, Lisa A. ;
Ries, Michael D. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2010, 92A (06) :1522-1525
[3]
Polyethylene wear in cases using femoral stems of similar geometry, but different metals, porous layer, and modularity [J].
Bal, BS ;
Vandelune, D ;
Gurba, DM ;
Jasty, M ;
Harris, WH .
JOURNAL OF ARTHROPLASTY, 1998, 13 (05) :492-499
[4]
HIP-JOINT LOADING DURING WALKING AND RUNNING, MEASURED IN 2 PATIENTS [J].
BERGMANN, G ;
GRAICHEN, F ;
ROHLMANN, A .
JOURNAL OF BIOMECHANICS, 1993, 26 (08) :969-990
[5]
High failure rates with a large-diameter hybrid metal-on-metal total hip replacement CLINICAL, RADIOLOGICAL AND RETRIEVAL ANALYSIS [J].
Bolland, B. J. R. F. ;
Culliford, D. J. ;
Langton, D. J. ;
Millington, J. P. S. ;
Arden, N. K. ;
Latham, J. M. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2011, 93B (05) :608-615
[6]
Bonnaig Nicolas S, 2011, Orthopedics, V34, P132, DOI 10.3928/01477447-20101221-36
[7]
Trunion fracture of a fully porous-coated femoral stem - Case report [J].
Botti, TP ;
Gent, J ;
Martell, JM ;
Manning, DW .
JOURNAL OF ARTHROPLASTY, 2005, 20 (07) :943-945
[8]
FRETTING CORROSION ACCELERATES CREVICE CORROSION OF MODULAR HIP TAPERS [J].
BROWN, SA ;
FLEMMING, CAC ;
KAWALEC, JS ;
PLACKO, HE ;
VASSAUX, C ;
MERRITT, K ;
PAYER, JH ;
KRAAY, MJ .
JOURNAL OF APPLIED BIOMATERIALS, 1995, 6 (01) :19-26
[9]
Femoral Stem Fracture and In Vivo Corrosion of Retrieved Modular Femoral Hips [J].
Carlson, J. Caitlin Huot ;
Van Citters, Douglas W. ;
Currier, John H. ;
Bryant, Amber M. ;
Mayor, Michael B. ;
Collier, John P. .
JOURNAL OF ARTHROPLASTY, 2012, 27 (07) :1389-1396
[10]
Chana R, 2012, J BONE JOINT SURG BR, V94B, P281, DOI [10.1302/0301-620X.94B2, 10.1302/0301-620X.94B2.27247]