Femoral Stem Fracture and In Vivo Corrosion of Retrieved Modular Femoral Hips

被引:59
作者
Carlson, J. Caitlin Huot [1 ]
Van Citters, Douglas W. [1 ]
Currier, John H. [1 ]
Bryant, Amber M. [1 ]
Mayor, Michael B. [1 ]
Collier, John P. [1 ]
机构
[1] Dartmouth Coll, Thayer Sch Engn, Hanover, NH 03755 USA
关键词
stem; fretting; corrosion; modularity; fracture; FRETTING CREVICE CORROSION; ON-METAL BEARINGS; PROSTHESES; WEAR; ARTHROPLASTY; REPLACEMENT; COMPONENTS;
D O I
10.1016/j.arth.2011.11.007
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
100224 [整形外科学];
摘要
A series of 78 retrieved modular hip devices were assessed for fretting and corrosion. Damage was common at both the head-neck junction (54% showing corrosion; 88% showing fretting) and at the stem-sleeve junction (88% corrosion; 65% fretting). Corrosion correlated to in vivo duration, patient activity, and metal (vs ceramic) femoral heads but did not correlate to head carbon content. Femoral stem fatigue fracture was observed in seven retrievals; all had severe corrosion, were under increased stress, and were in vivo longer than the non-fractured cohort. This study emphasizes the potential for stem fracture when small diameter femoral stems with large offsets are used in heavy and active patients. Designs which reduce fretting and corrosion in modular implants is warranted as patients demand longer lasting implants.
引用
收藏
页码:1389 / 1396
页数:8
相关论文
共 30 条
[1]
AMSTUTZ HC, 1996, CLIN ORTHOP S, V329, pS297
[2]
BOBYN JD, 1994, CLIN ORTHOP RELAT R, P27
[3]
Callaghan J.J., 2007, ADULT HIP, VII
[4]
CORROSION BETWEEN THE COMPONENTS OF MODULAR FEMORAL HIP PROSTHESES [J].
COLLIER, JP ;
SURPRENANT, VA ;
JENSEN, RE ;
MAYOR, MB ;
SURPRENANT, HP .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (04) :511-517
[5]
Collier JP, 1992, CLIN ORTHOP RELAT RE, V129
[6]
Collier JP, 1991, CLIN ORTHOP RELAT RE, V305
[7]
CORROSION AND WEAR AT THE MODULAR INTERFACE OF UNCEMENTED FEMORAL STEMS [J].
COOK, SD ;
BARRACK, RL ;
CLEMOW, AJT .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1994, 76B (01) :68-72
[8]
Corrosion at the Stem-Sleeve Interface of a Modular Titanium Alloy Femoral Component as a Reason for Impaired Disengagement [J].
Fraitzl, Christian R. ;
Moya, Luis E. ;
Castellani, Lorenzo ;
Wright, Timothy M. ;
Buly, Robert L. .
JOURNAL OF ARTHROPLASTY, 2011, 26 (01) :113-119
[9]
Fretting-corrosion of materials used as orthopaedic implants [J].
Geringer, J ;
Forest, B ;
Combrade, P .
WEAR, 2005, 259 :943-951
[10]
Fretting Crevice Corrosion of Stainless Steel Stem-CoCr Femoral Head Connections: Comparisons of Materials, Initial Moisture, and Offset Length [J].
Gilbert, Jeremy L. ;
Mehta, Manav ;
Pinder, Bryan .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2009, 88B (01) :162-173