The In Vivo Linear and Volumetric Wear of Hip Resurfacing Implants Revised for Pseudotumor

被引:45
作者
Glyn-Jones, Sion [1 ]
Roques, Anne [1 ]
Taylor, Adrian [1 ]
Kwon, Young-Min [1 ]
McLardy-Smith, Peter [1 ]
Gill, Harinderjit S. [1 ]
Walter, William [1 ]
Tuke, Mike [1 ]
Murray, David [1 ]
机构
[1] Univ Oxford, NDORMS, Biomed Res Unit, Oxford OX3 7LD, England
关键词
ON-METAL BEARINGS; FOLLOW-UP; ARTHROPLASTY; REPLACEMENT; COMPONENT; HYPERSENSITIVITY; SERIES; GROIN;
D O I
10.2106/JBJS.J.01206
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
100224 [整形外科学];
摘要
Background: Metal-on-metal arthroplasty-related pseudotumors can cause severe local destruction of bone and soft tissues. The cause of pseudotumors is unknown, although some authors have implicated metal wear debris. The aim of this study was to measure the location and magnitude of wear on resurfacing devices that were retrieved during revision procedures for pseudotumor (the pseudotumor group) and for other reasons (the control group). Methods: We examined thirty-six hip-resurfacing implants, which were divided into two groups: eighteen implants from patients with a diagnosis of pseudotumor and eighteen control implants. Implant orientation and patient demographics were recorded. Three-dimensional, contactless metrology was used to scan the surface of the femoral and acetabular components to a resolution of 20 nm. Linear and volumetric wear were measured, and the components were examined for evidence of edge wear. Results: There was three times more total linear wear and over six times more total volumetric wear of the femoral and acetabular components in the pseudotumor group as compared with that in the control group. The mean linear wear rate and standard deviation of the femoral components in the pseudotumor group (8.4 +/- 8.7 mu m/yr) were significantly greater than those in the control group (2.9 +/- 3.9 mu m/yr; p = 0.01). The mean volumetric wear. rate of the femoral components was also significantly greater in the pseudotumor group (3.3 +/- 5.7 mm(3)/yr) than it was in the control group (0.8 +/- 1.2 mm(3)/yr; p = 0.009). Seventeen of eighteen subjects in the pseudotumor group had edge wear, compared with six of eighteen in the control group (p < 0.001). Conclusions: Implants that were retrieved because of pseudotumor had a significantly higher wear rate and prevalence of edge wear than the control implants did. There was a strong association between pseudotumor and the high levels of wear debris that are generated during edge-loading. However, not all patients with high wear developed pseudotumors, and not all pseudotumors had high wear; therefore, other factors are most likely involved in the cause of pseudotumors. Level of Evidence: Therapeutic Level III. See Instructions for Authors for a complete description of levels of evidence.
引用
收藏
页码:2180 / 2188
页数:9
相关论文
共 38 条
[1]
Surface arthroplasty in young patients with hip arthritis secondary to childhood disorders [J].
Amstutz, HC ;
Su, EP ;
Le Duff, MJ .
ORTHOPEDIC CLINICS OF NORTH AMERICA, 2005, 36 (02) :223-+
[2]
Baur W, 2005, ORTHOPADE, V34, P225, DOI 10.1007/s00132-004-0761-x
[3]
Accuracy of EBRA-FCA in the measurement of migration of femoral components of total hip replacement [J].
Biedermann, R ;
Krismer, M ;
Stöckl, B ;
Mayrhofer, P ;
Ornstein, E ;
Franzén, H .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1999, 81B (02) :266-272
[4]
Improvement in the assessment of wear of total knee replacements using coordinate-measuring machine techniques [J].
Blunt, L. A. ;
Bills, P. J. ;
Jiang, X-Q ;
Chakrabarty, G. .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2008, 222 (H3) :309-318
[5]
Comparing laser and polychromatic confocal optical displacement sensors for the 3D measurement of cylindrical artefacts containing microscopic grooved structures [J].
Boltryk, P. J. ;
Hill, M. ;
McBride, J. W. .
WEAR, 2009, 266 (5-6) :498-501
[6]
Counsell A, 2008, ACTA ORTHOP BELG, V74, P870
[7]
Correlation between inclination of the acetabular component and metal ion levels in metal-on-metal hip resurfacing replacement [J].
De Haan, R. ;
Pattyn, C. ;
Gill, H. S. ;
Murray, D. W. ;
Campbell, P. A. ;
De Smet, K. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2008, 90B (10) :1291-1297
[8]
Risk factors for inflammatory pseudotumour formation following hip resurfacing [J].
Glyn-Jones, S. ;
Pandit, H. ;
Kwon, Y. -M. ;
Doll, H. ;
Gill, H. S. ;
Murray, D. W. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2009, 91B (12) :1566-1574
[9]
The relationship between head-neck ratio and pseudotumour formation in metal-on-metal resurfacing arthroplasty of the hip [J].
Grammatopoulos, G. ;
Pandit, H. ;
Murray, D. W. ;
Gill, H. S. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2010, 92B (11) :1527-1534
[10]
Cystic lesion of the groin due to metallosis:: A rare long-term complication of metal-on-metal total hip arthroplasty [J].
Gruber, Florian W. ;
Boeck, Andreas ;
Trattnig, Siegfried ;
Lintner, Felix ;
Ritschl, Peter .
JOURNAL OF ARTHROPLASTY, 2007, 22 (06) :923-927