Neutron scattering facility for the measurement of nuclear recoil quenching factors

被引:11
作者
Jagemann, T
Jochum, J
Feilitzsch, FV
机构
[1] Univ Tubingen, Inst Phys, D-72076 Tubingen, Germany
[2] Tech Univ Munich, Lehrstuhl AstroTeilchenphys, Dept Phys, D-85748 Garching, Germany
关键词
quenching factor; nuclear recoils; dark matter; neutron scattering;
D O I
10.1016/j.nima.2005.06.060
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A neutron scattering multidetector facility has been designed, set up and put into operation. It is commissioned for the determination of the reduced response of particle detectors where the primary interaction leads to a nuclear recoil instead of an electron recoil. Nuclear recoils are investigated by the scattering of a well collimated monoenergetic neutron beam of 11 MeV which is obtained by using an inverse (p,n) reaction. The performance of this facility is discussed. (c) 2005 Elsevier B.V. All rights reserved.
引用
收藏
页码:245 / 260
页数:16
相关论文
共 25 条
[11]  
GASCON J, 2004, ASTROPARTICLES COSMO, P91
[12]   DETECTABILITY OF CERTAIN DARK-MATTER CANDIDATES [J].
GOODMAN, MW ;
WITTEN, E .
PHYSICAL REVIEW D, 1985, 31 (12) :3059-3063
[13]   The scintillation efficiency of carbon and hydrogen recoils in an organic liquid scintillator for dark matter searches [J].
Hong, J ;
Craig, WW ;
Graham, P ;
Hailey, CJ ;
Spooner, NJC ;
Tovey, DR .
ASTROPARTICLE PHYSICS, 2002, 16 (03) :333-338
[14]  
Knoll G. F., 1989, RAD DETECTION MEASUR
[15]  
MESSOUS MY, 1992, THESIS U C BERNARD L
[16]  
PAXTOR C, 1984, NUCL INSTRUM METHODS, V227, P87
[17]   THE 4 GAP DOUBLE DRIFT BUNCHER AT MUNICH [J].
ROHRER, L ;
JAKOB, H ;
RUDOLPH, K ;
SKORKA, SJ .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1984, 220 (01) :161-164
[18]   Final results of the EDELWEISS-I dark matter search with cryogenic heat-and-ionization Ge detectors -: art. no. 122002 [J].
Sanglard, V ;
Benoit, A ;
Bergé, L ;
Blümer, J ;
Broniatowski, A ;
Censier, B ;
Chabert, L ;
Chapellier, M ;
Chardin, G ;
Charvin, P ;
Collin, S ;
De Jésus, M ;
Deschamps, H ;
Di Stefano, P ;
Dolgorouky, Y ;
Drain, D ;
Dumoulin, L ;
Eitel, K ;
Fesquet, M ;
Fiorucci, S ;
Gascon, J ;
Gerlic, E ;
Gerbier, G ;
Goldbach, C ;
Goyot, M ;
Gros, M ;
Hervé, S ;
Horn, M ;
Juillard, A ;
Karolak, M ;
Kikuchi, C ;
de Lesquen, A ;
Luca, M ;
Mallet, J ;
Marnieros, S ;
Mosca, L ;
Navick, XF ;
Nollez, G ;
Pari, P ;
Schoeffel, L ;
Stern, M ;
Vagneron, L ;
Villar, V .
PHYSICAL REVIEW D, 2005, 71 (12)
[19]   SICANE:: a detector array for the measurement of nuclear recoil quenching factors using a monoenergetic neutron beam [J].
Simon, E ;
Bergé, L ;
Broniatowski, A ;
Bouvier, R ;
Chambon, B ;
De Jésus, M ;
Drain, D ;
Dumoulin, L ;
Gascon, J ;
Hadjout, JP ;
Juillard, A ;
Martineau, O ;
Pastor, C ;
Stern, M ;
Vagneron, L .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2003, 507 (03) :643-656
[20]  
SIMON E, 2001, THESIS U CLAUDE B LY