Conversion coefficients for estimation of effective dose to patients from dose area product during fluoroscopy X-ray examinations

被引:29
作者
Gosch, D.
Gosch, K.
Kahn, T.
机构
[1] Univ Klin Leipzig AoR, Klin & Poliklin Diagnost & Intervant Radiol, D-04103 Leipzig, Germany
[2] Humboldt Univ, Charite, Med Klin Schwrpunkt Psychosomat, D-1086 Berlin, Germany
来源
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN | 2007年 / 179卷 / 10期
关键词
effective dose; dose area product; conversion coefficient;
D O I
10.1055/s-2007-963237
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The effective dose, which is related to the stochastic radiation risk, cannot be measured directly. Therefore, conversion coefficients for estimating the effective dose for patients were calculated for fluoroscopy X-ray examinations from the dose area product which can be easily measured. Materials and Methods: Conversion coefficients were calculated using the PC-based Monte Carlo program PCXMC in software version 1.5.1. Conversion coefficients were determined for tube voltages between 60 W and 110 W, for additional cupper filtration up to 0.9 mm, for different projections (a. p., p. a. and lateral) and X-ray field size (between 20 cm x 20 cm and 40 cm x 40 cm in image plane) and field position (from head to leg). Results: Tube voltage, filtration and field position have a significant effect on conversion coefficients. Conversion coefficients increase as the tube voltage and filtration increase. Depending on the radiation spectrum used, the conversion coefficients may differ up to a factor of 4 for the same X-ray examination. The highest conversion coefficients were found for irradiation of the body region followed by the head and leg region. The effect of the field size and projection is moderate. Conversion coefficients for a. p. projection are higher than those for p. a. and lateral projections. Conclusion: Effective dose can be easily estimated by multiplying relevant conversion coefficients by the dose area product. Using conversion coefficients from the literature that do not take additional cupper filtration into consideration can underestimate the effective dose by a factor or up to 2 or more depending on the filtration used.
引用
收藏
页码:1035 / 1042
页数:8
相关论文
共 33 条
[1]  
ARNAL M L, 1961, Fortschr Geb Rontgenstr Nuklearmed, V95, P323
[2]   Optimization of beam filtering, kV-mA regulation curve, and image intensifier entrance exposure rate to reduce radiation exposure in angiographic fluoroscopy. [J].
Barkhausen, J ;
Schoenfelder, D ;
Nagel, HD ;
Stöblen, F ;
Müller, RD .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 1999, 171 (05) :391-395
[3]  
BOER A, 1994, CIRCULATION, V89, P2710
[4]   Comparison of effective doses obtained from dose-area product and air kerma measurements in interventional radiology [J].
Bor, D ;
Sancak, T ;
Olgar, T ;
Elcim, Y ;
Adanali, A ;
Sanlidilek, U ;
Akyar, S .
BRITISH JOURNAL OF RADIOLOGY, 2004, 77 (916) :315-322
[5]  
CHRISTY M, 1980, ORNLNUREGTM367
[6]  
CHRISTY M, 1980, NUREGCR1159
[7]   EFFECTIVE DOSE - HOW EFFECTIVE FOR PATIENTS [J].
DREXLER, G ;
PANZER, W ;
PETOUSSI, N ;
ZANKL, M .
RADIATION AND ENVIRONMENTAL BIOPHYSICS, 1993, 32 (03) :209-219
[8]  
Drexler G., 1990, CALCULATION DOSE EXT
[9]  
Gosch D, 1989, Radiol Diagn (Berl), V30, P345
[10]   Radiation exposure with 3D rotational angiography of the skull [J].
Gosch, D. ;
Kurze, W. ;
Deckert, F. ;
Schulz, T. ;
Patz, A. ;
Kahn, T. .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2006, 178 (09) :880-885