Dose perturbations and image artifacts caused by carbon-coated ceramic and stainless steel fiducials used in proton therapy for prostate cancer

被引:44
作者
Cheung, Joey [1 ]
Kudchadker, Rajat J. [1 ,2 ]
Zhu, X. Ronald [2 ]
Lee, Andrew K. [3 ]
Newhauser, Wayne D. [1 ,2 ]
机构
[1] Univ Texas Houston, Grad Sch Biomed Sci, Houston, TX 77030 USA
[2] Univ Texas MD Anderson Canc Ctr, Dept Radiat Phys, Houston, TX 77030 USA
[3] Univ Texas MD Anderson Canc Ctr, Dept Radiat Oncol, Houston, TX 77030 USA
关键词
RADIOCHROMIC FILM DOSIMETRY; NONRADIOACTIVE GOLD SEED; MONTE-CARLO SIMULATIONS; RADIATION-THERAPY; MARKERS; BEAM; LOCALIZATION; RADIOTHERAPY; IRRADIATION; IMPACT;
D O I
10.1088/0031-9155/55/23/S13
中图分类号
R318 [生物医学工程];
学科分类号
100103 [病原生物学];
摘要
Image-guided radiation therapy using implanted fiducial markers is a common solution for prostate localization to improve targeting accuracy. However, fiducials that are typically used for conventional photon radiotherapy cause large dose perturbations in patients who receive proton radiotherapy. A proposed solution has been to use fiducials of lower atomic number (Z) materials to minimize this effect in tissue, but the effects of these fiducials on dose distributions have not been quantified. The objective of this study was to analyze the magnitude of the dose perturbations caused by select lower-Z fiducials (a carbon-coated zirconium dioxide fiducial and a plastic-coated stainless steel fiducial) and compare them to perturbations caused by conventional gold fiducials. Sets of phantoms were used to assess select components of the effects on dose. First, the fiducials were assessed for radiographic visibility using both conventional computed tomography (CT) and an on-board kilovoltage imaging device at our proton therapy center. CT streak artifacts from the fiducials were also measured in a separate phantom. Second, dose perturbations were measured downstream of the fiducials using radiochromic film. The magnitude of dose perturbation was characterized as a function of marker material, implantation depth and orientation with respect to the beam axis. The radiographic visibility of the markers was deemed to be acceptable for clinical use. The dose measurements showed that the perpendicularly oriented zirconium dioxide and stainless steel fiducials located near the center of modulation of the proton beam perturbed the dose by less than 10%, but that the same fiducials in a parallel orientation near the end of the range of the beam could perturb the dose by as much as 38%. This suggests that carbon-coated and stainless steel fiducials could be used in proton therapy if they are located far from the end of the range of the beam and if they are oriented perpendicular to the beam axis.
引用
收藏
页码:7135 / 7147
页数:13
相关论文
共 19 条
[1]
Radiochromic film for medical radiation dosimetry [J].
Butson, MJ ;
Yu, PKN ;
Cheung, T ;
Metcalfe, P .
MATERIALS SCIENCE & ENGINEERING R-REPORTS, 2003, 41 (3-5) :61-120
[2]
Monte Carlo simulations of dose near a nonradioactive gold seed [J].
Chow, James C. L. ;
Grigorov, Grigor N. .
MEDICAL PHYSICS, 2006, 33 (12) :4614-4621
[3]
Dose measurements near a non-radioactive gold seed using radiographic film [J].
Chow, JCL ;
Grigorov, GN .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (18) :N227-N234
[4]
Dose perturbations from implanted helical gold markers in proton therapy of prostate cancer [J].
Giebeler, Annelise ;
Fontenot, Jonas ;
Balter, Peter ;
Ciangaru, George ;
Zhu, Ronald ;
Newhauser, Wayne .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2009, 10 (01) :63-70
[5]
A study of image-guided intensity-modulated radiotherapy with fiducials for localized prostate cancer including pelvic lymph nodes [J].
Hsu, Annie ;
Pawlicki, Todd ;
Luxton, Gary ;
Hara, Wendy ;
King, Christopher R. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 68 (03) :898-902
[6]
Intraprostatic fiducials for localization of the prostate gland: Monitoring intermarker distances during radiation therapy to test for marker stability [J].
Kupelian, PA ;
Willoughby, TR ;
Meeks, SL ;
Forbes, A ;
Wagner, T ;
Maach, M ;
Langen, KM .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 62 (05) :1291-1296
[7]
MICROSCOPIC GOLD PARTICLE-BASED FIDUCIAL MARKERS FOR PROTON THERAPY OF PROSTATE CANCER [J].
Lim, Young Kyung ;
Kwak, Jungwon ;
Kim, Dong Wook ;
Shin, Dongho ;
Yoon, Myonggeun ;
Park, Soah ;
Kim, Jin Sung ;
Ahn, Sung Hwan ;
Shin, Jungwook ;
Lee, Se Byeong ;
Park, Sung Yong ;
Pyo, Hong Ryeol ;
Kim, Dae Yong ;
Cho, Kwan Ho .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 74 (05) :1609-1616
[8]
Comparison of megavoltage position verification for prostate irradiation based on bony anatomy and implanted fiducials [J].
Nederveen, AJ ;
Dehnad, H ;
van der Heide, UA ;
van Moorselaar, RJA ;
Hofman, P ;
Lagendijk, JJW .
RADIOTHERAPY AND ONCOLOGY, 2003, 68 (01) :81-88
[9]
Monte Carlo simulations of the dosimetric impact of radiopaque fiducial markers for proton radiotherapy of the prostate [J].
Newhauser, Wayne ;
Fontenot, Jonas ;
Koch, Nicholas ;
Dong, Lei ;
Lee, Andrew ;
Zheng, Yuanshui ;
Waters, Laurie ;
Mohan, Radhe .
PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (11) :2937-2952
[10]
Dosimetric impact of tantalum markers used in the treatment of uveal melanoma with proton beam therapy [J].
Newhauser, Wayne D. ;
Koch, Nicholas C. ;
Fontenot, Jonas D. ;
Rosenthal, Stanley J. ;
Gombos, Dan S. ;
Fitzek, Markus M. ;
Mohan, Radhe .
PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (13) :3979-3990