Sterilization and wear-related failure in first- and second-generation press-fit condylar total knee arthroplasty

被引:38
作者
Griffin, William L.
Fehring, Thomas K.
Pomeroy, Donald L.
Gruen, Thomas A.
Murphy, Jeffrey A.
机构
[1] Orthocarolina Hip & Knee Ctr, Charlotte, NC USA
[2] Orthocarolina Res Inst, Charlotte, NC USA
[3] Arthroplasty Fdn, Louisville, KY USA
[4] Orthoped Hosp, Joint Replacement Inst, Los Angeles, CA USA
[5] DePuy Johnson & Johnson Co, Warsaw, IN USA
关键词
D O I
10.1097/BLO.0b013e31814d4d19
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We compared the incidence of wear-related failures between two large cohorts of patients undergoing total knee arthroplasty implanted with identical modular tibial trays and polyethylene inserts sterilized by different methods. A total of 1183 second-generation press-fit condylar prostheses having inserts packaged and sterilized in an oxygen-free environment were assessed at a minimum 5-year followup (mean, 7.0 years). Wear-related failure was defined as (1) osteolysis greater than 100 mm(2) or (2) revision of the implant resulting from osteolysis, polyethylene wear, chronic synovitis, and/or effusion. Wear-related survivorship was calculated using Kaplan-Meier survival analysis. Results were compared with our previously published study of 1287 first-generation press-fit condylar modular knees having inserts sterilized by gamma irradiation in air at 5-year minimum followup (mean, 7.8 years). The wear-related failure rate for the seeond-generation design was 1.1% and 10-year survivorship was 97.0% compared with 8.3% failure and 87.7% 10-year survival for the first-generation design. For second-generation components, patient age was the only variable correlated with wear-related failure. For first-generation components sterilized in air, several variables were correlated to wear-related failure with shelf age of the polyethylene insert being the most important factor. These data emphasize the dramatic effect improvements in polyethylene manufacturing, specifically sterilization methods, can have on implant survivorship.
引用
收藏
页码:16 / 20
页数:5
相关论文
共 33 条
[1]
Ayers D C, 2001, Instr Course Lect, V50, P421
[2]
Ayers D C, 1997, Instr Course Lect, V46, P205
[3]
Bohl JR, 1999, CLIN ORTHOP RELAT R, P28
[4]
Radiographic definition of pelvic osteolysis following total hip arthroplasty [J].
Claus, AM ;
Engh, CA ;
Sychterz, CJ ;
Xenos, JS ;
Orishimo, KF ;
Engh, CA .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2003, 85A (08) :1519-1526
[5]
THE POSTERIOR STABILIZED TOTAL KNEE PROSTHESIS - ASSESSMENT OF POLYETHYLENE DAMAGE AND OSTEOLYSIS AFTER A 10-YEAR-MINIMUM FOLLOW-UP [J].
COLIZZA, WA ;
INSALL, JN ;
SCUDERI, GR .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1995, 77A (11) :1713-1720
[6]
Impact of gamma sterilization on clinical performance of polyethylene in the knee [J].
Collier, JP ;
Sperling, DK ;
Currier, JH ;
Sutula, LC ;
Saum, KA ;
Mayor, MB .
JOURNAL OF ARTHROPLASTY, 1996, 11 (04) :377-389
[7]
Shelf age of the polyethylene tibial component and outcome of unicondylar knee arthroplasty [J].
Collier, MB ;
Engh, CA ;
Engh, GA .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2004, 86A (04) :763-769
[8]
Backside wear of modular ultra-high molecular weight polyethylene tibial inserts [J].
Conditt, MA ;
Ismaily, SK ;
Alexander, JW ;
Noble, PC .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2004, 86A (05) :1031-1037
[9]
Backside wear of polyethylene tibial inserts: Mechanism and magnitude of material loss [J].
Conditt, MA ;
Thompson, MT ;
Usrey, MM ;
Ismaily, SK ;
Noble, PC .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2005, 87A (02) :326-331
[10]
Currier BH, 1997, CLIN ORTHOP RELAT R, P111