Effect of detector lag on CT noise power spectra

被引:17
作者
Baek, Jongduk [1 ]
Pelc, Norbert J. [2 ,3 ,4 ]
机构
[1] Stanford Univ, Dept Radiol, Stanford, CA 94305 USA
[2] Stanford Univ, Dept Radiol, Stanford, CA 94305 USA
[3] Stanford Univ, Dept Bioengn, Stanford, CA 94305 USA
[4] Stanford Univ, Dept Elect Engn, Stanford, CA 94305 USA
关键词
detector lag; NPS; noise correlation; CT; FLAT-PANEL IMAGERS; X-RAY-DETECTORS; ACTIVE-MATRIX; MERCURIC IODIDE; SIGNAL; PERFORMANCE; HGI2;
D O I
10.1118/1.3589135
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
100231 [临床病理学]; 100902 [航空航天医学];
摘要
Purpose: The authors examined the effect of detector lag on the noise power spectrum (NPS) of CT images reconstructed with filtered backprojection (FBP). Methods: The authors derived an analytical expression of the NPS with detector lag, and then verified it using computer simulations with parallel beam and fan beam geometries. The dependence of the NPS on the amount of lag, location within the scanned field of view (FOV), and the number of views used in the reconstruction (samples per rotation) was investigated using constant and view dependent noise in the raw data. Results: Detector lag introduces noise correlation in the azimuthal direction. The effect on the NPS is a frequency dependent reduction in amplitude. In small regions of the image, the effect is primarily in the frequencies corresponding to the azimuthal direction. The noise blurring and NPS filtering increases with increasing radial distance, and therefore regions at larger radial distances have lower noise power. With the same detector lag response function, the amount of noise correlation and NPS filtering decreases with increasing number of views. Conclusions: The shape of the NPS depends on the detector lag coefficients, location of the region, and the number of views used in the reconstruction. In general, the noise correlation caused by detector lag decreased the amplitude of the NPS. (C) 2011 American Association of Physicists in Medicine. [DOI: 10.1118/1.3589135]
引用
收藏
页码:2995 / 3005
页数:11
相关论文
共 30 条
[1]
Strategies to improve the signal and noise performance of active matrix, flat-panel imagers for diagnostic x-ray applications [J].
Antonuk, LE ;
Jee, KW ;
El-Mohri, Y ;
Maolinbay, M ;
Nassif, S ;
Rong, X ;
Zhao, Q ;
Siewerdsen, JH ;
Street, RA ;
Shah, KS .
MEDICAL PHYSICS, 2000, 27 (02) :289-306
[2]
Local and global 3D noise power spectrum in cone-beam CT system with FDK reconstruction [J].
Baek, Jongduk ;
Pelc, Norbert J. .
MEDICAL PHYSICS, 2011, 38 (04) :2122-2131
[3]
The noise power spectrum in CT with direct fan beam reconstruction [J].
Baek, Jongduk ;
Pelc, Norbert J. .
MEDICAL PHYSICS, 2010, 37 (05) :2074-2081
[4]
Dainty J.C., 1974, Image Science: Principles, Analysis and Evaluation of Photographic-type Imaging Processes
[5]
Investigation of the signal behavior at diagnostic energies of prototype, direct detection, active matrix, flat-panel imagers incorporating polycrystalline HgI2 [J].
Du, Hong ;
Antonuk, Larry E. ;
El-Mohri, Youcef ;
Zhao, Qihua ;
Su, Zhong ;
Yamamoto, Jin ;
Wang, Yi .
PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (05) :1325-1351
[6]
A small-signal approach to temporal modulation transfer functions with exposure-rate dependence and its application to fluoroscopic detective quantum efficiency [J].
Friedman, S. N. ;
Cunningham, I. A. .
MEDICAL PHYSICS, 2009, 36 (08) :3775-3785
[7]
Performance of a 41X41 cm2 amorphous silicon flat panel x-ray detector designed for angiographic and R&F imaging applications [J].
Granfors, PR ;
Aufrichtig, R ;
Possin, GE ;
Giambattista, BW ;
Huang, ZS ;
Liu, JQ ;
Ma, B .
MEDICAL PHYSICS, 2003, 30 (10) :2715-2726
[8]
DQE(f) of an amorphous silicon flat panel x-ray detector: Detector parameter influences and measurement methodology [J].
Granfors, PR ;
Aufrichtig, R .
MEDICAL IMAGING 2000: PHYSICS OF MEDICAL IMAGING, 2000, 3977 :2-13
[9]
Polycrystalline Mercuric Iodide Films on CMOS Readout Arrays [J].
Hartsough, Neal E. ;
Iwanczyk, Jan S. ;
Nygard, Einar ;
Malakhov, Nail ;
Barber, William C. ;
Gandhi, Thulasidharan .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2009, 56 (04) :1810-1816
[10]
Investigation of a solid-state detector for advanced computed tomography [J].
Hsieh, J ;
Gurmen, OE ;
King, KF .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2000, 19 (09) :930-940