Pre-clinical validation of a new partially cemented femoral prosthesis by synergetic use of numerical and experimental methods

被引:47
作者
Viceconti, M [1 ]
Cristofolini, L [1 ]
Baleani, M [1 ]
Toni, A [1 ]
机构
[1] Ist Ortoped Rizzoli, Lab Tecnol Med, I-40136 Bologna, Italy
关键词
hip prosthesis; finite element analysis; stress shielding; primary stability; cement stresses;
D O I
10.1016/S0021-9290(01)00029-X
中图分类号
Q6 [生物物理学];
学科分类号
071011 [生物物理学];
摘要
The present work reports the pre-clinical validation of an innovative partially cemented femoral prosthesis called cement-locked uncemented (CLU) prosthesis. The inventors of the device under investigation claimed that, when compared to a comparable fully cemented stem, the new stem would present various advantages. Two previous experimental studies confirmed that primary stability acid stress shielding were comparable to those of cemented stems. Aim of the present study was to investigate if the remaining claims were confirmed as well. A complete finite element model of the bone-implant complex was created from CT data. The model was validated against in vitro measurements of bone surface strains as well as against primary stability measurements. The peak stresses predicted in the CLU cement mantle were not found significantly lower than those reported in other studies on fully cemented stems. However, once the cement inlet geometry is optimised and the associated stress risers are eliminated, the CLU cement mantle should be subjected to much lower stresses. The stress induced in the sterns by both load cases was well below the fatigue limit of the Ti6Al4 V alloy. Finite element models predicted for all load cases relative motion between cement and metal lower than 60 mum. This amplitude may be fully accommodated by elastic deformations of the cement micro-ridges. The experimental and numerical results showed the validity of the new fixation concept, although a further optimisation of the geometry of the cement pockets is needed in order to further reduce the stresses in the cement. (C) 2001 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:723 / 731
页数:9
相关论文
共 42 条
[1]
ANGST M, 1998, T 8 ANN C EUR ORTH R, P5
[2]
[Anonymous], 1979, CLIN ORTHOP RELAT RE
[3]
*ASM, 1994, MAT PROPERTIES HDB T, P538
[4]
Baleani M, 2000, J BIOMED MATER RES, V50, P605, DOI 10.1002/(SICI)1097-4636(20000615)50:4<605::AID-JBM17>3.0.CO
[5]
2-P
[6]
HIP-JOINT LOADING DURING WALKING AND RUNNING, MEASURED IN 2 PATIENTS [J].
BERGMANN, G ;
GRAICHEN, F ;
ROHLMANN, A .
JOURNAL OF BIOMECHANICS, 1993, 26 (08) :969-990
[7]
A critical analysis of stress shielding evaluation of hip prostheses [J].
Cristofolini, L .
CRITICAL REVIEWS IN BIOMEDICAL ENGINEERING, 1997, 25 (4-5) :409-483
[8]
Mechanical validation of whole bone composite femur models [J].
Cristofolini, L ;
Viceconti, M ;
Cappello, A ;
Toni, A .
JOURNAL OF BIOMECHANICS, 1996, 29 (04) :525-535
[9]
INFLUENCE OF THIGH MUSCLES ON THE AXIAL STRAINS IN A PROXIMAL FEMUR DURING EARLY STANCE IN GAIT [J].
CRISTOFOLINI, L ;
VICECONTI, M ;
TONI, A ;
GIUNTI, A .
JOURNAL OF BIOMECHANICS, 1995, 28 (05) :617-624
[10]
In vitro measured strains in the loaded femur: Quantification of experimental error [J].
Cristofolini, L ;
McNamara, BP ;
Freddi, A ;
Viceconti, M .
JOURNAL OF STRAIN ANALYSIS FOR ENGINEERING DESIGN, 1997, 32 (03) :193-200