Determination of halides by microwave induced plasma and stabilized capacitive plasma atomic emission spectrometry after on-line continuous halogen generation

被引:21
作者
Camuna, JF
Montes, M
Pereiro, R
SanzMedel, A
Katschthaler, C
Gross, R
Knapp, G
机构
[1] UNIV OVIEDO,FAC CHEM,DEPT PHYS & ANALYT CHEM,OVIEDO,SPAIN
[2] GRAZ UNIV TECHNOL,INST ANALYT CHEM MICRO & RADIOCHEM,A-8010 GRAZ,AUSTRIA
基金
奥地利科学基金会;
关键词
continuous halogen; generation; halides; microwave induced plasma; stabilized capacitive plasma;
D O I
10.1016/S0039-9140(96)02055-3
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This paper describes a comparative study of the microwave induced plasma (MIP) and the stabilized capacitive plasma (SCP) for halide determinations. The MIP is generated in a Beenakker cavity TM010 using a tangential flow torch and the SCP consists of a 27.12 MHz discharge sustained in a liquid-cooled, fused silica tube surrounded by two annular electrodes. Both discharges are operated in helium at atmospheric pressure and detection was carried out by Atomic Emission Spectrometry (AES). The halides (I-, Br-, Cl-) are converted to volatile halogens by continuous flow generation based on chemical oxidation and on-line separation from the aqueous phase, via a gas-liquid separator, to be finally introduced into the plasma. The different factors affecting the emission intensity of the volatile halogens generated are compared for both discharges and the analytical performance characteristics are also evaluated. Detection limits of 17 ng ml(-1), 24 ng ml(-1) and 55 ng ml(-1) are obtained for the determination of Cl-, Br-, and I-, respectively, in the ultraviolet-visible (UV-VIS) region using the MIP-AES and 45 ng ml(-1), 135 ng ml(-1) and 400 ng ml(-1) for Cl-, Br-, and I- with the SCP-AES. Lines in the near infrared (NIR) region were also evaluated for the SCP-AES detection; improvements in detection limits higher than 30 times were observed in the NIR region as compared with the UV-VIS with detection limits in the NIR of 1.4 ng ml(-1) for Cl-, 3 ng ml(-1) for Br- and 13 ng ml(-1) for I-. (C) 1997 Elsevier Science B.V.
引用
收藏
页码:535 / 544
页数:10
相关论文
共 26 条
[11]  
FEUERBACHER H, 1988, WINT C PLASM SPECTR
[12]   NEAR-INFRARED NONMETAL ATOMIC EMISSION FROM A HELIUM MICROWAVE-INDUCED PLASMA - ELEMENT RATIO DETERMINATIONS [J].
FREEMAN, JE ;
HIEFTJE, GM .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1985, 40 (04) :653-664
[13]   ATOMIC EMISSION GAS-CHROMATOGRAPHIC DETECTION - CHEMICAL AND SPECTRAL INTERFERENCES IN THE STABILIZED CAPACITIVE PLASMA (SCP) [J].
GROSS, R ;
PLATZER, B ;
LEITNER, E ;
SCHALK, A ;
SINABELL, H ;
ZACH, H ;
KNAPP, G .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1992, 47 (01) :95-106
[14]   EVALUATION OF AN ELECTROTHERMAL VAPORIZATION SAMPLE INTRODUCTION SYSTEM INTO A STABILIZED CAPACITIVELY COUPLED HE-PLASMA (SCP) FOR THE DETERMINATION OF CHLORINE [J].
KATSCHTHALER, C ;
QUAN, X ;
KRIZOVA, H ;
GROSS, R ;
KNAPP, G .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1995, 50 (4-7) :453-462
[15]  
KATSCHTHALER C, 1996, THESIS GRAZ U TECHNO
[16]   HELIUM MICROWAVE-INDUCED PLASMAS FOR ELEMENT-SPECIFIC DETECTION IN CHROMATOGRAPHY [J].
LONG, GL ;
DUCATTE, GR ;
LANCASTER, ED .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1994, 49 (01) :75-87
[17]   SENSITIVE SELECTIVE GAS CHROMATOGRAPHY DETECTOR BASED ON EMISSION SPECTROMETRY OF ORGANIC COMPOUNDS [J].
MCCORMACK, AJ ;
TONG, SC ;
COOKE, WD .
ANALYTICAL CHEMISTRY, 1965, 37 (12) :1470-+
[18]   CONTINUOUS-FLOW DETERMINATION OF TRACE IODINE BY ATMOSPHERIC-PRESSURE HELIUM MICROWAVE-INDUCED PLASMA ATOMIC EMISSION-SPECTROMETRY USING GENERATION OF VOLATILE IODINE FROM IODIDE [J].
NAKAHARA, T ;
YAMADA, S ;
WASA, T .
APPLIED SPECTROSCOPY, 1990, 44 (10) :1673-1678
[19]   ANALYTE VOLATILIZATION PROCEDURE FOR CONTINUOUS-FLOW DETERMINATION OF BROMINE BY ATMOSPHERIC-PRESSURE HELIUM MICROWAVE-INDUCED PLASMA ATOMIC EMISSION-SPECTROMETRY [J].
NAKAHARA, T ;
MORIMOTO, S ;
WASA, T .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1992, 7 (02) :211-217
[20]   GAS-SAMPLING GLOW-DISCHARGE FOR OPTICAL-EMISSION SPECTROMETRY .2. OPTIMIZATION AND EVALUATION FOR THE DETERMINATION OF NONMETALS IN GAS-PHASE SAMPLES [J].
PEREIRO, R ;
STARN, TK ;
HIEFTJE, GM .
APPLIED SPECTROSCOPY, 1995, 49 (05) :616-622